Valid

GB/T 22232.2-2026Test method for the thermal stability of chemicals - Part 2: Adiabatic calorimetry (English PDF)

化学物质的热稳定性测定 第2部分:绝热量热法

Open the GB/T 22232.2-2026 preview as PDF

Preview — first pages of GB/T 22232.2-2026 (full document: 37 pages)

This is a limited preview

Buy now to download the full PDF (37 pages)

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

April 30, 2026

Implementation date

August 1, 2026

Scope

GB/T 22232.2-2026 is the English-translated version of 化学物质的热稳定性测定 第2部分:绝热量热法.

GB/T 22232.2-2026 is the Chinese national standard covering adiabatic calorimetry on a chemical - the test that lets a sample's own heat of reaction raise its temperature with no loss to the surroundings, which is the closest laboratory analogue of a runaway reaction in a full-size vessel. The onset temperature and the self-heat rate it produces are what a reactor's safety case is built on. Part 2, first edition, in force since 1 August 2026. It was issued on 30 April 2026 and has been in force since 1 August 2026, as a first edition. The document is under the responsibility of the Standardization Administration of China. This page is published from the official record of the 2026 edition; the clause text of a standard this recent is not yet in circulation, and the figures, limits and tables it contains are those of the document itself, delivered in full with the English translation.

Document preview — GB/T 22232.2-2026

National Standard of the People's Republic of China

ICS
13.300
Classification
A 80

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • 1 Scope
  • 2 Adiabatic heating method. --Part
  • 5 Instruments
  • 5.1 Adiabatic calorimeter
  • 5.2 Calorimeter Pool
  • 6 Calibration and Verification
  • 6.1 Calibration
  • 6.2 Verification
  • 7 Samples
  • 7.1 Requirements
  • 7.2 Sample Pretreatment
  • 8 Test Procedure
  • 8.1 Safety Measures
  • 8.2 Test Environment
  • 8.3 Parameter Settings
  • 8.4 Test Procedure
  • 8.5 Experimental Data Processing
  • 8.5.1 Thermal inertia factor (phi)
  • 8.5.2 Measured adiabatic temperature rise (DeltaTad)
  • 8.5.3 Correction for adiabatic temperature rise by phi value (DeltaTadphi)
  • 8.5.4 Heat released during the reaction (DeltaH)
  • 8.5.5 Measured time to reach maximum reaction rate (TMR)
  • 8.5.6 Correction of phi value for time to reach maximum reaction rate (TMRphi)
  • 9 Test Report

Foreword

This document conforms to GB/T 1.1-2020 "Standardization Work Guidelines Part

1.Structure and Drafting Rules of Standardization Documents". Drafting is scheduled. This document is Part 2 of GB/T 22232.GB/T 22232 has the following parts published. --Part

2.Adiabatic heating method. Please note that some content in this document may involve patents. The issuing organization of this document assumes no responsibility for identifying these patents. This document was proposed and is under the jurisdiction of the National Technical Committee on Standardization of Hazardous Chemicals Management (SAC/TC 251). This document was drafted by: Sinopec Safety Engineering Research Institute Co., Ltd., Chemical Registration Center of the Ministry of Emergency Management, and Ningbo Customs Technology Center. Heart, China University of Petroleum (Beijing), China National Accreditation Service for Conformity Assessment, Sinopec (Ningbo) Safety Technology Co., Ltd., Nanjing University of Science and Technology China Academy of Safety Science and Technology, Shanghai Chemical Industry Research Institute Testing Co., Ltd., Shenyang Chemical Industry Research Institute Co., Ltd., China Petroleum and Chemical Industry Research Institute Industry Federation, Hangzhou Yangyi Technology Co., Ltd., and Zhuhai Jintuan Chemical Co., Ltd. The main drafters of this document are. Dang Wenyi, Chen Youwei, Jin Manping, Chen Liping, Zhang Hongzhe, Zhang Jiedong, Li Yuntao, Zhang Fan, Sun Bing, and Sun Peiqin. Chen Sining, Wang Hao, Wu Juan, Wei Zhenyun, Li Lunhui, Cao Mengran, Xu Sen, Kong Xiangbei, Tu Yahui, Liu Xuan, Wang Ting, Sun Feng, Huang Fei, Zhang Jinmei, Tang Chenfei, Li Songping, Duan Dongquan, Wang Chen, Zhou Xintian, Yang Weihua, Li Wuping.

