GB/T 42414-2023Glass viscosity measurement - Rotational viscometer method (English PDF)
玻璃黏度测定 旋转黏度计法
Open the GB/T 42414-2023 preview as PDF
This is a limited preview
Buy now to download the full PDF (15 pages)
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
March 17, 2023
Implementation date
October 1, 2023
Scope
GB/T 42414-2023 is the English-translated version of 玻璃黏度测定 旋转黏度计法.
GB/T 42414-2023 describes how the viscosity of molten glass is measured with a rotational viscometer, for melts that behave as Newtonian liquids. The principle clause explains what the instrument actually senses: a rotor turning continuously in the melt drags against the glass at its surface, and the friction that resists it is read as torque. From there the document sets out the apparatus, how the sample is prepared, the test procedure, and the data processing that turns instrument parameters, rotor speed and torque at each temperature into a viscosity value through the given formula. A test report clause fixes what has to be recorded alongside the number, and an informative appendix lists the technical differences from ISO 7884-2:1987 and the reasons for them. Viscosity against temperature is the curve a glass plant runs on: melting, refining, conditioning and forming are each set at a point on it, and a curve measured on a poorly calibrated instrument, or on a melt that never reached thermal equilibrium, throws the whole schedule off and shows up later as cord, bubbles or a forming line that will not hold shape. For glass makers, furnace and forming engineers, and the laboratories that qualify new compositions.
Document preview — GB/T 42414-2023
National Standard of the People's Republic of China
- ICS
- 81.040.01
- Classification
- N64
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Terms and Definitions4
- 4 Principles5
- 5 Apparatuses6
- 6 Sample8
- 7 Test Procedures8
- 8 Data Processing10
- 9 Test Report11
- Appendix B (Informative) Technical Differences and Causes between this Document and ISO 7884-2:198713
Foreword
This document was issued on 17 March 2023 by the State Administration for Market Regulation; Standardization Administration of the PRC and takes effect on 1 October 2023.
It is a GB/T standard: recommended rather than compulsory, but it is the text a Chinese reviewer applies when assessing a submission.
1 Scope
GB/T 42414-2023 describes how the viscosity of molten glass is measured with a rotational viscometer, for melts that behave as Newtonian liquids. The principle clause explains what the instrument actually senses: a rotor turning continuously in the melt drags against the glass at its surface, and the friction that resists it is read as torque. From there the document sets out the apparatus, how the sample is prepared, the test procedure, and the data processing that turns instrument parameters, rotor speed and torque at each temperature into a viscosity value through the given formula. A test report clause fixes what has to be recorded alongside the number, and an informative appendix lists the technical differences from ISO 7884-2:1987 and the reasons for them. Viscosity against temperature is the curve a glass plant runs on: melting, refining, conditioning and forming are each set at a point on it, and a curve measured on a poorly calibrated instrument, or on a melt that never reached thermal equilibrium, throws the whole schedule off and shows up later as cord, bubbles or a forming line that will not hold shape. For glass makers, furnace and forming engineers, and the laboratories that qualify new compositions.
This Document describes the rotational viscometer test method for the glass viscosity
measurement.
This Document is applicable to the measurement of Newtonian liquid glass melt viscosity in
the range of 1Pa.s~104 Pa.s.
2 Normative References
The provisions in following documents become the essential provisions of this Document
through reference in this Document. For the dated documents, only the versions with the dates
indicated are applicable to this Document; for the undated documents, only the latest version
(including all the amendments) is applicable to this Document.
GB/T 16839.1 Thermocouples - Part 1: EMF specifications and tolerances (GB/T
16839.1-2018, IEC 60584-1:2013, IDT)
3 Terms and Definitions
For the purposes of this Document, the following terms and definitions apply.
3.1 Newtonian liquid
A liquid that obeys Newton's law of viscosity.
NOTE: The viscosity of this type of fluid has nothing to do with shear rate. The viscosity is constant at
a certain temperature.
3.2 Field of flow
The space filled by the glass melt and its boundaries with the inner surface of the crucible, the
rotating shaft and the outer surface of the rotor.
3.3 Special instrument constant; k*
Instrument constants integrated by taking into account the flow field coefficient, dimensional
coefficient, viscometer head coefficient, and rotor viscosity coefficient.
NOTE: k* has nothing to do with torque, speed and viscosity.
3.4 Melting temperature
The temperature corresponds to the viscosity of 10Pa.s.
3.5 Working temperature
The temperature corresponds to the viscosity of 103Pa.s.
3.6 VFT-equation
An empirical formula that describes the relationship between the viscosity of glass and
temperature.
NOTE: VFT is the abbreviation of Vogel-Fulcher-Tammann.
