HJ 599-2011Water quality. Determination of TNT. N-cetyl pridinium chloride. Sodium sulfite Spectrophotometric method (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (7 pages)

Need quotation for your management?

Get a formal quotation for HJ 599-2011 in 30 minutes.

Issued by

Ministry of Ecology and Environment

Level / Type

National · Mandatory

Issue date

February 10, 2011

Implementation date

June 1, 2011

Scope

HJ 599-2011 (Water quality. Determination of TNT. N-cetyl pridinium chloride. Sodium sulfite Spectrophotometric method) is available as an English-translated PDF.

HJ 599-2011 — This standard specifies the N-chlorohexadecylpyridine-sulfite spectrophotometric method for the determination of TNT in water. This standard applies to the determination of TNT in the ammunition charging industrial wastewater.

When using a 30 mm cuvette, the method detection limit is 0.05 mg/L and the measurement range is 0.2-4 mg/L. For samples with a TNT concentration higher than the upper limit of the method, the sample can be appropriately diluted and measured.

Found what you need?

Document preview — HJ 599-2011

National Standard of the People's Republic of China

ICS
13.060
Classification
Z 16
Replacing
GB/T 13903-1992

Issued by: Ministry of Ecology and Environment

Contents

  • Foreword
  • 1 Scope of application
  • 2 Principle of the method
  • 3 Interference and elimination
  • 4 Reagents and materials
  • 5 Instruments and equipment
  • 6 Samples
  • 6.1 Sample collection and preservation
  • 6.2 Preparation of samples
  • 6.3 Preparation of blank samples
  • 7 Analysis steps
  • 7.1 Drawing of the calibration curve
  • 7.2 Determination
  • 7.3 Blank test
  • 8 Calculation and representation of results
  • 8.1 Calculation of results
  • 8.2 Representation of results
  • 9 Precision and accuracy
  • 10 Quality Assurance and Quality Control
  • 11 Waste treatment
  • 12 Precautions

Foreword

To protect the environment and protect the human body in order to implement the Environmental Protection Law of the People's Republic of China and the Law of the People's Republic of China on Water Pollution Prevention and Control this standard is formulated for the determination of health and standardization of TNT in water.

This standard specifies the N-chlorohexadecylpyridine-sulfite spectrophotometric method for the determination of TNT in water.

This standard is a revision of the "Measurement Spectrophotometric Method for Water Quality TNT" (GB/T 13903-92).

This standard was first published in 1992. The original standard was drafted. Taiyang Jiangyang Chemical Plant. This standard is the first revision. Major repair the order is as follows.

- The standard name was changed to "Determination of water quality TNT by N-chlorohexadecylpyridine-sodium sulfite spectrophotometry";

- increased interference and elimination clauses;

-- Added blank test content;

-- Added quality assurance and quality control regulations;

-- Added provisions for waste disposal and precautions.

The national environmental protection standard "Water quality approved and issued by the former National Environmental Protection Agency on December 2, 1992 since the implementation of this standard.

The TNT's Determination Spectrophotometry (GB/T 13903-92) is abolished.

This standard was formulated by the Science and Technology Standards Department of the Ministry of Environmental Protection.

This standard is mainly drafted by. Beijing Zhongbing North Environmental Technology Development Co., Ltd., China Ordnance Industry Group Corporation and Liaoyang Qing Yang Special Chemical Co., Ltd.

This standard was approved by the Ministry of Environmental Protection on February 10, 2011.

This standard has been implemented since June 1, 2011.

This standard is explained by the Ministry of Environmental Protection.

Determination of water quality ladder N-chlorohexadecylpyridine-sodium sulfite spectrophotometry

1 Scope of application

This standard specifies the N-chlorohexadecylpyridine-sulfite spectrophotometric method for the determination of TNT in water.

This standard applies to the determination of TNT in the ammunition charging industrial wastewater.

When using a 30 mm cuvette, the method detection limit is 0.05 mg/L and the measurement range is 0.2-4 mg/L.

For samples with a TNT concentration higher than the upper limit of the method, the sample can be appropriately diluted and measured.

2 Principle of the method

The addition of TNT to the sodium sulfite in the sample is sensitized by N-chlorohexadecylpyridine to form a red complex.

The absorbance was measured at a wavelength of 466 nm. Within a certain concentration range, the TNT concentration and absorbance values are in accordance with Lambert-Beer law.

