GB/T 39948-2021Method of heat distribution test for food thermal sterilization equipments (English PDF)

食品热力杀菌设备热分布测试规程

Preview PDF

This is a limited preview

Buy now to download the full PDF (30 pages)

Need quotation for your management?

Get a formal quotation for GB/T 39948-2021 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

March 9, 2021

Implementation date

April 1, 2022

Scope

GB/T 39948-2021 is the English-translated version of 食品热力杀菌设备热分布测试规程.

GB/T 39948-2021 sets out how to carry out a heat distribution test on thermal sterilization equipment for food. It covers the test principle, the instruments, the preparation before the test, the records, the data analysis, the test report and the analysis of anomalies, and recommends how many tests to run and how long results remain valid.

It applies to high-pressure static, rotary, basketless and hydrostatic equipment, continuous atmospheric-pressure water-bath and tunnel equipment, steam boxes and water-bath tanks heated by steam, hot water, steam-air-water or steam-air mixtures, but not to microwave or flame sterilizers.

Preparation includes calibrating the temperature data acquisition instruments under JJF 1366, placing the temperature probes, testing with the smallest container size and collecting at least 25 readings per probe in the constant-temperature stage. It was issued on 9 March 2021 and has applied since 1 April 2022.

Found what you need?

Document preview — GB/T 39948-2021

National Standard of the People's Republic of China

ICS
67.050
Classification
X 70

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • Introduction4
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions6
  • 4 Principle of heat distribution test6
  • 5 Heat distribution test instruments6
  • 6 Preparation before heat distribution test7
  • 7 Heat distribution test records19
  • 8 Heat distribution data analysis19
  • 9 Test report21
  • 10 Anomaly analysis22
  • 11 Recommendations on the number of heat distribution tests and validity period for food thermal sterilization equipment23
  • Appendix A (Informative) Questionnaire of sterilization equipment and ancillary facilities24
  • Appendix B (Informative) Heat distribution test data analysis table of thermal sterilization equipment (example)26
  • Appendix C (Informative) Number of heat distribution tests for food thermal sterilization equipment29
  • Appendix D (Informative) Validity period of heat distribution test results30

1 Scope

This Standard specifies the terms and definitions, principle of heat distribution test, heat distribution test instruments, preparation before heat distribution test, heat distribution test records, heat distribution data analysis, test report, and anomaly analysis requirements of heat distribution test for food thermal sterilization equipment; and gives recommendations on the number of food heat penetration tests and validity period.

This Standard applies to the heat distribution test for thermal sterilization equipment, such as high-pressure static type (vertical, horizontal, including spray type, full water type), rotary type, high-pressure basketless type, hydrostatic pressure type, atmospheric-pressure water-bath continuous type, atmospheric-pressure tunnel continuous type, atmospheric-pressure steam box, and water-bath sterilization tank (pool), which use steam, hot water, steam-air-water, or steam-air mixture as heat-transfer medium.

This Standard does not apply to the heat distribution test of microwave heating and flame heating sterilization equipment.

2 Normative references

The following documents are indispensable for the application of this document.

For the dated references, only the editions with the dates indicated are applicable to this document. For the undated references, the latest edition (including all the amendments) are applicable to this document.

GB 8950 Hygienic specifications of cannery

GB/T 20938 Good manufacturing practice for canned food enterprises

JJF 1366 Calibration Specification of Temperature Data Acquisition Instruments

QB/T 5218 Terms of canned food industries

5 Heat distribution test instruments

5.2 Temperature display device of sterilization equipment

It shall meet the requirements of GB 8950.

5.3 Automatic temperature recorder of sterilization equipment

It shall meet the requirements of GB 8950.

5.4 Pressure gauge of sterilization equipment

It shall meet the requirements of GB 8950.

5.5 Pressure recorder of sterilization equipment

It shall meet the requirements of GB/T 20938.

6 Preparation before heat distribution test

6.1 Requirements for thermal sterilization equipment and ancillary facilities

6.1.1 The thermal sterilization equipment and sterilization ancillary facilities to be tested shall meet the requirements of GB/T 20938; including the requirements of devices such as exhaust pipe diameter, exhaust valve type, steam pipe diameter, electronic thermometer, air release valve, circulating pump. Before the test, it shall check and ensure that the supply of water, electricity, steam, compressed air, etc. is in a normal state. In the heat distribution test, the ratio OF the opening area of the loading bottom plate and packaging container separation plate used TO the entire plate area shall meet the following requirements: No less than 20% for steam sterilization; no less than 30% for hot water sterilization.

6.1.2 Before use, the temperature data acquisition instrument shall be calibrated. The calibration method shall meet the requirements of JJF 1366:

That is, in a standard constant-temperature oil bath or other constant-temperature facilities, it is checked against a mercury thermometer or electronic thermometer of known accuracy (division value needs to be 0.1 °C or better).

Generally, a fixed-point temperature calibration method is used. For example, fixed-point calibration is carried out on the commonly-used sterilization temperatures 100 °C, 116 °C, 121 °C or other specific sterilization temperature points. The deviation value BETWEEN the recorded value of the temperature measurement data recorder AND the mercury thermometer or electronic thermometer of known accuracy shall be recorded and archived; and corrected during data processing. […]

6.3.3.3 Atmospheric-pressure continuous sterilization equipment

According to the width of the conveyor chain, every 200 mm~250 mm, simultaneously put all the temperature measurement probes into the entrance end of the input chain; or put them in groups and staggered at the same time.

