Scope

GB/T 47684-2026 is the English-translated version of 仿生液体门控技术 跨膜压强测试方法.

GB/T 47684-2026 is the Chinese standard "Bioinspired liquid gating technology - Test method for transmembrane pressure". Its scope clause reads: This document describes a method for testing transmembrane pressure in biomimetic liquid gating technology.

This document applies to the transmembrane pressure testing of thermodynamically stable biomimetic liquid-gated systems (hereinafter referred to as "liquid-gated systems").

Its clauses include principles; testing apparatus; test conditions; preparation of the liquid valve system; system stability assessment; test steps; experimental data processing; test report.

Found what you need?

Document preview — GB/T 47684-2026

National Standard of the People's Republic of China

ICS
07.080
Classification
A 53

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • 1.Scope
  • 4.Principles
  • 5.Testing apparatus
  • 6 Test Conditions
  • 7.Preparation of the liquid valve system
  • 7.1 Porous Solid Membrane Materials

1 Scope

This document describes a method for testing transmembrane pressure in biomimetic liquid gating technology.

This document applies to the transmembrane pressure testing of thermodynamically stable biomimetic liquid-gated systems (hereinafter referred to as "liquid-gated systems").

4 Principles

By introducing functionally gated liquids into a solid porous framework, and controlling the opening and closing of the gates by applying pressure, controllable regulation of mass transport can be achieved.

Membrane transport behavior is characterized by the liquid-solid interface between a functionally gated liquid material and a porous solid membrane material, as well as the interaction between the transport fluid and the functionally gated liquid material.

The synergistic effect of the liquid (or gas)-liquid interface between materials.

As shown in Figure 1, the relative relationship between the transmembrane pressure and the critical transmembrane pressure determines the operating state of the liquid gate system.

---When the transmembrane pressure is lower than the critical transmembrane pressure (Pa)

The channel prevents the transported fluid from passing through the membrane;

---When the transmembrane pressure is higher than the critical transmembrane pressure (Pa >Pc), the transport fluid can pass through the membrane for transmembrane transport, and pores will form

Forming a stable gas-liquid or liquid-liquid interface layer;

---When the transmembrane pressure decreases below the critical transmembrane pressure, the functionally gated liquid material will refill the pores of the porous solid membrane, restoring...

The membrane is in the closed state.

5 Testing apparatus

Figure 2 shows a schematic diagram of the transmembrane pressure testing device. The functions and requirements of each component are as follows.

---Digital source meter. Provides a stable DC voltage to the pressure sensor, synchronizing the pressure sensor signal to the display unit, maximum voltage

It should be.200V, the maximum current should be 1A, the maximum measurement resolution should be 1pA or 100nV, and the basic measurement accuracy should be less than or equal to 100nV.

If the value is equal to 0.012%, the minimum sampling frequency should be 1Hz;

---Pressure sensor. Measures pressure in pipelines by outputting a current signal. The measurement range depends on the testing requirements, and the accuracy class should be within 0.5.

within;

---Temperature sensor. Tests the experimental temperature; the measuring range depends on the testing requirements; the measurement error should be less than ±(0.15°C 0.002×).

(range)

---Display unit. Operates the test software normally, acquiring and recording current data from the pressure sensor and temperature sensor data;

---Sample under test. Stable composite liquid-gated membrane, specific requirements are detailed in 7.1 and 7.2;

---Membrane clamping system. This system secures the gated membrane containing the measured liquid. It consists of a clamping plate and a sealing gasket, with a central opening for fluid passage. (Details omitted)

Please follow Appendix A;

---Three-way valve. Used to simultaneously connect sensor tubing, injection unit tubing, and membrane clamping system tubing;

---Fluid Piping. Provides a pathway for fluid transport and should be corrosion-resistant, pressure-resistant, and not easily deformed. Unless otherwise specified, fluid piping...

The total volume should be 10 mL;

---Pressure supply system. A stepper injection pump equipped with a syringe of appropriate capacity pressurizes the tubing, driving the fluid through a liquid-gated membrane, such as...

Unless otherwise specified, a 50mL syringe should be used, providing a flow rate range of 0mL/min to 100mL/min.

Note. For ease of representation, the SI units of flow rate (m3/s) have been converted, 1m3/s = 6 × 107 mL/min.

6 Test Conditions

Unless otherwise specified, the testing environment should be.

---Pressure. Complies with the standard atmospheric pressure of GB/T 45846;

---Temperature. 25°C±5°C;

---Relative humidity. (50±10)%.

7 Preparation of the liquid valve system

7.1 Porous Solid Membrane Materials

The preparation process for porous solid membrane materials is as follows.

a) Select.

1) Thickness. Should be a flat film of 10µm~1000µm;

2) Material. It should be an inorganic material, an organic material, or a hybrid material. Commonly used membrane materials are listed in Table 1.