3 Interference and elimination

When the wastewater has a certain color, it will interfere with the measurement. Dilute to 25.0 ml by collecting 10.0 ml of sample as a sample blank.

The absorbance was measured directly without any reagent. The absorbance of the sample blank is subtracted from the absorbance of the sample, and then calculated to eliminate the chromaticity pair the effect of the determination.

4 Reagents and materials

Analytical purification reagents that meet national standards are used for analysis, unless otherwise stated. The experimental water is freshly prepared deionized water or steamed.

Distilled water.

4.1 Concentrated hydrochloric acid. rho(HCl) = 1.19 g/ml.

4.2 Hydrochloric acid solution. 1 1.

4.3 Concentrated ammonia. rho(NH4OH) = 0.91 g/ml.

4.4 Ammonia solution. 1 1.

4.5 Ether. rho(C2H5OC2H5) = 0.71 g/ml.

4.6 Anhydrous ethanol. rho(C2H5OH) = 0.79 g/ml.

4.7 Sodium sulfite solution. rho(Na2SO3) = 0.1 g/ml.

Weigh 10.0 g of anhydrous sodium sulfite dissolved in an appropriate amount of water, dissolve and transfer to a 100 ml volumetric flask, add water to the mark, and mix. this the solution is valid for 3 days.

4.8 N-chlorohexadecylpyridine solution. rho[C6H5N(CH2)15CH3Cl·H2O]=2.5 g/L.

Weigh 0.500 g of cetylpyridinium chloride in an appropriate amount of water, dissolve it, transfer it to a.200 ml volumetric flask, add water to the mark, and mix.

The solution was stored in a brown glass ground bottle and stored stably for 30 days at room temperature.

4.9 TNT. 2,4,6-trinitrotoluene [CH3C6H2(NO2)3].

Weigh 1.0 g of TNT (industrial product) in a small beaker, dissolve it with 10 ml of absolute ethanol (4.6), and filter with medium-speed quantitative filter paper.

Place the filtrate in a fume hood and allow it to dry naturally. Recrystallization twice, spare.

4.10 TNT standard stock solution. rho(TNT) = 50.0 mg/L.

Weigh 0.050 0 g TNT (4.9) in a 1 000 ml beaker, add about 800 ml of water preheated to 70 ° C, and place at 75-79 ° C.

In a constant temperature water bath, stir while heating until completely dissolved. Remove, cool to room temperature and transfer to a 1 000 ml brown volumetric flask.

Add water to the mark and mix. The solution is stored at 2 to 5 ° C in the dark, valid for 30 days, or purchased commercially available certified standards.

4.11 TNT standard use solution. rho (TNT) = 10.0 mg/L.

Measure 50.0 ml of TNT standard stock solution (4.10) in a 250 ml brown volumetric flask, add water to the mark and mix. When used Now equipped.

5 Instruments and equipment

5.1 Visible spectrophotometer. with 30 mm cuvette.

5.2 Constant temperature water bath.

5.3 Separating funnel. 50 ml.

5.4 Plug scale colorimetric tube. 50 ml.

5.5 Common instruments and equipment used in general laboratories.

6 Samples

6.1 Sample collection and preservation

Samples should be collected in brown glass bottles, stored at 0 to 4 ° C in the dark, and measured within 5 days.

6.2 Preparation of samples

6.2.1 When the pH value of the water sample is not 4~9, adjust the pH value with hydrochloric acid solution (4.2) or ammonia solution (4.4).

6.2.2 If there is more suspended matter in the water sample, it can be measured by filtration with quantitative filter paper or after extraction with ether.

Extract with ether, the steps are as follows. Measure 25.0 ml of the sample in a 50 ml separatory funnel, add 15 ml of ether (4.5), vigorously shake Move for 2 min. After standing and layering, the water phase is discarded. The ether phase was transferred to a stoppered colorimetric tube and washed with 2 ml of diethyl ether (4.5).

The funnel, washed ether was incorporated into a plug scale tube. Place the colorimetric tube in a constant temperature water bath (temperature does not exceed 40 ° C) and evaporate to no Ether smell, to be tested.

6.3 Preparation of blank samples

A blank sample was prepared by replacing the sample with water and following the same procedure as in the preparation of the sample (6.2.2).