GB/T 22232 aims to establish a test method for the thermal stability of chemical substances, and is proposed to consist of three parts. --Part

2 Adiabatic heating method. --Part

3.Rapid Screening Calorimetry. Thermal stability is crucial for evaluating the safety of chemical substances and determining safe operating conditions during production, storage, transportation, and use. Parameters. Unstable chemical substances may undergo exothermic reactions such as decomposition and polymerization under heating conditions. If the accumulated heat cannot be dissipated in time, it will... This can cause a sharp rise in temperature and pressure, leading to serious accidents such as fires and explosions. GB/T 22232 aims to establish a system of methods for determining the thermal stability of chemical substances. Part 1, "Differential Scanning Calorimetry," is applicable to rapid... Quickly obtain basic thermal information such as phase transition, heat of reaction, and initial decomposition temperature of chemical substances for preliminary screening and comparison; Part 2 "Adiabatic" Calorimetry simulates the thermal behavior of chemical substances under adiabatic conditions to obtain the initial exothermic temperature, the amount of heat released during the reaction, the maximum reaction rate, and so on. The key safety parameters, such as arrival time, provide core data for process safety design and risk assessment; Part 3, "Rapid Screening Calorimetry," aims to... By utilizing a simple and efficient experimental design, we can quickly obtain basic thermal information such as the heat of reaction and initial decomposition temperature of chemical substances. Part 1, "Differential Scanning Calorimetry," and Part 3, "Rapid Screening Calorimetry," are typically used as preliminary screening and basic characterization methods. For early hazard identification and sample screening; Part 2, "Adiabatic Thermometry," is used for more accurate and realistic safety data measurement in hazardous scenarios. The three parts together constitute a complete methodology chain from rapid qualitative screening to accurate quantitative assessment, allowing users to tailor their approach to different security assessments. Depending on the stage and accuracy requirements, select the appropriate or combined methods. Determination of the thermal stability of chemical substances Part

2.Adiabatic Thermal Method Warning. This document should be used by personnel with prior experience in formal laboratory work. Use of this document may involve certain hazardous activities. Regarding materials, equipment, and operations, this document does not address all potential safety issues. Users are responsible for taking appropriate safety and health precautions. And ensure that it complies with the conditions stipulated by relevant national laws and regulations.

1.Scope This document describes the principles, instruments, calibration and verification, samples, test procedures, and test methods for determining the thermal stability of chemical substances using the adiabatic calorimetry method. Inspection report. This document applies to the determination of the thermal stability of liquid and solid chemical substances using adiabatic calorimetry.

5.1 Adiabatic calorimeter

5.1.1 The adiabatic calorimeter shall be able to measure, display, and record changes in temperature and pressure over time and/or temperature, and shall meet the following requirements.

a) The exothermic detection sensitivity is no greater than 0.02 °C/min;

b) The maximum test temperature shall not be less than 450 °C;

c) The temperature display resolution is not less than 0.01 °C;

d) The maximum test pressure shall not be less than 15 MPa;

e) The pressure display resolution is not less than 1 kPa.

5.1.2 The adiabatic calorimeter consists of a testing unit, a control unit, and an auxiliary and safety system. Its main components include the following parts.

a) The test unit consists of components such as the furnace cover, furnace body, furnace heater, radiant heater, and instrument housing. (Structure diagram follows.) See Figure

2.Heaters are embedded in the furnace cover and furnace body, and radiant heaters are installed at the bottom of the furnace cavity to provide uniform heating. Controlled heating. The furnace cavity temperature sensor accuracy should reach ±0.1 °C, and the sample temperature sensor accuracy should reach ±0.1 °C. The pressure sensor accuracy should reach ±1 kPa.

5.2 Calorimeter Pool

5.2.1 The material can be stainless steel, titanium, Hastelloy, or other metals or glass. The shape should be spherical and the volume should not be less than 7 mL.

5.2.2 Based on the installation location of the sample temperature sensor, calorimeters can be divided into those with the sample temperature sensor inserted internally and those with the sample temperature sensor externally connected. There are three types of calorimeters. the sample temperature sensor sleeve embedded, the sample temperature sensor sleeve embedded, and the structural schematic diagram of the calorimeter is shown in Appendix A.

5.2.3 If the calorimeter and its connecting pipes are reused, the calorimeter and any pipes that the sample may come into contact with should be cleaned with solvent after the test. Remove any remaining samples or residue from other parts and keep them clean and dry to prevent cross-contamination and corrosion.

5.3 Balance The balance should have a weighing range of no less than 100 g and an accuracy of ±0.0001 g.

6.1 Calibration

6.1.1 The adiabatic calorimeter should be calibrated before initial use and when there are significant changes to the testing equipment system, such as replacing the temperature sensor or pressure sensor. Force sensors, heaters, furnaces, etc.; should be calibrated regularly according to the testing frequency during use.

6.1.2 If the entire instrument cannot be calibrated, then the calorimeter's calorimeter temperature sensor, furnace temperature sensor, and pressure sensor should be checked. The sensor readings are calibrated to meet the following calibration requirements.

a) The temperature indication error of the calorimeter pool is within the calibration range of 0 °C~350 °C, and the difference between the temperature indication and the actual value shall not exceed ±2 °C;

b) The temperature indication error of the heating furnace, with a calibration range of 0 °C ~ 350 °C, shall not differ from the temperature indication of the sample temperature sensor by more than [amount not specified]. ±3 °C;

c) Temperature rise rate indication error, the calibration range is the temperature rise rate within 20 minutes starting from the moment the calorimeter pool temperature reaches 100 °C, compared with the set value. The difference between the set values shall not exceed ±3%;

d) Pressure indication error, calibration range 0 kPa~15 MPa, the difference between the pressure indication and the actual value of the sealed pipeline connected to the calorimeter. Not exceeding ±10%.

6.2 Verification

6.2.1 After calibration, the adiabatic calorimeter should be verified using a reference material. During use, it should be verified periodically according to the testing frequency. For verification methods, please refer to Appendix B.

6.2.2 The commonly used reference material is a 20% (mass fraction) di-tert-butyl peroxide (DTBP) toluene solution.

7.1 Requirements

7.1.1 The sample volume should not exceed 2/3 of the calorimeter filling volume.

7.1.2 For highly reactive samples with unknown hazards, preliminary screening should be conducted using differential scanning calorimetry or an equivalent rapid screening calorimetry method. test.

7.1.3 For samples that are prone to chemical reactions with oxygen or water, such as alkyl aluminum, the samples should not come into contact with oxygen or water during the test operation.

7.1.4 For liquid samples with obvious backflow during the exothermic process, a flow-stopping device should be installed at the inlet of the calorimeter to prevent backflow.

7.2 Sample Pretreatment

7.2.1 Heterogeneous liquid samples, such as suspensions, should be homogenized beforehand; viscous liquid samples should be preheated.

7.2.2 Solid samples with a particle size larger than the diameter of the calorimeter inlet tube should be pre-sieved or crushed/ground. Before crushing or grinding the sample... It should be confirmed first that the crushing and grinding operations will not change the physicochemical properties of the sample.

7.2.3 For highly reactive solid samples, their thermal stability and impact/friction sensitivity should be assessed before crushing and grinding. Crushing and grinding should be carried out in accordance with... Adhere to the safe operating procedures for hazardous chemicals, and take effective protective and control measures in light of factors such as sample characteristics, operating environment, and equipment conditions. Measures include, but are not limited to, wearing protective gear, controlling the force and speed of operation, and using specialized protective equipment.