6.4 Specification of packaging container for heat distribution test

6.4.1 The packaging container used for testing shall be the smallest-specification packaging container used by the production of the enterprise under the test.

6.4.2 If the enterprise is producing a product with the smallest-specification packaging container, use that product for testing. If the packaging container of the production line is not the smallest specification, the smallest-specification packaging container is filled with water, to simulate the product (commonly known as water can) or other simulants test.

6.4.3 The heat distribution test of the rolling atmospheric-pressure continuous sterilization equipment is not limited by the container specification. However, the heat distribution test of the bucket or tank type atmospheric-pressure continuous sterilization equipment still needs to be carried out using the product packaged in the smallest specification.

6.5 Initial temperature of canned food

When testing the heat distribution, the initial temperature of canned food shall be slightly lower than or refer to the initial temperature in the sterilization specification.

6.6 Temperature data sampling interval

The temperature sampling time can be self-defined. The temperature data is usually collected every 0.5 min. If the sterilization time is short, the interval time for collecting the temperature can be shortened. Based on the sterilization constant-temperature stage, each temperature measurement probe shall collect no less than 25 temperature data.

Use filled can for heat distribution test. No matter what type of temperature data acquisition instrument, only after the entire sterilization process is completed, can the temperature data acquisition be terminated.

If the heat distribution test is performed with a water can as a load, after 15 min of constant temperature, the data acquisition can be ended; and enter the […]

6.7.5 High-pressure hot water sterilization

For high-pressure and high-temperature hot water sterilization operations using hot water as the heat-transfer medium (including full water type and spray type), after the water is injected to the specified water level, start to heat up. When the water temperature reaches 80 °C, write down the time; and continue to heat to the sterilization constant temperature. The time period from 80 °C to the sterilization constant temperature is defined as the come-up time (CUT for short).

After reaching the sterilization temperature, the sterilization constant temperature timing can be started.

At the end of the constant temperature, enter the cooling stage and complete the whole process of hot water sterilization.

Note 1: In view of the fact that the enthalpy of hot water is not as high as that of steam, the heat transfer rate during sterilization is not as fast as steam; and the temperature in the sterilization equipment is not easy to reach uniformity; so a reasonable come-up time needs to be specified. The U.S. FDA "Minimum CUT" is quoted here, which means: After the minimum come-up time of the sterilization equipment, the temperature in the whole pot has become uniform. If the specified minimum come-up time is 15 min, but the actual come-up takes 14 min to reach the sterilization constant temperature, at this time, the sterilization timing shall not be started. It is necessary to wait for another 1 min and reach the specified 15 min (that is, the two conditions of temperature and time are met at the same time) before the sterilization timing can be started. The purpose is to ensure that the temperature in the sterilization equipment is uniform.

Note 2: Due to the different inlet water temperature, products, and seasons of the sterilization operation, the actual initial water temperature in the sterilization equipment is bound to be inconsistent. In order to have a unified starting point for the set minimum come-up time, this Standard stipulates that the minimum come-up time starts at 80 °C.

Note 3: After obtaining the minimum (shortest) come-up time from the heat distribution test data, combined with the heating capacity of the sterilization equipment, the time can be specified after adding a 10% to 15% time safety factor.

6.7.6 Basketless high-pressure sterilization

After the temperature measurement probes are placed from the top of the […]

8 Heat distribution data analysis

a) After 3 min of constant temperature, the temperature discrete value of all temperature measurement probes shall be reduced. Compared with the reference temperature, the standard deviation (sigma i) of the temperature at different locations in the sterilization equipment during the entire constant-temperature process shall be less than 0.5 °C.

b) After the come-up stage of thermal sterilization is over, and within 1 min of the constant temperature, the lowest temperature of the temperature measurement probe is 1.67 °C (3.0 °F) lower than the reference temperature (refers to the temperature measurement value of the temperature measurement probe placed next to the mercury thermometer, the same below), which is regarded as acceptable.

c) If the come-up stage is over, and within 3 min of constant temperature, the lowest temperature of the temperature measurement probe is 0.56 °C (1.0 °F) lower than the reference temperature, which is considered acceptable.

8.3.2 Hot water sterilization equipment

If the following two conditions are met at the same time, the heat distribution can be considered acceptable:

a) If the come-up stage is over, and within 6 min of constant temperature, the lowest temperature of the temperature measurement probe is 1.1 °C (2.0 °F) lower than the reference temperature, which is considered acceptable.

b) The average temperature value Xi of each temperature measurement probe in the constant-temperature stage is compared with the re-average temperature value X of the average values X1, X2, X3,..., Xn of all other temperature measurement probes. The deviation value is within the range of 0.56 °C.

8.3.3 Atmospheric-pressure sterilization equipment

When using atmospheric-pressure continuous tunnel sterilization equipment to sterilize acidic foods, in the constant-temperature stage, in the heat distribution test, the lowest temperature of the temperature measurement probe shall not be lower than the sterilization reference temperature specified in the process.

9 Test report

It is recommended that the heat distribution test report includes the following: […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 30 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 39948-2021

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
30 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 39948-2021

$270.00