Valid

GB/T 37930-2019Non-destructive testing instruments—Technical specifications of X-ray real-time imaging system for automobile wheel (English PDF)

无损检测仪器 汽车轮毂X射线实时成像检测仪技术要求

Preview PDF

This is a limited preview

Buy now to download the full PDF (24 pages)

Need quotation for your management?

Get a formal quotation for GB/T 37930-2019 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

August 30, 2019

Implementation date

March 1, 2020

Scope

GB/T 37930-2019 is the English-translated version of 无损检测仪器 汽车轮毂X射线实时成像检测仪技术要求.

GB/T 37930-2019 lays down the system composition, X-ray detector, testing environment, testing technology, imaging technology, image quality, image display and observation, image evaluation, test report, image storage and process evaluation for X-ray real-time imaging testing instruments used on automobile wheels.

Clause 1 applies it to those instruments and records no exclusions.

Its contents list names clause 4 on testing personnel, clause 5 on the X-ray real-time imaging testing system, clause 6 on imaging technology, clause 7 on image quality, clause 8 on image display and observation, clauses 9 to 11 on image evaluation, the test report and image storage, clause 12 on process evaluation, and three normative appendices covering the standard image grayscale test paper, the testing methods for resolution and image unsharpness, and process evaluation.

Clause 4 requires operators to hold non-destructive testing qualifications of the corresponding items and grades, to command the testing technique and basic computer operation, and to identify 25 pieces of full data on the Appendix A image within one minute.

Clause 5 describes the system as an X-ray machine, X-ray detector, computer image processing system, mechanical transmission system and test equipment and tooling, caps the X-ray tube focus at 3.0 mm, lets the detector be a flat panel, a linear diode array, an image intensifier tube or an equivalent, sets the leakage air kerma rate 1 m from the focus against Table 1, and calls for the system resolution to be verified at least once every 30 days with the resolution indicator of GB/T 23903 and the method of Appendix B. Clause 6 sets the maximum tube voltage by material and transillumination thickness through Figure 1, distinguishes static imaging, whose multi-frame images serve as the basis for storage and quality evaluation, from dynamic imaging, which may not be used to grade welding quality, fixes the information to be written irreversibly into the image files, and requires unwanted and scattered rays to be shielded with lead plates, copper filter plates or collimators.

Clause 7 asks for the grayscale range of the valid evaluation zone to lie between 40% and 90% of the system maximum, puts protection against ionizing radiation under GB 18871, and sets the operating chamber at 15 to 25 degrees Celsius and the exposure chamber at 5 to 30 degrees Celsius with relative humidity not above 80% and mains fluctuation within plus or minus 10%.

The remaining normative references are JB/T 4730.2-2016 and JB/T 7902 on radiographic testing of pressure equipment and on wire-type image quality indicators, and ASTM E155 on reference radiographs for aluminium and magnesium castings.

The standard carries ICS code 19.100 and Chinese classification N 78, was issued on 30 August 2019 and has applied since 1 March 2020.

Found what you need?

Document preview — GB/T 37930-2019

National Standard of the People's Republic of China

ICS
19.100
Classification
N 78

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative References4
  • 3 Terms and Definitions4
  • 4 Testing Personnel8
  • 5 X-ray Real-time Imaging Testing System8
  • 6 Imaging Technology10
  • 7 Image Quality14
  • 8 Image Display and Observation16
  • 9 Image Evaluation17
  • 10 Test Report17
  • 11 Image Storage17
  • 12 Process Evaluation18
  • Appendix A (Normative) Standard Image Grayscale Test Paper20
  • Appendix B (Normative) Testing Methods for Resolution and Image Unsharpness21
  • Appendix C (Normative) Process Evaluation24

1 Scope

This Standard stipulates the system composition, X-ray detector, testing environment, testing technology, imaging technology, image quality, image display and observation, image evaluation, test report, image storage and process evaluation of X-ray real-time imaging testing instrument for automobile wheel.

This Standard is applicable to X-ray real-time imaging testing instrument for automobile wheel.

2 Normative References

The following documents are indispensable to the application of this document. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.

GB 18871 Basic Standards for Protection against Ionizing Radiation and for the Safety of Radiation Sources;

GB/T 23903 Resolution Indicators for Ray Image;

JB/T 4730.2-2016 Non-destructive Testing of Pressure Equipment - Part 2: Radiographic Testing;

JB/T 7902 Non-destructive Testing - Specification for Wire-type Image Quality Indicators;

ASTM E155 Reference Radiographs for Inspection of Aluminum and Magnesium Castings

3 Terms and Definitions

The following terms and definitions are applicable to this document.

