Scope

GB/T 8647.9-2006 is the English-translated version of 镍化学分析方法 碳量的确定.

GB/T 8647.9-2006 - This Part specifies the method for the determination of carbon content in nickel. This Part is applicable to the determination of carbon content in nickel. Determination range: 0.0020%~0.20%.

Found what you need?

Document preview — GB/T 8647.9-2006

National Standard of the People's Republic of China

ICS
77.040.30
Classification
H 13

Issued by: General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China

Contents

  • Foreword...3
  • 1 Scope...5
  • 2 Method summary...5
  • 3 Reagents...5
  • 4 Instruments...6
  • 5 Analysis steps...6
  • 6 Precision...7
  • 7 Quality assurance and control...8

Foreword

GB/T 8647 "The methods for chemical analysis of nickel" is divided into the following

10 parts:

GB/T 8647.1, The methods for chemical analysis of nickel -- Determination of iron

content -- Sulfosalicylic acid spectrophotometric method

GB/T 8647.2, The methods for chemical analysis of nickel -- Determination of

aluminium content -- Electrothermal atomic absorption spectrometric method

GB/T 8647.3, The methods for chemical analysis of nickel -- Determination of silicon

content -- Molybdenum blue spectrophotometric method

GB/T 8647.4, The methods for chemical analysis of nickel -- Determination of

phosphorous content -- Molybdenum blue spectrophotometric method

GB/T 8647.5, The methods for chemical analysis of nickel -- Determination of

magnesium content -- Flame atomic absorption spectrometric method

GB/T 8647.6, The methods for chemical analysis of nickel -- Determination of cadmium,

cobalt, copper, manganese, lead and zinc contents -- Flame atomic absorption

spectrometric method

GB/T 8647.7, The methods for chemical analysis of nickel -- Determination of arsenic

antimony bismuth tin and lead contents -- Electrothermal atomic absorption

spectrometric method

GB/T 8647.8, The methods for chemical analysis of nickel -- Determination of sulphur

content -- Infra-red absorption method after high frequency induction furnace

combustion

GB/T 8647.9, The methods for chemical analysis of nickel -- Determination of carbon

content -- Infra-red absorption method after high frequency induction furnace

combustion

GB/T 8647.10, The methods for chemical analysis of nickel -- Determination of arsenic

cadmium lead zinc antimony bismuth tin cobalt copper manganese magnesium silicon

aluminium iron contents -- Atomic emission spectrometric method

This is Part 9.

This Part is not equivalent to ISO 7524:1985 "Nickel, ferronickel and nickel alloys --

Determination of carbon content -- Infra-red absorption method induction furnace

combustion".

The methods for chemical analysis of nickel - Determination

of carbon content -- Infra-red absorption method after high

frequency induction furnace combustion

1 Scope

This Part specifies the method for the determination of carbon content in nickel.

This Part is applicable to the determination of carbon content in nickel. Determination

range: 0.0020%~0.20%.

2 Method summary

In the presence of flux, oxygen flow is introduced into the high-frequency induction

furnace. The test material is burned at high temperature. Carbon generates carbon

dioxide gas and enters the infrared absorption cell. The instrument automatically

measures its absorption of infrared energy, calculates and displays the result.

3 Reagents

3.1 Anhydrous magnesium perchlorate [Mg(ClO4)2].

3.2 Caustic soda asbestos.

3.3 Glass wool.

3.4 Cotton wool.

3.5 Platinum-coated silicone.

3.6 Flux: low sulfur tungsten, tin, iron: W+Sn+Fe (1 g+0.2 g~0.25 g+0.2 g).

3.7 Crucible

Ceramic crucible (Phi24 mm×24 mm). Before use, it should be burned in a furnace with

air or oxygen at a temperature greater than 1100°C for 1 h~1.5 h. Take it out and place

it in a non-oiled desiccator to cool it for use (valid for two days).

3.8 Standard steel sample: carbon content is 0.002%~0.3%.

3.9 Standard steel sample or pure iron sample: carbon content is about 0.002%.

standard steel sample is stable within the error range.

NOTE: When selecting a standard steel sample, the carbon content should be greater than the

carbon content of the specimen being tested.

5.4 Calibration blanks

5.4.1 Weigh 1.000 g of low carbon standard steel sample (3.9) and place it in a crucible

(3.7). Add flux (3.6).

5.4.2 Place the crucible on the support of the furnace. Raise it to the burning position.

Follow the "automatic" blank calibration steps in the instrument manual. Repeat for 3

to 5 steel samples to obtain an average result with good reproducibility. The mass

fraction of carbon in the standard steel sample is deducted by "automatic" blank

correction. The value is expressed in %. The blank value obtained is stored in the

computer. (When the carbon content of the sample is > 0.001%, the blank value should

be < 0.0005%. When the carbon content of the sample is > 0.1%, 5.4 is omitted)

5.4.3 After the blank value is determined, repeat the test of the standard steel sample

according to (5.3). The test result should be stable within the error range. Then select a

standard sample with the same carbon content as the sample being tested for retest.

5.5 Determination

5.5.1 Weigh 1.000 g of specimen and place it in a crucible (3.7). Add flux (3.6).
5.5.2 Place the crucible on the support of the furnace. Raise it to the combustion position.

Follow the "automatic" analysis steps in the instrument manual. The instrument

automatically deducts the blank value and displays and prints the mass fraction of

carbon. The value is expressed in %.

6 Precision

6.1 Repeatability limit

For the determination values of two independent test results obtained under

repeatability conditions, the absolute difference between the two test results does not

exceed the repeatability limit (r) within the range of the average values given below.

The case where the repeatability limit (r) is exceeded does not exceed 5%. The intra-

laboratory repeatability is listed in Table 2.

......

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 8 pages — is available in the English PDF.

Referenced standards

How to Download GB/T 8647.9-2006

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
8 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 8647.9-2006

$109.00