GB/T 20975.36-2020Methods for chemical analysis of aluminium and aluminium alloys—Part 36:Determination of silver content—Flame atomic absorption spectrometry (English PDF)
铝及铝合金化学分析方法 第36部分:银含量的测定 火焰原子吸收光谱法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 20975.36-2020 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
June 2, 2020
Implementation date
May 1, 2021
Scope
GB/T 20975.36-2020 is the English-translated version of 铝及铝合金化学分析方法 第36部分:银含量的测定 火焰原子吸收光谱法.
GB/T 20975.36-2020 is the part of the Chinese series on the chemical analysis of aluminium and aluminium alloys that determines the silver content. It gives the flame atomic absorption spectrometric method, for 0.0050 % to 2.00 % silver.
It applies to the arbitration determination of silver in aluminium and aluminium alloys. Aluminium producers and analytical laboratories use it. Issued on 2 June 2020, it has applied since 1 May 2021.
Found what you need?
Document preview — GB/T 20975.36-2020
National Standard of the People's Republic of China
- ICS
- 77.120.10
- Classification
- H 12
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Method summary
- 5 Reagents
- 6 Instruments
- 7 Sample
- 8 Analysis steps
- 9 Test data processing
- 10 Precision
- 11 Test report
Foreword
GB/T 20975 "Methods for Chemical Analysis of Aluminum and Aluminum Alloys" is divided into 37 parts.
---Part 1: Determination of mercury content;
---Part 2: Determination of arsenic content;
---Part 3: Determination of copper content;
---Part 4: Determination of iron content;
---Part 5: Determination of silicon content;
---Part 6: Determination of cadmium content;
---Part 7: Determination of manganese content;
---Part 8: Determination of zinc content;
---Part 9: Determination of lithium content by flame atomic absorption spectrometry;
---Part 10: Determination of tin content;
---Part 11: Determination of lead content;
---Part 12: Determination of titanium content;
---Part 13: Determination of vanadium content;
---Part 14: Determination of nickel content;
---Part 15: Determination of boron content;
---Part 16: Determination of magnesium content;
---Part 17: Determination of strontium content;
---Part 18: Determination of chromium content;
---Part 19: Determination of Zirconium Content;
---Part 20: Determination of Gallium Content Butyl Rhodamine B Spectrophotometry;
---Part 21: Determination of calcium content;
---Part 22: Determination of beryllium content;
---Part 23: Determination of antimony content;
---Part 24: Determination of total rare earth content;
---Part 25: Determination of element content by inductively coupled plasma atomic emission spectrometry;
---Part 26: Determination of carbon content by infrared absorption method;
---Part 27: Determination of cerium, lanthanum and scandium content by inductively coupled plasma atomic emission spectrometry;
---Part 28: Determination of cobalt content by flame atomic absorption spectrometry;
---Part 29: Determination of molybdenum content thiocyanate spectrophotometric method;
---Part 30: Determination of hydrogen content by heating extraction thermal conductivity method;
---Part 31: Determination of phosphorus content molybdenum blue spectrophotometric method;
---Part 32: Determination of Bismuth Content;
---Part 33: Determination of potassium content by flame atomic absorption spectrometry;
---Part 34: Determination of sodium content by flame atomic absorption spectrometry;
---Part 35: Determination of Tungsten Content Thiocyanate Spectrophotometric Method;
---Part 36: Determination of silver content by flame atomic absorption spectrometry;
--- Part 37: Determination of Niobium Content.
This part is part 36 of GB/T 20975.
This section was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This part of GB/T 20975 specifies the flame atomic absorption spectrometry for the determination of silver content in aluminum and aluminum alloys.
This section applies to the arbitration determination of silver content in aluminum and aluminum alloys. Flame atomic absorption spectrometry measurement range. 0.0050%~ 2.00%.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated references, the latest version (including all amendments) is used in this document.
GB/T 8005.2 Terminology of Aluminum and Aluminum Alloy Part 2: Chemical Analysis
GB/T 8170-2008 Numerical rounding rules and the expression and determination of limit values
3 Terms and definitions
The terms and definitions defined in GB/T 8005.2 apply to this document.
4 Method summary
The sample was dissolved with hydrochloric acid and nitric acid. In an acid medium, at the wavelength of 328.1nm of flame atomic absorption spectrometer, air-acetylene fire Flame, measuring the absorbance of silver to determine the silver content.
5 Reagents
Unless otherwise stated, only reagents and laboratory secondary water confirmed to be analytically pure are used in the analysis.
5.1 Hydrochloric acid (rho=1.19g/mL).
5.2 Nitric acid (rho=1.42g/mL).
5.3 Hydrochloric acid (1 1).
5.4 Nitric acid (1 1).
5.5 Hydrochloric acid (1 4).
5.6 Silver standard storage solution. weigh 0.5000g silver (wAg>=99.99%) into a 250mL beaker, add 20mL nitric acid (5.4), Cover with a watch glass, heat at low temperature to completely dissolve, remove and cool, transfer to a 500mL volumetric flask, add 40mL nitric acid (5.2), and dilute with water to Scale and mix well. 1mL of this solution contains 1mg of silver.
5.7 Silver standard solution. Pipette 10.00mL silver standard stock solution (5.6) into a.200mL volumetric flask, add 20mL nitric acid (5.2), Dilute to the mark with water and mix well. 1mL of this solution contains 50µg of silver.
6 Instruments
Atomic absorption spectrometer with silver hollow cathode lamp. The instrument should meet the following conditions.
---Characteristic concentration. In a solution consistent with the matrix of the measurement solution, the characteristic concentration of silver should not be greater than 0.03µg/mL.
---Precision. Use the highest concentration of standard solution to measure the absorbance 10 times, and the standard deviation should not exceed 1.0% of the average absorbance; Use the lowest concentration standard solution (not the "zero" concentration standard solution) to measure the absorbance 10 times, and the standard deviation should not exceed the highest The concentration standard solution is 0.5% of the average absorbance.
---Linear working curve. Divide the working curve into five equal sections according to the concentration, the difference between the absorbance difference of the highest section and the absorbance difference of the lowest section The ratio should not be less than 0.70.
7 Sample
Process the sample into chips not larger than 1mm.
8 Analysis steps
8.1 Sample
Weigh the sample with a mass (m) of 0.20g (Chapter 7), accurate to 0.0001g.
8.2 Parallel test
Do two tests in parallel and take the average value.
8.3 Blank test
Do a blank test with the sample (8.1).
8.4 Determination
8.4.1 Place the sample (8.1) in a 100mL beaker, add 20mL of hydrochloric acid (5.3), cover with a watch glass, and after the violent reaction stops, add 1mL of nitric acid (5.2), heat until the sample is dissolved, boil to remove nitrogen oxides, remove, cool to room temperature, add 15mL of hydrochloric acid (5.1). Will melt Transfer the liquid into a 100mL volumetric flask (V0), dilute to the mark with water, and mix. Divide the test solution (V1) into an appropriate volumetric flask (V2) according to Table 1. Dilute hydrochloric acid (5.5) to the mark and mix well.
8.4.2 Using an air-acetylene flame, at the wavelength of 328.1nm of the atomic absorption spectrometer, and at the same time as the series of standard solutions, adjust to zero with water and measure For the absorbance of the blank test solution and the sample solution, the silver concentration in the blank solution (rho0) and the silver concentration (rho) in the test solution are measured from the working curve. versus Simultaneous determination of a series of standard solutions.
8.5 Drawing of working curve
8.5.1 Pipette 0mL, 0.20mL, 1.00mL, 2.00mL, 3.00mL, 4.00mL, 5.00mL silver standard solution (5.7), Group 100mL volumetric flask, dilute to the mark with hydrochloric acid (5.5), and mix.
8.5.2 Under the same conditions as the test solution determination, adjust to zero with water, measure the absorbance of the series of standard solutions, and subtract the "zero" concentration in the series of standard solutions. For the absorbance of the solution, use the concentration of silver as the abscissa and the corresponding absorbance as the ordinate to draw a working curve.
9 Test data processing
The silver content is calculated as the silver mass fraction wAg, calculated according to formula (1).
When the silver content is less than 1.00%, the calculation result retains two significant digits; when the silver content is >=1.00%, the calculation result is expressed to two decimal places Bit. The rounding of values shall be implemented in 3.2 and 3.3 of GB/T 8170-2008.
10 Precision
10.1 Repeatability
The measured value of two independent test results obtained under repeatability conditions, within the average range given below, these two test results The absolute value of r does not exceed the repeatability limit r, and the repeatability limit r does not exceed 5%. The repeatability limit r is linearly interpolated according to the data in Table 2. Method and extension method.
10.2 Reproducibility
The measured values of two independent test results obtained under reproducibility conditions are within the range of the average value given in Table 5. The absolute difference does not exceed the reproducibility limit R, and the reproducibility limit R does not exceed 5%. The reproducibility limit R is linear according to the data in Table 3. Obtained by interpolation and extension methods.
11 Test report
The test report should include the following.
a) Number and name of this part;
b) All necessary information about identification of samples, laboratory, analysis date, report date, etc.;
c) Express the test results in an appropriate form;
d) Abnormal phenomena during the test;
e) Signatures of review and approval personnel.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download GB/T 20975.36-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 8005.2-2011 — Terms of aluminium and aluminium alloys -- Part 2: Chemical analysis
GB/T 8170-2008 — Rules of rounding off for numerical values & expression and judgement of limiting values
GB/T 20975.37-2020 — Methods for chemical analysis of aluminium and aluminium alloys—Part 37:Determination of niobium content