3.1 X-ray Digital Radioscopy

X-ray digital radioscopy refers to an imaging method, which adopts photoelectric

3.7 Flat panel detector

Flat panel detector refers to an X-ray detector which receives and converts into image data output through flat panel two-dimensional image detector array after X-ray is converted into optical (electric) signal through a conversion screen.

3.8 Linear Diode Array; LDA

Linear diode array refers to an X-ray detector which receives and converts into digital signals through linear array image sensor after X-ray is converted into optical (electric) signal through a conversion screen.

NOTE: linear diode array needs to utilize the relative motion with objects to form a digital image of the testing zone.

3.9 Image Intensifier Tube; IIT

Image intensifier tube refers to a device which utilizes the method of photoelectric multiple to obtain high-brightness visible light image on the output screen after X-ray is converted into visible light through a scintillator.

NOTE: generally speaking, it couples with CCD (charge coupled device) or CMOS (complementary metal oxide semiconductor), then, outputs video electrical signals or direct digital signals.

3.10 Sensitivity

Sensitivity refers to the minimum diameter of image quality indicator that is visible to the naked eyes in the displayed perspective image.

NOTE: it shall be expressed in image quality indicators.

3.11 Dynamic Range

Dynamic range refers to the ratio of the maximum value that a detector can output to the dark field signal standard deviation under the condition of conducting no calibration.

3.12 Pixel

Pixel refers to the minimum constitutional unit of digital image and the minimum geometric dimension that can be identified in display image.

NOTE: different grayscale values of various pixels constitute different brightness of digital image.

3.13 Line Pair

Line pair is constituted of a line and a spacing, and the width of the spacing equals to the width of the line.

NOTE: the number of identifiable line pairs within the range of each millimeter of width is

3.14 Signal to noise ratio

NOTE: signal to noise ratio is used to evaluate digital image's noise amplitude; it is a decisive influencing factor in sensitivity testing.

4 Testing Personnel

Personnel engaged in X-ray digital imaging testing shall obtain non-destructive testing qualification of corresponding items and grades, then, they may be allowed in corresponding work.

Testing personnel shall possess knowledge related with the testing technology and grasp corresponding basic computer operating methods.

Testing personnel shall be able to identify 25 pieces of full data on the image in Appendix A from a digital imaging system display within 1 min.

5 X-ray Real-time Imaging Testing System

5.1 System Composition

X-ray real-time imaging system is mainly constituted of X-ray machine, X-ray detector, computer image processing system, mechanical transmission system and testing equipment and tooling, etc.

5.2 X-ray Machine

In accordance with the thickness and texture of the tested part of automobile wheel, and the focal length, select X-ray machine's energy range. The dimension of X-ray tube focus shall be not more than 3.0 mm.

5.3 X-ray Detector

In accordance with different testing requirements and testing conditions, the following X-ray detectors may be selected:

---Flat panel detector;

---Linear diode array detector;

---Image intensifier tube;

---Other detectors that have similar functions as mentioned above.

5.4 X-ray Air Kerma Rate

Under the rated working conditions, the leaking ray air Kerma rate of X-ray device at a distance of 1 m from X-ray tube focus shall comply with the requirements in Table 1.

5.5

20 lp/mm ~ 28 lp/mm, static sensitivity is superior to 2.5% and dynamic sensitivity is superior to 3.5%. When resolution is excessively low, image sharpness might become worse; when resolution is excessively high, the signal to noise ratio and dynamic range of images would become lower.

NOTE: 1 in = 25.4 mm.

Or, in accordance with detailed requirements, flat panel detector and other detectors may be adopted.

5.6.2 Testing of system resolution

After a system is determined or altered, its resolution shall be tested.

Adopt the radiographic testing system resolution indicator described in GB/T 23903 to test system resolution. Please refer to Appendix B for system resolution testing method.

5.6.3 Validation of system resolution

Under the condition of unchanged system, in every 30 d, system resolution shall be validated at least once; validation record shall also be kept. When abnormal image quality is found, system resolution shall be validated at any time.

6 Imaging Technology

6.1 X-ray Tube Voltage

X-ray energy shall select relatively low tube voltage. Figure 1 stipulates the maximum X-ray tube voltage that is allowed to be adopted under different materials and different transillumination thicknesses.

6.5.3 Imaging mode
6.5.3.1 Static imaging

Under the static state of automobile wheel, the detector absorbs a larger radiation dose, then, generates image data, which obtains testing images through multi-frame overlaying (or uniform) processing. The testing images shall be considered as a basis for original image data storage and quality evaluation.

6.5.3.2 Dynamic imaging

When automobile wheel is in uniform motion, dynamically observe the testing images.

This method may be applied to general quality inspection beyond the testing range.

