Scope
GB/T 6730.76-2017 is the English-translated version of 铁矿石 钾、钠、钒、铜、锌、铅、铬、镍、钴含量的测定 电感耦合等离子体发射光谱法.
GB/T 6730.76-2017 is the Chinese standard "Iron ores - Determination of potassium, sodium, vanadium, copper, zinc, lead, chromium, nickel, cobalt elements - Inductively coupled plasma optical emission spectrometry".
Its scope clause reads: This Part of GB/T 6730 specifies the method for the determination of potassium, sodium, vanadium, copper, zinc, lead, chromium, nickel and cobalt content by inductively coupled plasma optical emission spectrometer (ICP-OES).
This Part is applicable to iron ore, iron concentrate, sinter and pellets. The range of measured element content is shown in Table 1. Its clauses include principle; reagents and materials; instruments and equipment; sampling and sample preparation; analysis steps; result calculation and its representation; test report.
It was issued by the General Administration of Quality Supervision, Inspection and Quarantine on OCTOBER 14, 2017 and took effect on JULY 01, 2018.
Found what you need?
Document preview — GB/T 6730.76-2017
National Standard of the People's Republic of China
- ICS
- 73.060.10
- Classification
- D 31
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Principle5
- 4 Reagents and materials5
- 5 Instruments and equipment6
- 6 Sampling and sample preparation6
- 7 Analysis steps7
- 8 Result calculation and its representation10
- 9 Test report15
- Annex A (normative) Acceptance procedure for specimen analysis results16
- Annex B (normative) Spectrometer performance test17
- Annex C (normative) Recommended calibration curves20
- Annex D (informative) Statistics and other data from interlaboratory test results21
1 Scope
This Part of GB/T 6730 specifies the method for the determination of potassium, sodium, vanadium, copper, zinc, lead, chromium, nickel and cobalt content by inductively coupled plasma optical emission spectrometer (ICP-OES).
This Part is applicable to iron ore, iron concentrate, sinter and pellets. The range of measured element content is shown in Table 1.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 602, Chemical reagent - Preparations of standard solutions for impurity
GB/T 6379.1, Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions
GB/T 6379.2, Measurement methods and results - Accuracy (trueness and precision) - Part 2: Determine the standard methods of measurement repeatability and reproducibility of the basic method
GB/T 6682, Water for analytical laboratory use - Specification and test methods
GB/T 6730.1, Iron ores - Preparation of pre-dried test samples for chemical analysis
GB/T 8170, Rules of rounding off for numerical values and expression and judgement of limiting values
GB/T 10322.1, Iron ores - Sampling and sample preparation procedures
GB/T 12806, Laboratory glassware - One-mark volumetric flasks
GB/T 12808, Laboratory glassware - One-mark pipettes
3 Principle
Use mixed acid of hydrochloric acid, hydrofluoric acid, nitric acid and perchloric acid to decompose the specimen. Use inductively coupled plasma spectrometer to select the appropriate spectral line to measure the spectral intensity of the element to be detected. Calculate the final element content according to the calibration made from the standard solution.
4 Reagents and materials
In the analysis, unless otherwise stated, only use the confirmed guaranteed reagents and water of grade two and above in accordance with GB/T 6682.
4.1 High-purity iron: purity >99.99%.
4.2 Mixed flux: 2 portions of anhydrous sodium carbonate and 1 portion of boric acid are ground and mixed, dried and set aside for future use.
4.3 Hydrochloric acid: rho is about 1.16g/mL
Thoroughly mix the laboratory specimens. Use indentation to sample. Dry the specimen at 105°C±2°C according to GB/T 6730.1.
7 Analysis steps
7.1 Number of determinations
In accordance with the procedures in Annex A, the same pre-dried specimen is independently determined at least twice. Take the average value.
NOTE: "Independently" means that the results of any subsequent and subsequent determinations are not affected by the results of previous determinations. In this analytical method, this condition means repeated measurements by the same operator at different times or by different operators, including the use of appropriate recalibration.
7.2 Amount of test material
Weigh about 1.0g of the pre-dried specimen (6.2), accurate to 0.0002g.
NOTE: The test material is quickly weighed to avoid re-absorbing moisture.
7.3 Blank test and verification test
7.3.1 Blank test
Conduct the blank test with the test material. Use high purity iron (see 4.1) instead of test material to test. Make sure it is consistent with the iron content in the material to be tested. All reagents shall be taken from the same reagent bottle.
7.3.2 Verification test
The same type of standard sample is analyzed along with the test material to do the verification test.
