GB/T 40267-2021Determination of Levodopa in plant derived products—High performance liquid chromatography (English PDF)
植物源产品中左旋多巴的测定 高效液相色谱法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 40267-2021 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
May 21, 2021
Implementation date
December 1, 2021
Scope
GB/T 40267-2021 is the English-translated version of 植物源产品中左旋多巴的测定 高效液相色谱法.
GB/T 40267-2021 sets out a high performance liquid chromatography method for measuring the levodopa content of plant derived products. It applies to dry plant samples such as Vicia faba L., Stizolobium cochinchinensis (Lour.)
Tang et Wang and Mucuna sempervirens Hemsl. The levodopa is extracted ultrasonically with a 2% hydrochloric acid solution, separated on a reversed-phase C18 column and detected with an ultraviolet or diode array detector at 280 nm, then identified by retention time and quantified by the external standard method.
The standard covers the reagents and materials, the instruments, sample preparation by grinding and sieving, specimen processing, the chromatographic reference conditions, the standard working curve, parallel tests, data processing and precision.
Laboratories testing plant raw materials use it to obtain comparable levodopa results. It was issued on 21 May 2021 and has applied since 1 December 2021.
Found what you need?
Document preview — GB/T 40267-2021
National Standard of the People's Republic of China
- ICS
- 07.080
- Classification
- A 40
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Terms and Definitions4
- 4 Principle4
- 5 Reagents and Materials4
- 6 Instruments and Equipment6
- 7 Sample6
- 8 Test Procedures6
- 9 Test Data Processing7
- 10 Precision8
- 11 Others8
- Appendix A (informative) High Performance Liquid Chromatogram9
1 Scope
This Standard describes the determination of levodopa content in plant derived products through the high performance liquid chromatography.
This Standard is applicable to the determination of levodopa content in plant derived dry samples, such as: Vicia faba L., Stizolobium cochinchinensis (Lour.) Tang et Wang and Mucuna sempervirens Hemsl., etc.
2 Normative References
The content of the following documents constitutes indispensable clauses of this document through normative references in the text. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB/T 6682 Water for Analytical Laboratory Use - Specification and Test Methods
3 Terms and Definitions
There are no terms and definitions that need to be defined in this document.
4 Principle
After the specimen is ultrasonically extracted with 2% (volume fraction) hydrochloric acid solution, it is separated through reversed-phase chromatographic column and detected by ultraviolet (or diode array) detector. It is qualitatively determined by retention time and quantitatively determined by the external standard method.
5 Reagents and Materials
5.1 Reagents
Unless it is otherwise specified, the reagents used by this Method shall be analytically pure. The water shall be Grade-1 water specified in GB/T 6682.
Microporous membrane (aqueous phase): with a pore size of 0.45 µm.
6 Instruments and Equipment
6.1 High performance liquid chromatograph: equipped with ultraviolet detector (UV) or diode array detector (DAD).
6.2 Analytical balance: with a division value of 0.01 mg and 0.0001 g.
6.3 Ultrasonic generator: with a power of 250 W and a working frequency of 40 kHz.
6.4 Centrifuge: with a rotating speed of not less than 8,000 g.
6.5 Acidity meter: with an accuracy of 0.01 pH unit.
6.6 Vortex mixer.
6.7 Electric pulverizer.
7 Sample
Take about 250 g samples of Vicia faba L., Stizolobium cochinchinensis (Lour.) Tang et Wang and Mucuna sempervirens Hemsl.; use the electric pulverizer to smash them; let them pass through a sieve with an inner diameter of 355 µm ± 13 µm. Mix the prepared specimen and equally divide it into two parts. Put it into a clean specimen container; seal and label it. Store it at room temperature. Keep it sealed and protected from light.
8 Test Procedures
8.1 Specimen Processing
Weigh-take 1 g of specimen (accurate to 0.001 g) into a 50 mL centrifuge tube with stopper. Add 10 mL of 2% (volume fraction) hydrochloric acid solution. After vortex mixing, ultrasonically extract it for 30 min; every 15 min, shake it once. Then, at 8,000 (x g), centrifuge for 5 min; transfer the supernatant to a 25 mL volumetric flask. Use 10 mL of 2% (volume fraction) hydrochloric acid solution to re-extract the residue.
Combine the two extracts. Then, use 2% (volume fraction) hydrochloric acid solution to reach a constant volume; shake it well. Let it pass through a 0.45 µm microporous membrane; reserve it for later test.
8.2 Determination
8.2.1 Liquid chromatography reference conditions
The liquid chromatography reference conditions are shown below:
a) Chromatographic column: C18 chromatographic column, 4.6 mm x 250 mm, particle size 5 µm, or others with equivalent performance;
b) Mobile phase: methanol-10 mmol/L dipotassium hydrogen phosphate solution = 5:95;
c) Flow rate: 0.7 mL/min;
d) Column temperature: 20 °C;
e) Injection volume: 10 µL;
f) Detection wavelength: 280 nm.
8.2.2 Drawing of standard working curve
In accordance with the levodopa content in the specimen, select the standard working solution with an appropriate range of mass concentration. From low concentration to high concentration, successively inject specimen for analysis. Take the chromatographic peak area of levodopa as the y-coordinate; take the mass concentration of the solution as the x-coordinate; draw the standard working curve.
The chromatogram of the standard working solution is shown in Figure A.1 of Appendix A.
8.2.3 Sample determination
Inject the specimen solution into the high performance liquid chromatograph. Conduct qualitative determination by the retention time; record the peak area. In accordance with the standard working curve, obtain the mass concentration of levodopa in the specimen solution. The mass concentration of levodopa in the specimen solution shall be within the range of mass concentration of the standard working curve. Samples that exceed the upper limit of the mass concentration of the standard working curve shall be diluted and injected for analysis. The chromatogram of a typical specimen solution is shown in Figure A.2.
8.3 Parallel Test
In accordance with the stipulations of 8.1 and 8.2, conduct two parallel determinations on the same specimen.
9 Test Data Processing
The levodopa content in the specimen shall be calculated in accordance with Formula (1): […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
How to Download GB/T 40267-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
GB/T 10003-2026 — Biaxially oriented polypropylene (BOPP) film
GB/T 10004-2026 — Plastic/Plastic laminated films and pouches for packaging
Secure payment via Stripe
Payments accepted
GB/T 40267-2021
$165.00