Due to the relatively small absorbed dose, large noise and low sharpness of the detector, dynamic real-time image cannot be considered as a basis for the grading of welding quality.

6.5.4 Image marking

6.5.4.1 On the testing images, information that is advantageous to image evaluation and validation, such as workpiece No., base metal thickness, external diameter, testing personnel and testing date, shall at least be marked. When the same welding seam is saved as multiple testing image files, each image shall contain the above-mentioned information. In image storage, marking information shall be directly written into image files by software and shall be irreversible; there is no need to use typeface images here.

6.5.4.2 When the same test part is under continuous testing, the serial No. of multiple testing images shall be continuous and automatically set up by system software.

6.5.4.3 Through system software, provide positioning indication of the central position of testing images and the range of one-time transillumination length. In the images, center marking, and typeface image, which overlaps and connects the marking are unnecessary.

6.5.5 Shielding of unwanted rays and scattered rays

Unwanted rays and scattered rays shall be shielded as much as possible. Proper measures, such as lead plate, copper filter plate and collimator (grating) may be adopted to shield unwanted rays and scattered rays; restrict the range of the irradiation field.

7 Image Quality

7.1 Sensitivity of Image Quality Indicator

7.1.1 Requirements of image quality indicator

7.3 Image Grayscale Distribution Range

Grayscale range of testing images within the valid evaluation zone should be controlled between 40% ~ 90% of the system's maximum grayscale range. Through image grayscale histogram, directly measure image grayscale distribution range.

7.4 Timing of Image Evaluation

After the quality of testing images satisfies the stipulated requirements, defect grading evaluation of the test parts may be conducted.

7.5 Testing Environment

Protection against ionizing radiation shall comply with the stipulations in GB 18871.

Temperature in operating chamber: 15 °C ~ 25 °C; relative humidity <= 80%.

Temperature in X-ray exposure chamber: 5 °C ~ 30 °C; relative humidity <= 80%; there shall be air ventilation device inside the exposure chamber.

The fluctuation of power source and voltage shall be not more than ± 10%.

The outer shell of the testing equipment shall have favorable ground connection.

X-ray source high-voltage generator shall have independent ground electrode; resistance <= 4 ohm.

8 Image Display and Observation

8.1 Image Display

Testing images may be displayed on black-and-white display or color display through the mode of positive image or negative image. Through the adjustment of the brightness and contrast ratio of display, it shall guarantee that all data in the image in Appendix A can be identified.

8.2 Image Observation

Under the environment of soft light, observe testing images. Image display screen shall be clean; there shall be no obvious light reflection.

8.3 Paper Print Output of Images

Use high-definition printer to output paper testing images. The size of the printed images shall be identical to the full size of the test part, so as to make it convenient for on-site validation of defect locations and on-site quality analysis.

12 Process Evaluation

12.1 Through process test and evaluation, determine process parameters that can satisfy the requirements of image quality, as it is shown in Appendix C.

12.2 After process conditions are changed, process evaluation shall be re-conducted.

12.3 There shall be necessary testing process files, which shall include the following content:

---Basis of testing;

---Scope of application;

---Personnel requirements;

---Equipment conditions;

---Parameters of workpieces;

---Spot check proportion;

---Timing of testing;

---System performance;

---Image quality requirements;

---Selection of image quality indicator;

---Stipulations of image quality indicator placement;

---Resolution tester;

---Mode of transillumination;

---Method of transillumination;

---Direction of transillumination;

---Geometric parameters and sketch maps;

---Parameters of transillumination;

---Magnification;

---One-time transillumination length;

---Number of image frames;

B.3 Testing Methods for Image Resolution and Unsharpness

B.3.1 Testing methods

Place the resolution indicator for X-ray testing images in the position of the test part.

B.3.2 Determination of image resolution

On the display screen, observe images of the resolution indicator for X-ray testing images. When a line pair with separated grid is observed, then, the corresponding value of this line pair shall be image resolution.

B.3.3 Determination of image unsharpness

On the display screen, observe images of the resolution indicator for X-ray testing images. When a line pair with superposed grid is observed, then, the corresponding grid width of this line pair shall be image unsharpness.

B.4 Relations between System Resolution and Image Resolution

System resolution is image resolution when magnification equals to or approaches 1.

It eliminates the influence of process factors on image quality and purely reflects the resolution of X-ray digital imaging equipment itself. When magnification is larger than 1, if the X-ray source adopts small focus, then, image resolution is generally higher than system resolution; if focus dimension is relatively large, image resolution might be lower than system resolution due to the influence of geometric unsharpness.

B.5 Conversion Relations between Image Resolution and Image Unsharpness

The conversion relations of image resolution and image unsharpness in magnitude is "one-half of the reciprocal". […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 37930-2019

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included