GB/T 4325.13-2013Methods for chemical analysis of molybdenum - Part 13: Determination of calcium content - Flame atomic absorption spectrometry (English PDF)

钼化学分析方法 第13部分:钙量的测定 火焰原子吸收光谱法

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for GB/T 4325.13-2013 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

May 9, 2013

Implementation date

February 1, 2014

Scope

GB/T 4325.13-2013 is the English-translated version of 钼化学分析方法 第13部分:钙量的测定 火焰原子吸收光谱法.

GB/T 4325.13-2013 is the Chinese standard "Methods for chemical analysis of molybdenum - Part 13: Determination of calcium content - Flame atomic absorption spectrometry".

Its scope clause reads: GB/T 4325 This section specifies the method for determination of molybdenum of calcium. This section applies to molybdenum powder, molybdenum, molybdenum trioxide, ammonium molybdate determination of calcium.

Measurement range. 0.0005% to 0.030%. Its clauses include method summary; reagents; instrument; sample.

Found what you need?

Document preview — GB/T 4325.13-2013

National Standard of the People's Republic of China

ICS
77.120.99
Classification
H 63
Replacing
GB/T 4325.14-1984

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • 1 Scope
  • 2 Method summary
  • 3 Reagents
  • 4 Instrument
  • 5 Sample

1 Scope

GB/T 4325 This section specifies the method for determination of molybdenum of calcium.

This section applies to molybdenum powder, molybdenum, molybdenum trioxide, ammonium molybdate determination of calcium. Measurement range. 0.0005% to 0.030%.

2 Method summary

Sample decomposition of hydrogen peroxide, citric acid, hydrochloric acid medium, using standard addition method using air - acetylene flame atomic absorption of light

Spectrometer to measure calcium absorbance at the wavelength of 422.7nm, and calculate the amount of calcium.

3 Reagents

Unless otherwise indicated, this part of the reagents used are in line with national standards or industry standard analytical reagent, the water is deionized water.

3.1

paramolybdate (w>=99.99%).

3.2

Hydrogen peroxide (30%), pure class distinctions.

3.3

hydrochloric acid (rho = 1.19g/mL), pure class distinctions.

3.4

citric acid (500g/L), pure class distinctions.

3.5 Calcium standard stock solution

Weigh 2.4970 grams of calcium carbonate (mass fraction >=99.99%) (pre-bake at 105 °C 1h, and in the dryer

Cooled to room temperature) in the 250mL beaker, add 30mL water, 15mL hydrochloric acid (3.3), on the home heating furnace until completely dissolved, boil rid

Carbon dioxide, after cooling transferred 1000mL volumetric flask, dilute with water to volume, and mix. This solution 1mL containing 1.0mg of calcium.

3.6 Calcium standard solution

Pipette 5.0mL calcium standard stock solution (3.5), placed in 500mL volumetric flask, dilute with water to volume, and mix.

This solution 1mL containing 10µg calcium.

4 Instrument

Atomic absorption spectrometer, with a calcium hollow cathode lamp.

Under optimum working conditions for those who can reach the following indicators can be used.

--- Characteristic concentration. measurement solution with the matrix consistent solution, characterized in calcium concentration should not exceed 0.05µg/mL;

--- Precision. standard solution with the highest concentration of 10 times the absorbance measurements, the standard deviation should not exceed 1.5% of the average absorbance;

With the lowest concentration of standard solution (not "zero" concentration of the standard) measuring the absorbance of 10 times, the standard deviation should not exceed the maximum

The average concentration of 0.5% absorbance;

--- Linear curve. the curve according to concentration, divided into five sections, the difference between the absorbance of the absorbance difference between the highest and the lowest segment segment

Ratio should not be less than 0.7.

5 Sample

Molybdenum should smash through the 0.75mm standard sieve.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Similar standards

Editions of GB/T 4325.13

EditionTitleRevisionStatus
GB/T 4325.13-2013Methods for chemical analysis of molybdenum - Part 13: Determination of calcium content - Flame atomic absorption spectrometrycurrent editionCurrent
GB/T 4325.14-1984Methods for chemical analysis of molybdenum - Part 13: Determination of calcium content - Flame atomic absorption spectrometryprevious editionSuperseded

This page sells the current edition, GB/T 4325.13-2013. Earlier editions are listed for reference only.

How to Download GB/T 4325.13-2013

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 4325.13-2013

$109.00