Scope

HJ 488-2009 is the English-translated version of 水质 氟化物的测定 氟试剂分光光度法.

HJ 488-2009 — This standard specifies the fluorine reagent spectrophotometric method for the determination of fluoride in surface water, groundwater and industrial wastewater.

This standard applies to the determination of fluoride in surface water, groundwater and industrial wastewater. The detection limit of this method is 0.02 mg/L, and the lower limit of determination is 0.08 mg/L.

Found what you need?

Document preview — HJ 488-2009

National Standard of the People's Republic of China

ICS
13.060
Classification
Z 16
Replacing
GB/T 7483-1987

Issued by: Ministry of Ecology and Environment

Contents

  • Foreword
  • 1 Scope of application
  • 2 Principle of the method
  • 3 Reagents and materials
  • 4 Instruments and equipment
  • 5 Interference and elimination
  • 6 Samples
  • 6.1 Acquisition and preservation
  • 6.2 Preparation of samples
  • 7 Analysis steps
  • 7.1 Calibration curve
  • 7.2 Determination
  • 7.3 Blank test
  • 8 Calculation of results
  • 9 Precision and accuracy

Foreword

In order to implement the "Environmental Protection Law of the People's Republic of China" and the "Water Pollution Prevention and Control Law of the People's Republic of China" Body health, standardize monitoring methods for fluoride in water, and develop this standard.

This standard specifies methods for the determination of fluoride in surface water, groundwater and industrial wastewater.

This standard amends the "Fluorine Spectrophotometric Method for the Determination of Fluoride in Water Quality" (GB 7483-87), the original standard drafting list the location is the Baotou City Environmental Monitoring Station. This is the first revision.

The main revisions are as follows.

- Broaden the linear range of the method by adjusting the cuvette;

-- Adjust and revise some text and sentence structure.

The national environmental protection standard "Water quality approved and issued by the former National Environmental Protection Agency on March 14, 1987 since the implementation of this standard.

Determination of Fluoride Fluorescence Spectrophotometry (GB 7483-87) is abolished.

This standard was formulated by the Science and Technology Standards Department of the Ministry of Environmental Protection.

This standard is mainly drafted by. China Environmental Monitoring Center, Liaoning Provincial Environmental Monitoring Center Station.

This standard was approved by the Ministry of Environmental Protection on September 27, 2009.

This standard has been implemented since November 1, 2009.

This standard is explained by the Ministry of Environmental Protection.

Water quality - Determination of fluoride - Fluorescent reagent spectrophotometric method

1 Scope of application

This standard specifies the fluorine reagent spectrophotometric method for the determination of fluoride in surface water, groundwater and industrial wastewater.

This standard applies to the determination of fluoride in surface water, groundwater and industrial wastewater.

The detection limit of this method is 0.02 mg/L, and the lower limit of determination is 0.08 mg/L.

2 Principle of the method

The fluoride ion reacts with the fluorine reagent and cerium nitrate in an acetate buffer medium with a pH of 4.1 to form a blue ternary complex.

The absorbance at 620 nm is proportional to the fluoride ion concentration and the fluoride (F-) is quantified.

3 Reagents and materials

The reagents used in this standard are analytically pure reagents that meet the national standards unless otherwise stated. The experimental water is newly prepared.

Ionic water or fluoride-free distilled water.

3.1 Hydrochloric acid solution. c = 1 mol/L.

Take 8.4 ml of hydrochloric acid dissolved in 100 ml of deionized water.

3.2 Sodium hydroxide solution. c = 1 mol/L.

Weigh 4 g of sodium hydroxide in 100 ml of deionized water.

3.3 Acetone (CH3COCH3).

3.4 Sulfuric acid (H2SO4). rho20 = 1.84 g/ml.

Take 300 ml of sulfuric acid into a 500 ml beaker, place it on a hot plate for 1 h, cool it, and put it into a bottle for later use.

3.5 Glacial acetic acid (CH3COOH).

3.6 Fluoride standard stock solution. Weigh the pure sodium fluoride (NaF) 0.221 0 g which has been dried at 105 °C for 2 h, dissolved in deionized water.

Transfer to a 1 000 ml volumetric flask, dilute to the mark, mix and store in a polyethylene bottle, and use 100 µg of fluoride per ml.

3.7 Fluoride standard use solution. Pipette 20.00 ml of sodium fluoride standard stock solution (3.6), transfer to 1 000 ml volumetric flask, use deionized water Dilute to the mark and store in a polyethylene bottle. This solution contains 2.00 µg of fluoride per ml.

3.8 Fluorine reagent solution. c = 0.001 mol/L.

Weigh 0.193 g of fluoro reagent [3 - methylamine - alizarin - diacetic acid, referred to as ALC, C14H7O4 · CH2N (CH2COOH) 2], add 5 ml Wet in deionized water, add sodium hydroxide solution (3.2) to dissolve it, add 0.125 g of sodium acetate (CH3COONa·3H2O), use salt the acid solution (3.1) was adjusted to pH 5.0, diluted to 500 ml with deionized water, and stored in a brown bottle.