7 Analysis steps

7.1 Drawing of the calibration curve

7.1.1 Measure 0.00, 0.50, 1.00, 3.00, 5.00, 7.00, 10.00 ml TNT standard use solution (4.11) in 7 points the TNT content in the liquid funnel was 0.00, 5.00, 10.0, 30.0, 50.0, 70.0, 100 µg, respectively. Add water to 25 ml separately, according to 6.2.2 the same procedure is used for the extraction operation.

7.1.2 Add 2 ml of absolute ethanol (4.6) to the wall of the tube with a plug scale of the TNT standard series, add about 10 ml of water.

Shake well.

7.1.3 Add 3 ml of sodium sulfite solution (4.7), mix, add 5 ml of N-chlorohexadecylpyridine solution (4.8), add water to 25 ml scribe line, shake well and let stand for 15 min. The absorbance was measured with a 30 mm cuvette at a wavelength of 466 nm using water as a reference. To suck the luminosity is the ordinate, and the corresponding TNT content (µg) is the abscissa, and a calibration curve is drawn.

Note. When the TNT standard series is not directly extracted, only 0.00, 0.50, 1.00, 3.00, 5.00, 7.00, 10.00 ml TNT standard is used.

The liquid (4.11) was placed in 7 plug-type colorimetric tubes, mixed, and the calibration procedure was drawn according to the same procedure as in 7.1.3.

7.2 Determination

Measure 10.0 ml of the sample filtered by the quantitative filter paper in a stoppered colorimetric tube, and measure the absorbance according to the same procedure as in 7.1.3.

Or a sample of 10.0 ml of diethyl ether extracted in a plug-type colorimetric tube, measured in accordance with the same procedure as in 7.1.2 and 7.1.3.

Absorbance.

Note. When the water sample is colorless and turbid, measure 10.0 ml of the sample in a stoppered colorimetric tube and measure the absorbance according to the same procedure as in 7.1.3.

7.3 Blank test

The blank sample to be tested (6.3) was measured for absorbance according to the same procedure as in 7.2.

8 Calculation and representation of results

8.1 Calculation of results

The mass concentration rho of the TNT in the sample is calculated according to formula (1).

0m m

rho -= (1)

Where. rho--the mass concentration of TNT in the sample, mg/L;

M--the TNT content of the sample obtained on the calibration curve, µg;

M0--the TNT content of the blank sample found on the calibration curve, µg;

V--sample volume, ml.

8.2 Representation of results

When the measurement result is less than 1 mg/L, the two decimal places are retained, and when the measurement result is greater than or equal to 1 mg/L, the three significant numbers are retained.

word.

9 Precision and accuracy

In the laboratory, six parallel determinations were made on the actual samples with TNT mass concentrations of 4.80 mg/L and 10.00 mg/L.

The standard deviations were 1.10% and 0.65%, respectively, and the spiked recovery was 95.9% to 102%.

Five laboratories measured and spiked the uniform samples with a TNT mass concentration of 2.00 mg/L.

The standard deviation was 2.6%, and the spiked recovery was 95% to 110%.

10 Quality Assurance and Quality Control

10.1 At least one full program blank test shall be performed for each batch of samples, and the measured value shall not exceed the method detection limit.

10.2 The correlation coefficient of the calibration curve should be greater than or equal to 0.999.

10.3 Each batch of sample analysis shall have an intermediate checkpoint, and the relative deviation of the measured value from the corresponding point of the calibration curve shall not exceed 10%.

10.4 At least 10% of parallel samples should be determined for each batch of samples, and at least one parallel sample should be determined for less than 10 samples.

The deviation should be less than 20%.

10.5 Each batch of samples should be tested by at least one spiked recovery analysis, and the actual sample spike recovery should be controlled at 90% to 110%.

11 Waste treatment

The waste liquid generated during the test analysis and the remaining actual samples should be collected and disposed of properly.

12 Precautions

12.1 Ether should be stored at low temperature and protected from light. When a milky white precipitate appears when an appropriate amount of sodium sulfite solution (4.7) is added, the new ether should be replaced.

12.2 If there are more petroleum products in the water sample, it can be measured by filtering with double-layer quantitative filter paper.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 7 pages — is available in the English PDF.

How to Download HJ 599-2011

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
7 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 599-2011

$214.00