4 Principles
When a rotational viscometer is used to measure the high-temperature viscosity of glass melt,
the motor drives the rotor to continuously rotate in the glass melt, and friction force is generated
between the rotor surface and the contacting glass melt interface. The friction force is related
to the viscosity of the glass melt. Therefore, changes in the rotation speed and torque of the
rotor reflect the viscosity of the glass melt. The greater the torque is, the greater the viscosity
of the glass melt is. The higher the rotation speed is, the smaller the viscosity of the glass melt
is. vice versa. The torque and rotational speed of the rotor in the glass melt are measured
experimentally, and the viscosity of the glass melt is calculated by using Formula (1).
Where:
eta - viscosity, in Pa.s;
k* - constant of special instrument;
tau - torque applied to the rotator, in N.m;
n - speed of rotator.
The high-temperature viscosity of the glass melt is mainly related to the composition and
temperature of the glass. The viscosity plot of the glass melt with a specific composition
measured at different temperatures constitutes the viscosity-temperature curve of the glass.
by the reference glass and the same viscosity value in Appendix C is greater than 10 degrees C, the
instrument status shall be adjusted to meet the usage requirements.
7.2 Testing
7.2.1 Place the crucible containing the glass sample to be measured into the high-temperature
furnace of the viscometer. Make sure that the crucible is placed stably on the base.
7.2.2 Set the test condition parameters, including the heating rate of the high-temperature
furnace (recommended 5 degrees C/min~10 degrees C/min), upper limit temperature, lower limit temperature,
holding time, cooling rate during the test (recommended 1 degrees C/min~3 degrees C/min), reheating rate after
testing. When setting the upper limit temperature, in addition to considering the possible
chemical reactions between the glass melt and the crucible and rotor materials, it is also
necessary to consider the volatilization and devitrification tendencies of the glass.
For volatile glass, the upper limit temperature should not be too high, as long as the bubbles
can be fully discharged (refer to the temperature corresponding to 100Pa.s of the glass melt).
For easily devitrified glass, it should use a faster cooling rate during measurement. The
temperature change rate should refer to the data set when calibrating the instrument using
standard glass. When setting parameters, measurements should be made within a short period
of time to avoid changes in the chemical or physical properties of the glass during long-term
measurements. The viscosity measurement can be carried out under uniform cooling conditions;
or the insulation setting can be performed at a specific temperature during the cooling process,
and the viscosity can be measured continuously.
7.2.3 Start heating and temperature-rising the sample to the set upper limit temperature, and
keep it warm for no less than 10 min.
7.2.4 Set the initial speed of the viscometer.
7.2.5 Slowly lower the rotor and start the rotor rotation. Once the bottom of the rotor contacts
the molten glass, the torque on the viscometer head starts to rise from 0, indicating that the rotor
has contacted the surface of the glass melt. Then lower the rotor to ensure that the rotor is
completely immersed in the glass melt and ensure that the location where the rotor is immersed
into the glass melt is consistent during each test.
7.2.6 The measurement officially starts, and the rotor speed and torque are continuously read
during the cooling process.
7.2.7 Stop the rotor rotation after the furnace temperature drops to the set lower limit
temperature according to the set testing procedure or when the measured viscosity reaches the
maximum set value. However, if it is found that the rotor speed decreases sharply, the rotor
rotation can be stopped when the furnace temperature is still higher than the set lower limit
temperature to avoid deformation of the rotating shaft due to excessive viscosity of the glass
melt.
7.2.8 When the furnace temperature quickly returns to the set value, lift the rotor to completely
leave the surface of the glass melt. After the residual glass melt drips from the rotor, lower the
crucible base, take out the crucible and pour out the glass melt.
In the event of a sudden failure of the instrument or phase separation or crystallization of the
glass, the rotor rotation shall be stopped manually; and the heating program shall be reset
immediately to increase the glass melt temperature and facilitate the lifting of the rotor and
meter head.
7.2.9 After lifting the crucible base and resetting it, stop the instrument heating system.
7.2.10 Shut down the instrument or after continuing to collect sample temperature and furnace
temperature to shut down the instrument.
7.3 End of test
7.3.1 After the test sample is cooled to room temperature, its appearance non-uniformity (such
as devitrification, traces of dissolution of crucible and rotor materials, bubbles, discoloration)
shall be observed. All unusual changes in materials in contact with the glass melt shall be
considered and recorded.
7.3.2 Soak and clean the cooled platinum rotor with hydrofluoric acid. If a platinum crucible is
used, the residual glass frit can be removed by pickling.
Remaining clauses in the full document
- 5 Apparatuses
- 6 Sample
- 7 Test Procedures
- 8 Data Processing
- 9 Test Report
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
GB/T 16839.1 Thermocouples - Part 1: EMF specifications and tolerances (GB/T
Similar standards
GB 38031-2025|GB/T 16839.1|GB/T 16839.1-2018|GB/T 42387|GB/T 42413|GB/T 35598|GB/T 35595|GB 38031-2020
How to Buy GB/T 42414-2023
- 1Add to cart. Click the "Buy GB/T 42414-2023" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards