GB/T 23367.1-2024Methods for the chemical analysis of lithium cobalt oxide - Part 1: Determination of the cobalt content - EDTA titration and potentiometric titration (English PDF)
钴酸锂化学分析方法 第1部分:钴含量的测定 EDTA滴定法和电位滴定法
Open the GB/T 23367.1-2024 preview as PDF
This is a limited preview
Buy now to download the full PDF (18 pages)
Issued by
State Administration for Market Regulation; Standardization Administration of China
Level / Type
National · Recommended
Issue date
November 28, 2024
Implementation date
June 1, 2025
Scope
GB/T 23367.1-2024 is the English-translated version of 钴酸锂化学分析方法 第1部分:钴含量的测定 EDTA滴定法和电位滴定法.
China's national method for determining the cobalt content of lithium cobalt oxide, by EDTA titration and by potentiometric titration. It is Part 1 of GB/T 23367 and specifies the determination of cobalt in the cathode material. Cobalt is the expensive element in a lithium-ion battery and, in this material, the majority constituent: lithium cobalt oxide is roughly sixty per cent cobalt by mass, and cobalt has been among the most volatile commodities in the world, with supply concentrated in one country and a price that has moved by a factor of three within a year. So the cobalt assay is not a quality check but the basis on which the material is bought and sold, and the tonnage of a shipment is settled on it. That is why the method is titrimetric rather than instrumental. A titration is an absolute measurement: the analyte reacts stoichiometrically with a standardised reagent and the result depends on the purity of that reagent and on the accuracy of a burette, not on a calibration curve built from standards that might themselves be wrong. For a determination with money attached, at a concentration where precision of a fraction of a per cent matters, that traceability is worth more than the speed of an instrumental method. Two routes are given: complexometric titration with EDTA against a visual indicator, which needs no instrument at all, and potentiometric titration, which locates the endpoint electrically and is more precise and less dependent on the analyst. The standard fixes the dissolution of a material that is not easily dissolved, the removal or masking of the interferences, the standardisation of the titrant, both endpoint procedures and the precision. Issued on 28 November 2024 and in force since 1 June 2025, it replaces GB/T 23367.1-2009.
Document preview — GB/T 23367.1-2024
National Standard of the People's Republic of China
- ICS
- 77.160
- Classification
- H 71
- Replacing
- GB/T 23367.1-2009
Issued by: State Administration for Market Regulation; Standardization Administration of China
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Method 1: EDTA titration
- 4.2 Reagents or materials
- 4.2.3 Nitric acid (rho =
- 4.2.4 Perchloric acid (rho =
- 4.2.6 Ammonia-ammonium chloride buffer solution (pH ~ 10): WEIGH
- 4.2.7 Murexide indicator: WEIGH
- 4.5.1 Specimen Weigh
- 4.5.4 Determination
- 4.5.4.3 TRANSFER
- 5 Method 2: potentiometric titration...
- 5.2.3 Nitric acid (rho =
- 5.2.4 Perchloric acid (rho =
- 5.2.6 Ammonium chloride-ammonium citrate-ammonia mixed solution (pH ~ 10): WEIGH
- 5.2.7 Preparation of potassium ferrocyanide (K3[Fe(CN)6]) standard solution (about
- 5.2.8 Standardization of potassium ferrocyanide (K3[Fe(CN)6]) standard solution (about
- 5.3 Instruments and equipment
- 5.4 Samples
- 5.5 Test steps
- 5.5.1 Specimen Weigh
- 5.5.3 Determination
- 6 Precision...
- 7 Test report...
Foreword
Lithium-ion batteries are the core of 3C consumer electronic products, and lithium cobalt oxide positive electrode materials are important raw materials for 3C lithium-ion batteries. Therefore, vigorously developing lithium cobalt oxide materials is a prerequisite for achieving high-quality development of 3C consumer electronic products. GB/T 23367.1-2009 "Methods for chemical analysis of lithium cobalt oxide - Part 1: Determination of cobalt content - EDTA titration" was issued in March 2009 and implemented in January 2010. It has been more than ten years since then. The implementation and application of this Standard has unified the determination method of cobalt content in lithium cobalt oxide, significantly reducing the communication cost and trade cost of both supply and demand sides for the determination of cobalt content in lithium cobalt oxide products. In recent years, with the iteration of lithium cobalt oxide products and the upgrading of testing technology, the determination method of cobalt content in lithium cobalt oxide products has also gradually developed from the traditional EDTA manual titration to the EDTA automatic titration and redox potentiometric titration, and more stringent requirements have been put forward for the test conditions. GB/T 23367 "Methods for chemical analysis of lithium cobalt oxide" consists of 2 parts. - Part 1: Determination of cobalt content - EDTA titration and potentiometric titration; - Part 2: Determination of lithium, nickel, manganese, magnesium, aluminium, iron, sodium, calcium and copper content - Inductively coupled plasma atomic emission spectrometry. This document is formulated based on the needs of domestic and foreign lithium cobalt oxide trade. The revision of this Standard can make the determination method of cobalt content in lithium cobalt oxide products further adapt to the current level of technological development, promote the progress of related testing technologies, provide technical guidance for related domestic industries, and promote the continuous healthy and stable development of the lithium-ion battery positive electrode material industry. It is convenient for suppliers and customers to use unified analytical methods for testing, which is conducive to reducing commercial disputes caused by differences in testing methods, and has good economic and social benefits. Methods for chemical analysis of lithium cobalt oxide - Part 1: Determination of cobalt content - EDTA titration and potentiometric titration
1 Scope
China's national method for determining the cobalt content of lithium cobalt oxide, by EDTA titration and by potentiometric titration. It is Part 1 of GB/T 23367 and specifies the determination of cobalt in the cathode material. Cobalt is the expensive element in a lithium-ion battery and, in this material, the majority constituent: lithium cobalt oxide is roughly sixty per cent cobalt by mass, and cobalt has been among the most volatile commodities in the world, with supply concentrated in one country and a price that has moved by a factor of three within a year. So the cobalt assay is not a quality check but the basis on which the material is bought and sold, and the tonnage of a shipment is settled on it. That is why the method is titrimetric rather than instrumental. A titration is an absolute measurement: the analyte reacts stoichiometrically with a standardised reagent and the result depends on the purity of that reagent and on the accuracy of a burette, not on a calibration curve built from standards that might themselves be wrong. For a determination with money attached, at a concentration where precision of a fraction of a per cent matters, that traceability is worth more than the speed of an instrumental method. Two routes are given: complexometric titration with EDTA against a visual indicator, which needs no instrument at all, and potentiometric titration, which locates the endpoint electrically and is more precise and less dependent on the analyst. The standard fixes the dissolution of a material that is not easily dissolved, the removal or masking of the interferences, the standardisation of the titrant, both endpoint procedures and the precision. Issued on 28 November 2024 and in force since 1 June 2025, it replaces GB/T 23367.1-2009.
This document specifies the method for determining the cobalt content in lithium cobalt oxide. Method 1 is EDTA titration, and method 2 is potentiometric titration. This document applies to the determination of cobalt content in lithium cobalt oxide. Determination range: 56.00 % ~ 62.00 %.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 6682 Water for analytical laboratory use - Specification and test methods
GB/T 8170 Rules of rounding off for numerical values & expression and judgement of limiting values
GB/T 17433 Foundation terms for chemical analysis of metallurgical products
3 Terms and definitions
For the purpose of this document, the terms and definitions defined in GB/T 17433 apply.
4 Method 1: EDTA titration
4.1 Principle After the sample is dissolved in hydrochloric acid (or a mixed solution of nitric acid and perchloric acid), in an ammonia-ammonium chloride buffer solution, with murexide as the indicator and EDTA standard titration solution as the titrant, the cobalt content is calculated based on the volume of EDTA standard titration solution consumed. It can use EDTA manual titration or EDTA automatic titration.
4.2 Reagents or materials
4.2.1 Water, GB/T 6682, grade 3 and above.
4.2.2 Hydrochloric acid (1 + 1).
4.2.3 Nitric acid (rho =
1.42 g/mL), analytical grade or above.
4.2.4 Perchloric acid (rho =
1.67 g/mL), analytical grade or above.
4.2.5 Cobalt standard solution: WEIGH 1.0000 g of metallic cobalt (wCo >= 99.98 %) in a 400 mL beaker, ADD 30 mL of hydrochloric acid (4.2.2), COVER with a watch glass, DISSOLVE completely at low temperature, COOL to room temperature, TRANSFER to a 1000 mL volumetric flask, DILUTE to the mark with water, and MIX well. 1 mL of this solution contains
1.0 mg of cobalt. Commercially available certified cobalt standard solution can also be used.
4.2.6 Ammonia-ammonium chloride buffer solution (pH ~ 10): WEIGH
54.5 g of ammonium chloride and dissolve in water, ADD 350 mL of ammonia, DILUTE to 1000 mL with water, and MIX well.
4.2.7 Murexide indicator: WEIGH
0.4 g of murexide, ADD 50 g of potassium sulfate, and GRIND thoroughly in a mortar. Store in a desiccator.
4.2.8 Preparation of disodium ethylenediaminetetraacetate (C10H14N2O8Na2 - 2H2O, referred to as EDTA) standard titration solution (about
0.01 mol/L). WEIGH
3.72 g of EDTA into a 400 mL beaker, ADD water to dissolve with slight heat, COOL to room temperature, TRANSFER to a 1000 mL volumetric flask, DILUTE to the mark with water, and MIX well.
4.2.9 Standardization of disodium ethylenediaminetetraacetate (C10H14N2O8Na2 - 2H2O, referred to as EDTA) standard titration solution (about
0.01 mol/L) (calibration before use): TRANSFER
15.00 mL of cobalt standard solution (4.2.5) into the titration cup, record as V1. ADD 50 mL of water and 20 mL ~ 22 mL of EDTA standard titration solution (preferably 3 mL ~ 5 mL before the end point of the reaction process), then ADD 10 mL of ammonia-ammonium chloride buffer solution (4.2.6), and finally ADD about
0.1 g of murexide indicator (4.2.7). If EDTA manual titration is selected, directly titrate with EDTA standard titration solution until the solution changes from orange-yellow to purple-red, which is the end point. If EDTA automatic titration is selected, insert the photometric electrode into the automatic potentiometric titrator, and titrate to the end point with EDTA standard titration solution under stirring according to the titration end point recognition program set by the instrument.
4.5.1 Specimen Weigh
0.10 g of sample (4.4) to an accuracy of 0.0001 g. The mass of the weighed specimen is recorded as m1.
4.5.2 Parallel test Perform two tests in parallel and take the average value.
4.5.3 Blank test Perform a blank test along with the specimen, and the volume of the EDTA standard titration solution consumed is recorded as V5.
4.5.4 Determination
4.5.4.1 PLACE the specimen (4.5.1) in a 100 mL beaker, ADD 10 mL of hydrochloric acid (4.2.2), COVER with a watch glass, and HEAT on a low-temperature electric hot plate until completely dissolved. After cooling, TRANSFER to a 100 mL volumetric flask, DILUTE to the mark with water, MIX well, and RECORD the constant volume as V6.
4.5.4.2 If the specimen cannot be dissolved by hydrochloric acid (4.2.2), take another specimen (4.5.1) and PLACE it in a 100 mL beaker, ADD a mixed solution of 3 mL nitric acid (4.2.3) and 3 mL perchloric acid (4.2.4) to dissolve, COVER with a watch glass, and HEAT on a low-temperature electric hot plate until completely dissolved. Then PLACE the solution at 120 °C ~ 250 °C and HEAT until the solution volume remains 1 mL ~ 2 mL. REMOVE and cool slightly, RINSE the watch glass and the beaker wall with water, and HEAT at low temperature to dissolve the salt until the solution is clear. After cooling, TRANSFER To a 100 mL volumetric flask, DILUTE to the mark with water, MIX well, and RECORD the constant volume as V6.
4.5.4.3 TRANSFER
25.00 mL of the test solution (
4.5.4.1 or 4.5.4.2) to the titration cup, RECORD it as V7. ADD about 50 mL of water and 19 mL ~ 21 mL of EDTA standard titration solution (depending on the cobalt content of lithium cobaltate, preferably 3 mL ~ 5 mL before the end of the reaction process). After stirring, ADD an appropriate amount of ammonia-ammonium chloride buffer solution (4.2.6) to adjust the pH of the solution to be tested to ~ 10. Finally, ADD about
0.1 g of murexide indicator (4.2.7). If EDTA manual titration is selected, titrate with EDTA standard titration solution (4.2.8) until the solution changes from orange-yellow to purple-red, which is the end point. If EDTA automatic titration is selected, insert the photometric electrode (4.3.2) into the automatic potentiometric titrator (4.3.1), and titrate to the end point with EDTA standard titration solution(4.2.8) under stirring according to the titration end point recognition program set by the instrument.
5.2.2 Hydrochloric acid (1 + 1).
5 Method 2: potentiometric titration...
5.2.3 Nitric acid (rho =
1.42 g/mL), analytical grade and above.
