GB/T 7962.7-2025Test methods of colourless optical glass — Part 7: Extent of striae (English PDF)
无色光学玻璃测试方法 第7部分:条纹度
Open the GB/T 7962.7-2025 preview as PDF
This is a limited preview
Buy now to download the full PDF (15 pages)
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
June 30, 2025
Implementation date
January 1, 2026
Scope
GB/T 7962.7-2025 is the English-translated version of 无色光学玻璃测试方法 第7部分:条纹度.
GB/T 7962.7-2025 is the Chinese national standard covering grading the streaks of differing refractive index inside optical glass — the striae meter with its collimated beam and its projection screen or image sensor, the darkened room, the three sample forms of a polished blank, a finely ground one wetted against a mounting plate, and a moulded preform immersed in refractive liquid because it has no flat face to look through, and the reading of the striae image. Part 7 of the series. It replaces GB/T 7962.7-1987, under the China Machinery Industry Federation. In force from 1 January 2026. Issued on 30 June 2025, it has been in force since 1 January 2026, replacing GB/T 7962.7-1987.
Document preview — GB/T 7962.7-2025
National Standard of the People's Republic of China
- ICS
- 81.040.01
- Classification
- L 50
- Replacing
- GB/T 7962.7-1987
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- Introduction5
- 1 Scope8
- 2 Normative References8
- 3 Terms and Definitions8
- 4 Test Principle8
- 5 Instruments9
- 6 Sample11
- 7 Test Conditions11
- 8 Test Procedures12
- 8.1 Preparation12
- 8.2 Test Operation12
- 9 Result Expression13
- 10 Test Report14
- Appendix A (informative) Measurement Method for Extent of Striae of Optical Elements - Interferometry15
1 Scope
This document describes the test principle, instruments, sample, test conditions, test procedures, result expression, and test report for the extent of striae of colorless optical glass.
This document applies to the testing of the extent of striae of colorless optical glass. It can be used as a reference for testing the extent of striae of other materials that transmit light in the test wave band.
2 Normative References
GB/T 903-2019
3 Terms and Definitions
The following terms and definitions are applicable to this document.
3.1 Striae
The non-uniformity of a material caused by the difference in refractive index between localized areas and surrounding areas within the glass.
3.2 Extent of striae
The degree, to which striae are present within the glass under test.
NOTE. it includes the thickness, length and density of the striae.
4 Test Principle
When striae are present in glass, there is a difference in refractive index between the striae and the glass matrix. When a beam of light passes through a striae-containing glass sample, the presence of the striae causes the light to be deflected, resulting in a striae image on the beam cross-section. This method uses a parallel beam of light passing through the glass under test.
Via optical imaging, the striae image is projected onto a projection screen for observation by the human eye or received by an image sensor. Against a dark background, the shape, size and distribution of the striae are tested to determine the extent of striae.
When the light-transmitting surface of the glass sample under test is a polished parallel plane, conventional testing is used; when the light-transmitting surface of the glass sample under test is a finely ground parallel plane, the light-transmitting surface is fitted with a refractive liquid using a mounting plate, and testing is performed using the mounting plate method; when the light-transmitting surface of the glass sample under test is a spherical, aspherical, or other non- parallel planes, for example, a molded pre-form, the sample is immersed in a refractive liquid using a sample cell, and testing is performed using the immersion method.
5 Instruments
The basic composition of a striae meter using the projection method is shown in Figure 1.
Depending on the mode of acquiring the striae image by the image receiving system, it is divided into two modes. human eye observation via projection screen reception and direct image acquisition via image sensor. The instrument mainly includes a light source system (light source, filter, aperture slot), a beam expansion and collimation system, an image receiving system, and a sample stage. For the mounting plate method and the immersion method, accessories such as mounting plates and sample cells are also included. Specific requirements are as follows.
a) Light source system. primarily composed of light source, filter and aperture slot, providing an illumination of 50 lx and a matched test wavelength for striae testing.
— Light source. adjustable intensity. When using a projection screen to receive striae images, the illuminance projected onto the projection screen shall be 50 lx. When using an image sensor to receive striae images, the intensity of illumination on the image sensor shall meet its operational requirements, forming a clear electronic image without overexposure or underexposure.
— Filters. when using a projection screen to receive striae images, a visible light filter suitable for human eye observation is used. When using an image sensor to receive striae images, a matching filter suitable for the image sensor is used to protect the sensor.
— Aperture slot. adjustable aperture, with an aperture of 0.5 mm to 4 mm, matched with a beam expansion and collimation system to adjust beam quality, ensuring a uniform and spotless beam illuminating the sample.
b) Beam expansion and collimation system. F-number (i.e., the reciprocal of the relative aperture) 10.
c) Image receiving system. depending on the mode of acquiring the striae image, either a projection screen or an image sensor can be used.
— The illuminated surface of the projection screen shall be a uniform clean white board coated with 250-mesh fine white gypsum powder, or an equivalent white board. The screen size shall not be smaller than the projection spot size.
— -The operating wavelength of the image sensor shall match the light source system, with a pixel count of not less than 1 K 1 K and a spatial resolution 0.05 mm.
d) Sample stage. the sample height and state can be adjusted in accordance with the sample specifications, and the angle can be adjusted left and right.
e) Mounting plate. made of materials, for example, glass with high transmittance, good chemical stability, and extent of striae meeting Grade A as specified in GB/T 903- 2019, with a thickness 2 mm, a polished working surface, a roughness Ra 0.05 m, and a parallelism 1.
f) Sample cell. made of materials, for example, glass with high transmittance, good chemical stability, and extent of striae meeting Grade A as specified in GB/T 903- 2019 The light-transmitting surface is polished, with a roughness Ra 0.05 m and a parallelism 1.
b) Direct image acquisition mode via image sensor Description of indexes. 1---light source; 2---filter; 3---aperture slot;
4---beam expansion and collimation system; 5---sample testing station; 6---sample stage;
7---sample testing using the immersion method (including sample cell and refractive liquid);
8---sample testing using conventional methods;
9---sample testing using the mounting plate method (including mounting plate and refractive liquid); 10---projection screen; 11---human eye observation; 12---image sensor and lens; 13---computer.
Figure 1 -- Schematic Diagram of Striae Meter Using Projection Method (continued)
6 Sample
The striae samples used for testing are divided into three categories, with the following requirements for each category.
a) Routine test samples. sample specifications are based on natural dimensions, with polished light-transmitting surfaces, roughness Ra 0.05 m, and parallelism 2.
For sampled striae glass, the sample length is 50 mm 0.5 mm.
b) Test samples for the mounting plate method. sample specifications are based on natural dimensions, with finely ground light-transmitting surfaces, roughness Ra 1 m, and thickness difference 0.02 mm.
c) Test samples for the immersion method. all surfaces of the samples must be clean.
7 Test Conditions
Testing shall be conducted in a well-ventilated dark room, but local illumination that does not affect human eye observation, projection screen brightness, and image sensor is permissible. ......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Editions of GB/T 7962.7
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 7962.7-2025 | Test methods of colourless optical glass - Part 7: Extent of striae | current edition | Current |
This page sells the current edition, GB/T 7962.7-2025. Earlier editions are listed for reference only.
How to Buy GB/T 7962.7-2025
- 1Add to cart. Click the "Buy GB/T 7962.7-2025" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.