3) Aperture. should be 0.1µm~100µm;

4) Dimensions. should be consistent with the dimensions of the sealing elastomer, the contact surface should be flat and wrinkle-free, and the sample should be concentric with the sealing elastomer.

b) Cleaning.

1) The cleaning solution should be distilled water that meets or exceeds the laboratory grade III standard as required by GB/T 6682, with a purity of 95%.

Organic reagents such as methanol, ethanol, acetone, n-hexane, petroleum ether, or mixtures thereof, and cleaning solutions.

It should not chemically react with porous solid membrane materials and should not cause the porous solid membrane materials to swell;

2) Immerse the porous solid membrane material to be tested in the cleaning solution and clean the surface of the material using ultrasonic cleaning technology;

3) Wrap the cleaned porous solid membrane material in a clean, soft, lint-free cloth and let it air dry naturally in a clean bench, or use a clean, soft cloth to dry it.

Dry air or inert gas is used to gently blow away any residual cleaning fluid from the surface and pores.

7.2 Functionally Gated Liquid Materials

Based on the requirements of the experimental project, select or prepare a porous solid membrane material that exhibits flow characteristics under the test conditions and can fully wet and stabilize the membrane.

The functional gated liquid material is used to fill the internal cavity. Commonly used functional gated liquid material categories are shown in Table 2.

7.3 Transfer Fluid

The transport fluid should be immiscible with the functionally gated liquid material, and the affinity between the transport fluid and the porous solid membrane material should be lower than that of the functionally gated liquid material.

The affinity between gated liquids and porous solid membrane materials, and the types of transported fluids are as follows.

---Gas. Types include. air, nitrogen, argon, oxygen, carbon dioxide, sulfur dioxide, hydrogen sulfide, ammonia, formaldehyde, and mixtures thereof.

According to the test requirements, obtain the gas and fill it into a syringe. Harmful gases should be tested in a well-ventilated environment.

Released in accordance with GB 16297.

---Liquid. Commonly used fluid transfer materials are listed in Table 2.Obtain the liquid according to the experimental requirements and fill it into the syringe.

8 System stability assessment

8.1 Calculation

Before conducting the transmembrane pressure test, the liquid gate system should be calculated and judged to determine whether it is in a thermodynamically stable state according to formulas (1) and (2).

8.2 Judgment

Based on the principle of minimum energy, the thermodynamic stability of the liquid-valve system is determined as follows.

---When deltaE1<=0 or deltaE2<=0, it is a thermodynamically unstable state, and the functionally gated liquid material or porous solid membrane should be replaced.

Material;

---When deltaE1 >0 and deltaE2 >0, it is a thermodynamically stable state, and transmembrane pressure testing can be performed.

9 Test Steps

Follow these steps to perform the test.

a) Preparation of liquid-gated membranes.

1) Completely immerse the porous solid membrane material in the functionally gated liquid material for at least 2 minutes;

2) Remove the liquid-gated membrane and remove excess functionally gated liquid material from the surface of the porous solid membrane until there is no obvious liquid level on the surface.

A protrusion.

b) Prepare the instruments and equipment and connect the pipelines according to the requirements of Chapter 5.

c) According to the device diagram of the clamping system in Appendix A, install the liquid gated membrane into the clamp and connect the membrane clamping system to the pipeline.

d) Connect the power cord, turn on the device switch, start the test software in the display unit interface, and set the test parameters (sampling interval, test...).

The sampling interval should be set to 1 second for parameters such as time and pressure supply system operating parameters (flow rate, etc.).

e) As the test progresses, the system is gradually pressurized, and transmembrane pressure tests are performed within the accuracy class of the pressure sensor until the current-time curve is obtained.

The line reaches its peak value and stabilizes [stable state means that the fluctuation range of the current does not exceed ±(accuracy class % × output current range)].

Example. A pressure sensor with an output current of 4mA~20mA, a measuring range of 0kPa~100kPa, and an accuracy class of 0.5, exhibiting a current fluctuation range.

=±(0.5%×16mA)=±0.08mA.

f) Repeat steps d) and e) three times.

g) Save and record the test data.

h) Turn off the instrument and equipment and disconnect the connecting pipelines.

10 Experimental Data Processing

Convert the test current value into a pressure value according to formula (3).

11 Test Report

The test report should include the following content; see Appendix D for a sample test report.

a) The name and number of the standard on which the test was based (including the year of publication);

b) Report number, submitter, submitting organization, submission date, and project name;

c) Porous solid membrane materials, functionally gated liquid materials, and fluid information transmission;

d) Test temperature, humidity, atmospheric pressure, and flow rate;

e) Pressure sensor range;

f) Transmembrane pressure curve;

g) Critical transmembrane pressure test results, including the measured values, average values, and standard deviations of all tests;

h) Any anomalies observed;

i) Test date;

j) Information about the tester, reviewer, approver, and laboratory.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 30 pages — is available in the English PDF.

How to Download GB/T 47684-2026

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
30 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 47684-2026

$240.00