5.2.4 Perchloric acid (rho =
1.67 g/mL), analytical grade and above.
5.2.5 Cobalt standard solution: WEIGH 3.0000 g of metallic cobalt (wCo >= 99.98 %) into a 400 mL beaker, ADD 30 mL of hydrochloric acid (5.2.2), COVER with a watch glass, DISSOLVE completely at low temperature, COOL to room temperature, TRANSFER to a 1000 mL volumetric flask, DILUTE to the mark with water, and MIX well. 1 mL of this solution contains
3.0 mg of cobalt. Commercially available certified cobalt standard solution can also be used.
5.2.6 Ammonium chloride-ammonium citrate-ammonia mixed solution (pH ~ 10): WEIGH
62.5 g of ammonium chloride and 50 g of ammonium citrate, DISSOLVE in water, ADD 350 mL of ammonia water, DILUTE to 1000 mL with water, and MIX well.
5.2.7 Preparation of potassium ferrocyanide (K3[Fe(CN)6]) standard solution (about
0.05 mol/L). WEIGH
17.8 g of potassium ferrocyanide, DISSOLVE in water, FILTER and TRANSFER into a 1000 mL brown volumetric flask, DILUTE to the mark with water, MIX well, and STORE in a dark place.
5.2.8 Standardization of potassium ferrocyanide (K3[Fe(CN)6]) standard solution (about
0.05 mol/L) (calibration before use): ADD
25.00 mL of potassium ferrocyanide standard solution (5.2.7) to a 250 mL beaker, RECORD it as V9, and then ADD 80 mL of ammonium chloride-ammonium citrate-ammonia mixed solution (5.2.6), and MIX well. INSERT the electrode into the potentiometric titrator and titrate with the cobalt standard solution (5.2.5) under stirring until the potential jump, which is the end point. The volume of the cobalt standard solution consumed is recorded as V8. Calculate the titration coefficient of the potassium ferrocyanide standard solution equivalent to the cobalt standard solution according to formula (3): where: K - titration coefficient, the volume ratio of the cobalt standard solution consumed by a unit volume of potassium ferrocyanide standard solution; V8 - volume of the cobalt standard solution consumed during titration, in milliliters (mL); V9 - volume of the potassium ferrocyanide standard solution added, in milliliters (mL). Perform three parallel calibrations. The range of the volume of the cobalt standard solution consumed shall not be greater than
0.05 mL. Take the average value. Otherwise, recalibrate. The calculation result is retained to four decimal places, and the rounding off of numerical values is carried out in accordance with the provisions of GB/T 8170.
5.3 Instruments and equipment
5.3.1 Automatic potentiometric titrator with stirring device.
5.3.2 Redox electrodes matching the instrument.
5.4 Samples
5.4.1 The sample particle size shall not be greater than 100 µm.
5.4.2 Before analysis, the sample shall be dried at 110 °C ± 5 °C for 2 h and placed in a desiccator to cool to room temperature.
5.5.1 Specimen Weigh
1.00 g of sample (5.4) to an accuracy of 0.0001 g. The mass of the weighed specimen is recorded as m2.
5.5.2 Parallel test Perform two tests in parallel and take the average value.
5.5.3 Determination
5.5.3.1 PLACE the specimen (5.5.1) in a 100 mL beaker, ADD a small amount of water to moisten it, slowly ADD 20 mL of hydrochloric acid (5.2.2), COVER with a watch glass, and HEAT at low temperature until the specimen is completely dissolved. After cooling, TRANSFER to a 250 mL volumetric flask, DILUTE to the mark with water, SHAKE well, and RECORD the constant volume as V12.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
Editions of GB/T 23367.1
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 23367.1-2024 | Methods for the chemical analysis of lithium cobalt oxide - Part 1 | first revision | Current |
| GB/T 23367.1-2009 | Methods for the chemical analysis of lithium cobalt oxide - Part 1 | first issue | Superseded |
This page sells the current edition, GB/T 23367.1-2024. Earlier editions are listed for reference only.
How to Buy GB/T 23367.1-2024
- 1Add to cart. Click the "Buy GB/T 23367.1-2024" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 17433-2014 — Foundation terms for chemical analysis of metallurgical products
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
GB/T 8170-2008 — Rules of rounding off for numerical values & expression and judgement of limiting values
Secure payment via Stripe
Payments accepted
GB/T 23367.1-2024
$200.00