7.4 Measurement
7.4.1 Decomposition of test material
Place the test material (see 7.2) in a 100mL PTFE beaker (see 5.1). Use water to wet. Add 10mL of hydrochloric acid (see 4.3), 1mL of nitric acid (see 4.4), 4mL of hydrofluoric acid (see 4.5), 3mL of perchloric acid (see 4.6). Cover with PTFE cap. Heat on a hot plate. The temperature is controlled at around 150°C till it is dissolved completely. Open the lid and evaporate to dryness. Add 10mL of hydrochloric acid (see 4.7) and dissolve it on a discharge hot plate. After cooling, transfer to a 100mL volumetric flask (see 5.2). Use water to dilute to the scale. Mix well. necessary, perform as many operations as possible until the obtained value is better than the value in Table 2. When the element concentration in the test solution is higher than 5000 × DL, RSDN is the only performance parameter to evaluate. The measured value shall be lower than the RSDNmin recommended in Table 2.
7.4.3 Calibration curve
The calibration solution is defined as the solution required to draw the analytical element calibration curve. Its concentration range in the solution is expressed in mass fraction (%). It depends on the performance parameters and linear sensitivity of the instrument. A minimum of 6 calibration solutions are required to cover the ranges shown in Table 1. For specimens with a narrow concentration range, the calibration solution shall include the effective area. If the element concentration in the solution is higher than 5000 × DL, an additional calibration curve shall be drawn to cover this range. When the calibration curve is nonlinear, it can choose a spectral line with a lower sensitivity or dilute the sample and calibration solution equally.
Table C.1 in Annex C lists the calibration curve concentrations in general. It is diluted from stock solution (see 4.9). The matrix of the calibration solution is as consistent as possible with the matrix in the solution to be tested. Generally, calculate according to the iron content in the test material which is 60%. The concentration of hydrochloric acid in the solution to be tested is 5%. If it is not a ready-made calibration solution, check its strength before use to determine if it can be used.
7.4.4 Measurement
Determine the calibration solutions in order from low concentration to high concentration. Start the measurement from the blank calibration solution. Read the reading three times for each measurement. Take the average value. Wash with water between each two tests. When determining each element, the iron content in the blank solution shall not be greater than the value in Table C.2.
NOTE 1: If the test solution exceeds the calibration curve range, the test solution and blank can be diluted, or the calibration solution can be reconstituted.
NOTE 2: Select spectral lines to see if there is any background interference. The blank solution and sample solution with iron added can be sucked and sprayed to test.
NOTE 3: After the calibration curve is completed, two-point recalibration can be used to deal with instrument drift after a period of time. See 8.3 for details.
After determining the calibration solution, determine the standard substance solution first. Then determine the sample solution. Use water to wash between each two tests. Standard substance solution and sample solution shall be […]
Annex B (normative) Spectrometer performance test
B.1 General requirements for spectrometer performance
The purpose of the performance tests given in this appendix is to provide the means. By optimizing the ICP spectrometer, different types of instruments can be used and different operating conditions can be found. This enables the use of different plasma spectrometers to ultimately produce consistent results. The whole process controls three basic parameters: detection limit (DL), background equivalent concentration (BEC) and short-term precision (RSDNmin). The best performance of the instrument to determine the element to be tested is shown in Table 2.
B.2 Definitions
B.2.1 Detection limit (DL)
When the element produces a minimum concentration signal, any false background signal with a certain specified level can be considered to be exceeded. Additionally, the elemental concentration yields a signal that is three times the standard deviation of the background intensity.
B.2.2 Background equivalent concentration (BEC)
The background intensity that is equal to the net signal intensity produced by the analyte concentration. It is a measure of sensitivity to a given wavelength.
B.2.3 Short-term precision (RSDN): The relative standard deviation of a series of instrument intensity readings obtained under predetermined conditions.
B.3 Reference solution
Use the three calibration solutions in Table C.1: blank solution, solution 1 (within the lowest concentration range determined in Table 1), and solution 2 for detection.
B.4 Procedure
The plasma spectrometer shall first be adjusted according to the manufacturer's recommendations and the laboratory's quantitative analysis practices. Aspirate the blank and take 10 intensity readings. Repeat this for solution 1, solution 2.
Use formula (B.1) to calculate the slope of the analytical curve: […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 602Chemical reagent - Preparation of standard solutions for impurity determination
- GB/T 6379.1Accuracy (trueness and precision) of measurement methods and results -- Part 1: General principles and definitions
- GB/T 6379.2Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
- GB/T 6682Water for analytical laboratory use - Specification and test methods
- GB/T 6730.1Iron ores -- Preparation of pre-dried test samples for chemical analysis
- GB/T 8170Rules of rounding off for numerical values & expression and judgement of limiting values
Similar standards
How to Download GB/T 6730.76-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 10322.1-2000 — Iron ores. Sampling and sample preparation procedures
GB/T 10322.1-2014 — Iron ores - Sampling and sample preparation procedures
GB/T 10322.1-2023 — Iron ores - Sampling and sample preparation procedures
Secure payment via Stripe
Payments accepted
GB/T 6730.76-2017
$235.00