GB/T 3399-1982Test method for thermal conductivity of plastics by means of the guarded hot plate (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (7 pages)

Need quotation for your management?

Get a formal quotation for GB/T 3399-1982 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

December 29, 1982

Implementation date

October 1, 1983

Scope

GB/T 3399-1982 (Test method for thermal conductivity of plastics by means of the guarded hot plate) is available as an English-translated PDF.

GB/T 3399-1982 — 1.1 Heat flux Phi. The amount of heat passing through a surface per unit time, measured in W.

Found what you need?

Document preview — GB/T 3399-1982

Contents

  • 1 Definitions3
  • 2 Principle3
  • 3 Specimen3
  • 4 Test Conditions4
  • 5 Test Instrument4
  • 6 Test Procedure5
  • 7 Test Results5
  • 8 Test Report7
  • Additional Information7

1 Definitions

1.1 Heat flux Phi. The amount of heat passing through a surface per unit time, measured in W.

1.2 Heat flux density q. The amount of heat passing through a unit area of a surface per unit

time, measured in W/m².

1.3 Thermal conductivity lambda. The heat flux per unit area per unit temperature gradient

perpendicular to the unit area under steady-state conditions, measured in W/(m·K).

2 Principle

2.1 This method is based on the principle of unidirectional stable heat conduction. When the upper and lower surfaces of the specimen are at different stable temperatures, the heat flux through the specimen's effective heat transfer area, as well as the temperature difference and thickness between the two surfaces, are measured to calculate the thermal conductivity.

3 Specimen

3.1 The specimen shall be a homogeneous hard material, with both surfaces smooth and parallel, free of cracks or other defects. For flat specimens, the unevenness is required to be within 0.5 mm/m.

When testing soft materials or granular materials, a wooden frame is required, and the specimen shall be generally homogeneous.

3.2 The specimen shall be a representative portion of the product, either cut directly from the product or prepared according to the product standard.

3.2.1 For specimens measured by the double-plate thermal conductivity testers, they shall be as similar as possible.

3.3 The specimen shall be circular or square, with a diameter or side length equal to that of the guarded hot plate and a thickness of no less than 5 mm. The maximum thickness is determined by the instrument and shall not exceed 1/8 of its diameter or side length.

3.4 Before testing, the specimen shall be conditioned according to the product standard. If no

product standard available, the specimen shall be conditioned for 24 h according to the conditions specified in GB 2918-1982 Plastics - Standard Atmospheres for Conditioning and Testing.

3.5 Each set of specimens shall consist of at least two specimens.

4 Test Conditions

4.1 Test Environment. It shall comply with the normal temperature and humidity requirements

specified in GB 2918-1982.

4.2 Test conditions

Hot plate temperature. Below 333 K.

Cold plate temperature. Room temperature or the desired temperature.

The temperature difference between the hot and cold plates shall be no less than 10 K, and the temperature gradient across the specimen shall be between 400 K/m and 2000 K/m.

NOTE. If no anti-dew protection measures are in place, the cold plate temperature shall be at least 5 K above the ambient dew point.

5 Test Instrument

The instrument required by this method is a flat-plate thermal conductivity meter with a guarded hot plate. It consists of a hot plate (including a main hot plate and a guarded hot plate), a cold plate, a temperature measuring instrument, and a calorimeter.

5.1 Hot plate. The main hot plate shall have a diameter of at least 100 mm; and the minimum

width of the guarded hot plate shall be at least 1/4 of the hot plate's side length or diameter. A suitable gap shall be provided between the main and guarded hot plates, with the flat area of the gap not exceeding 8% of the main hot plate's area. The hot plates shall be flat and smooth, with surface unevenness within 0.25 mm/m. Temperature shall be uniform, with temperature fluctuations on the main hot plate not exceeding 2% of the temperature difference between the two sides of the specimen; on the guarded hot plate, not exceeding 5%. An average temperature balance detector shall be located between the main and guarded hot plates.

5.2 Cold plate. It shall have the same dimensions and unevenness requirements as the hot plate.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 7 pages — is available in the English PDF.

How to Download GB/T 3399-1982

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
7 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 3399-1982

$155.00