8.1 Safety Measures

8.1.1 Some chemicals involved in this test may release toxic, corrosive, or flammable gases. All test procedures should be conducted in an environment with effective ventilation. When operating in windy conditions, personnel should wear personal protective equipment appropriate for the hazards.

8.1.2 Samples with high reactivity should undergo preliminary screening tests. Based on the results of these tests, an appropriate sample quantity should be selected to prevent adverse reactions due to improper testing. Excessive sample volume caused the calorimeter to rupture.

8.1.3 The pipeline connecting the calorimeter and the pressure sensor should be checked regularly for blockages, and cleaned or replaced in a timely manner.

8.1.4 Before opening the pressure relief valve to reduce pressure and disassembling the calorimeter after the test, the potential hazards of pressure release and remaining samples in the calorimeter should be analyzed. Potential hazards to test operators (e.g., isocyanate generates hydrogen cyanide at high temperatures, acrylonitrile generates hydrogen cyanide at high temperatures, alkyl aluminum thermal testing) (The vehicle may still spontaneously combust after testing), and necessary safety measures should be taken, including but not limited to wearing a gas mask, protective gloves, and goggles, in well-ventilated areas. Operate in well-ventilated areas, equipped with fire extinguishers, etc., to avoid injury to test personnel.

8.2 Test Environment

8.2.1 The test shall be conducted in an environment with a temperature of 15 °C to 35 °C and a relative humidity of no more than 85%.

8.2.2 The adiabatic calorimeter should be installed in a well-ventilated environment, or the outlet of the pressure relief valve should be connected to a suitable well-ventilated environment; The area is free from strong vibrations and corrosive gases, and other cold and heat sources should be avoided.

8.3 Parameter Settings

8.3.1 Select appropriate parameter settings based on the actual situation. The experimental parameter settings are as follows:

a) The test start temperature should be at least 10 °C lower than the initial exothermic temperature of the sample;

b) The end temperature of the test should not exceed the highest test temperature of the adiabatic calorimeter;

c) The temperature rise interval should be set within the range of 3°C to 10°C, and preferably set to 5°C;

d) Waiting time, set within the range of 15 min ~ 30 min, preferably set to 15 min;

e) Search time, set within the range of 5 min ~ 10 min, preferably set to 5 min;

f) The exothermic detection sensitivity should not exceed 0.02 °C/min.

8.3.2 Select the headspace atmosphere of the calorimeter pool according to the test requirements.

8.4 Test Procedure

8.4.1 Select the amount of material that will not chemically react with the sample based on the physicochemical properties of the sample and the maximum pressure range during the test. The hot pool, weigh and record the mass of the hot pool (mb), excluding the mass of the nut and sealing gasket.

8.4.2 Check the airtightness of the system by conducting a pressure test using inert gas to ensure that the calorimeter, piping, and connections can withstand the pressure during the test. It maintains an effective seal and prevents leakage.

8.4.3 Place the weighed calorimeter on the balance, zero the balance after weighing, add an appropriate amount of sample, weigh and record the mass of the added sample. ms, install the calorimeter containing the sample into the adiabatic heating furnace in preparation for testing.

8.4.4 Set up and run the test program until the test is completed.

8.4.5 Record the temperature and pressure data during the test, and observe the initial heat release temperature (T0), the maximum heat release temperature (Tmax), and the termination temperature. Parameters such as thermal temperature (Tf) and time to reach the measured maximum reaction rate (TMR).

8.4.6 After the test is completed and the temperature of the exothermic reaction system of the sample returns to room temperature, record the temperature and pressure of the calorimeter bath after it has cooled to room temperature.

9 Test Report

The report should include at least the following.

a) Basic information about the sample;

b) Test standards (this document);

c) The method of sample preparation or pretreatment (if any);

d) Experimental apparatus and the selected calorimeter;

e) Test conditions;

f) Test results;

g) Start and end dates of the experiment.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 37 pages — is available in the English PDF.

How to Buy GB/T 22232.2-2026

  1. 1Add to cart. Click the "Buy GB/T 22232.2-2026" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
37 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 22232.2-2026

$365.00

$310.00for partners