3.9 cerium nitrate solution. c = 0.001 mol/L.

Weigh 0.443 g of lanthanum nitrate [La(NO3)3·6H2O], dissolve it with a small amount of hydrochloric acid solution (3.1), and adjust with 1 mol/L sodium acetate solution.

The pH was 4.1 and diluted to 1 000 ml with deionized water.

3.10 Buffer solution. pH = 4.1.

Weigh 35 g of anhydrous sodium acetate (CH3COONa) dissolved in 800 ml of deionized water and add 75 ml of glacial acetic acid (CH3COOH) (3.5).

Dilute to 1 000 ml with deionized water and adjust the pH to 4.1 on a pH meter with acetic acid or sodium hydroxide solution.

3.11 Mix the developer.

Take the fluorine reagent solution (3.8), buffer solution (3.10), acetone (3.3) and cerium nitrate solution (3.9) at a volume ratio of 3.1.3.3 Mix it. Formulated when ready to use.

4 Instruments and equipment

4.1 Spectrophotometer. A cuvette with an optical path of 30 mm or 10 mm.

4.2 pH meter.

5 Interference and elimination

In 25 ml of color developing solution containing 5 µg of fluoride, the following ions are present in excess of the following contents, which may interfere with the measurement and should be pre-steamed first.

Distillation. Cl- 30 mg; SO42- 5.0 mg; NO3- 3.0 mg; B4O72- 2.0 mg; Mg2 2.0 mg; NH4 1.0 mg; Ca2 0.5 mg.

6 Samples

6.1 Acquisition and preservation

The fluoride sample was measured and collected and stored in a polyethylene bottle.

6.2 Preparation of samples

Unless it is proved that the pretreatment of the sample is unnecessary, the sample may be directly prepared for colorimetry, otherwise it shall be pre-distilled according to Appendix A.

7 Analysis steps

7.1 Calibration curve

Add fluoride standard solution (3.7) 0.00, 1.00, 2.00, 4.00, 6.00, 8.00 ml to 6 25.0 ml volumetric flasks.

Add deionized water to 10 ml, accurately add 10.0 ml of mixed developer (3.11), dilute to the mark with deionized water, shake and place for 30 min.

Absorbance was measured using a 30 mm or 10 mm cuvette at 620 nm with pure water as a reference. Deducting reagent blank (zero concentration) the absorbance is plotted against the absorbance of the fluoride content, which is the calibration curve.

7.2 Determination

Accurately draw 1.00 ~ 10.00 ml of sample (depending on fluoride content in water) in a 25.0 ml volumetric flask, add deionized water to 10 ml, accurately add 10.0 ml of mixed developer (3.11), dilute to the mark with deionized water, and shake well. The following is done according to 7.1. Blank After calibration, the fluoride (F-) content was found on the calibration curve from the absorbance value.

7.3 Blank test

The sample was replaced with water and measured according to the procedure of measuring the sample.

8 Calculation of results

The fluoride (F-) mass concentration in the sample is calculated as follows.

rho =

Where. rho--the concentration of fluoride (F-) in the sample, mg/L;

M--calibration curve of the sample fluorine content, µg;

V--take the sample volume, ml.

The result is accurate to two decimal places.

9 Precision and accuracy

Three laboratories analyzed a uniform distribution standard solution of fluoride containing 0.5 mg/L with a relative standard deviation of 1.2% in the laboratory.

The relative standard deviation is 1.2% and the relative error is -0.8%.

Appendix A

(normative appendix) Determination of fluoride in water quality with pretreatment of interfering ion samples A.1 For water samples with strong acidity and alkalinity, adjust to neutral after 1 mol/L sodium hydroxide solution or 1 mol/L hydrochloric acid solution before measurement.

Line measurement.

A.2 Pre-distillation A.2.1 Distillation unit is shown in Figure A.1.

1-1 000 ml three-necked flask; 2-500 ml three-necked flask; 3-safety tube; 4-250 °C thermometer;

5-condensation tube; 6-receiver bottle; 7-universal electric furnace; 8-water vapor conduit; 9-bolt water

Figure A.1 Distillation unit diagram

A.2.2 Take 20 ml of the test solution and place it in a 500 ml three-necked flask (A.2.1). Add 20 ml of sulfuric acid (3.4) and mix slowly while shaking.

Connect the device according to Figure A.1 and heat up to introduce water vapor at a temperature of 145 °C. Collect the distillate at a distillation rate of 6 to 7 ml per minute to 200 ml, reserved for color development.

Note. The distillation temperature should be strictly controlled at (145 ± 5) °C, otherwise the sulfuric acid will be distilled out, affecting the measurement results.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 7 pages — is available in the English PDF.

Referenced standards

How to Download HJ 488-2009

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
7 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

HJ 488-2009

$85.00