GB/T 23595.6-2009Test methods of rare earth yellow phosphor for white LED lamps - Part 6: Determination of conductivity (English PDF)
白光LED灯用稀土黄色荧光粉试验方法 第6部分:电导率的测定
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 23595.6-2009 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
April 23, 2009
Implementation date
February 1, 2010
Scope
GB/T 23595.6-2009 is the English-translated version of 白光LED灯用稀土黄色荧光粉试验方法 第6部分:电导率的测定.
GB/T 23595.6-2009 is the Chinese standard "Test methods of rare earth yellow phosphor for white LED lamps - Part 6: Determination of conductivity". Its scope clause reads: This section provides 440nm ~ 480nm blue light excitation white LED measurement method using rare earth yellow phosphor conductivity.
This section applies to 440nm ~ 480nm blue light excitation measurement of electric conductivity of rare earth yellow phosphor white LED lights. Determination Fan Wai.
0µS/cm ~ 100µS/cm. Its clauses include principle of the method; instruments and reagents; test procedure.
Found what you need?
Document preview — GB/T 23595.6-2009
National Standard of the People's Republic of China
- ICS
- 77.120.99
- Classification
- H 14
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- 1 Scope
- 2 principle of the method
- 3 Instruments and reagents
- 4 Test Procedure
1 Scope
This section provides 440nm ~ 480nm blue light excitation white LED measurement method using rare earth yellow phosphor conductivity.
This section applies to 440nm ~ 480nm blue light excitation measurement of electric conductivity of rare earth yellow phosphor white LED lights. Determination Fan
Wai. 0µS/cm ~ 100µS/cm.
2 principle of the method
Under certain ambient temperature, the solution is directly proportional to the ion concentration and conductivity.
3 Instruments and reagents
3.1 Deionized water
Conductivity <=2µS/cm.
3.2 Balance
Precision 0.001g.
3.3 Conductivity Meter
Measuring range 0µS/cm ~ 1000µS/cm, with temperature compensation, precision ± 0.2µS/cm.
3.4 plastic beaker
50mL.
3.5 magnetic stirrer
power <=90W, adjustable speed timing.
3.6 Magnetic sub
about 25mm.
3.7 Pipette
20mL.
4 Test Procedure
4.1 Production Solution
4.1.1
Using an analytical balance (3.2) weigh the sample into 2g 50mL plastic beakers (3.4), and with a pipette (3.7) was added
20mL of deionized water (3.1), into a length of about 25mm magnetic sub (3.6), cover the slides.
4.1.2
The beaker containing the sample and water on a magnetic stirrer (3.5) After stirring 20min standing on stratification.
4.1.3
clear liquid upper layer was removed, was measured.
4.2 Test
4.2.1 Instrument Calibration
according to manufacturer's instructions prescribed method to calibrate the instrument.
4.2.2
Setting the cell constant and temperature compensation coefficient.
4.2.3
The temperature sensor and an electrode to be immersed in the sample with deionized water, shaking the liquid, read deionized water conductivity (deltaO) steady.
4.2.4
The temperature sensor and the electrode in the sample solution, shaking the liquid, when the display is stable, read the solution conductivity (deltaS).
5 Test results are expressed 5.1 conductivity of the sample solution (delta) according to equation (1).
delta = deltaS-deltaO (1)
Where.
deltaS --- conductivity of the solution, in units of µS/cm;
deltaO --- Deionized water conductivity, unit µS/cm.
5.2
Sample test three times in a row, the mean value. When delta<=20µS/cm, the poor value <=2µS/cm; when 20µS/cm < delta<= […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download GB/T 23595.6-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 23595.1-2025 — Test methods of rare earth phosphors for LED — Part 1: Determination of spectrum
GB/T 23595.2-2025 — Test methods of rare earth phosphors for LED — Part 2: Determination of relative brightness
GB/T 23595.3-2025 — Test methods of rare earth phosphors for LED — Part 3: Determination of chromaticity coordinates