GB/T 19972-2018Sterilization of health care products -- Biological indicators -- Guidance for the selection, use and interpretation of results (English PDF)

医疗保健产品灭菌 生物指示物 选择、使用和结果判断指南

Preview PDF

This is a limited preview

Buy now to download the full PDF (50 pages)

Need quotation for your management?

Get a formal quotation for GB/T 19972-2018 in 30 minutes.

Issued by

State Administration for Market Regulation, China National Standardization Administration

Level / Type

National · Recommended

Issue date

March 15, 2018

Implementation date

October 1, 2018

Scope

GB/T 19972-2018 is the English-translated version of 医疗保健产品灭菌 生物指示物 选择、使用和结果判断指南.

GB/T 19972-2018 guides the selection and use of biological indicators, and the judgement of their results, when they are used in the development, validation and routine monitoring of sterilization processes for health care products.

It is an identical adoption of ISO 14161:2009.

It defines the terms of the field, from bioburden, D value and z value to process challenge device, sterility assurance level and survival-kill window, and then covers the characteristics of biological indicators, the choice of supplier, their use in process development, in sterilization validation and in routine monitoring, the reading of results, culture conditions, third-party requirements, staff training, storage, management and disposal.

It does not cover processes that only physically remove microorganisms, such as filtration, nor liquid sterilization. It replaces GB/T 19972-2005, was issued on 15 March 2018 and has applied since 1 October 2018.

Found what you need?

Document preview — GB/T 19972-2018

National Standard of the People's Republic of China

ICS
11.080.01
Classification
C 47

Issued by: State Administration for Market Regulation, China National Standardization Administration

Contents

  • Foreword
  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and Definitions
  • 4 General
  • 5 biological indicator characteristics
  • 6 supplier selection
  • 7 Biological indicators in process development
  • 8 Sterilization confirmed biological indicators
  • 9 Biological indicators in routine monitoring
  • 10 result
  • 11 Application of biological indicator standards
  • 12 Culture conditions
  • 13 Requirements for third parties
  • 14 Staff Training
  • 15 Storage and management
  • 16 Biological Indicator Processing
  • Appendix A (informative) Microbial inactivation mechanics and counting methods
  • Appendix B (informative) Process Challenge Device
  • Appendix C (informative) Partial negative method for determining D value
  • Appendix D (informative) Example of a transcript of a biological indicator prepared by the user
  • Appendix E (informative) Calculation of z-values
  • Appendix F (informative) Survival curve method measures D value
  • Appendix G (informative) Survival-killing response characteristics
  • Reference

Foreword

This standard was drafted in accordance with the rules given in GB/T 1.1-2009.

This standard replaces GB/T 19972-2005 "Guidelines for the Selection, Use and Result Judgment of Sterilized Biological Indicators for Healthcare Products".

The main technical differences between this standard and GB/T 19972-2005 are as follows.

--- Added "duration", "installation identification", "running identification", "performance identification", "process variable", "reference microorganism", "spore log reduction" Terms and definitions of "less value" "sterility assurance level" and "survival-kill interval" (see 3.6, 3.8, 3.11, 3.12, 3.16, 3.17, 3.19, 3.21 and 3.26);

--- Added the informative appendix "calculation of z values" (see Appendix E);

--- Added the informative appendix "survival curve method to measure D value" (see Appendix F);

--- Added the informative appendix "Survival-killing response characteristics (see Appendix G)".

This standard uses the translation method equivalent to the ISO 14161:2009 "healthcare product sterilization biological indicator selection, use and knot Judgment Guide.

The documents of our country that have a consistent correspondence with the international documents referenced in this standard are as follows.

---GB 18278.1-2015 Health care products Sterilization and moist heat Part 1: Development, validation and validation of medical device sterilization processes General control requirements (ISO 17665-1:2006, IDT)

---GB 18279.1-2015 Health care products Sterilization of ethylene oxide Part 1: Development of medical device sterilization process, indeed Recognition and routine control requirements (ISO 11135-1:2007, IDT)

--- GB 18281.1-2015 Sterilization of biological indicators for health care products - Part 1: General (ISO 11138-1:2006, IDT)

--- GB 18281.2-2015 Sterilization of biological indicators for health care products - Part 2: Biological indications for the sterilization of ethylene oxide (ISO 11138-2:2006, IDT)

--- GB 18281.3-2015 Sterilization of biological indicators for health care products - Part 3: Biological indicators for moist heat sterilization (ISO 11138-3:2006, IDT)

--- GB 18281.4-2015 Sterilization of biological indicators for health care products - Part 4: Biological indicators for dry heat sterilization (ISO 11138-4:2006, IDT)

--- GB 18281.5-2015 Sterilization of biological indicators for health care products - Part 5: Indicator (ISO 11138-5:2006, IDT)

--- GB/T 19973.1-2015 Microwaveological methods for the sterilization of medical devices - Part 1 Ding (ISO 11737-1:2006, IDT)

---GB/T 19974-2005 Characteristics of Sterilization Factor of Healthcare Products and Setting and Confirmation of Sterilization Process of Medical Devices General requirements for general control and control (ISO 14937:2000, IDT)

Introduction

This standard provides guidance on the selection, use, and determination of biological indicators when applied to the development, validation, and monitoring of sterilization processes.

The procedures described in this standard are of a general nature and are not themselves a comprehensive development, validation and monitoring process for the sterilization of healthcare products. The purpose of this standard is not to force the use of biological indicators in a process, but to correctly select them if biological indicators are used. Choose and use guidelines to avoid misleading results.

In this standard, users can obtain biological indicator selection guides for specific sterilization processes and key parameters, and use the health correctly. A guide to objects.

The user should select a biological indicator that is appropriate for the particular sterilization process they employ. Biological indications due to large differences in sterilization processes, it is difficult for the manufacturer to anticipate all possible uses of the product, so the manufacturer can only identify its biological indicator for a particular use. For mining The specific sterilization process used should be properly selected, used, recycled, and judged by the user.

Conditions for storage and transport of biological indicators prior to use, use of biological indicators or sterilizer process parameters may be indicative of biological indicators Adverse effects, in addition to the culture techniques used after exposure to the process, including culture temperature, media type, manufacturer and specific batches may also It is the measured resistance associated with recovery and growth. Therefore, it should be stored and used in accordance with the recommendations of the biomarker manufacturer. After sterilization, biological indicators should be aseptically transferred (if applicable) and cultured as specified by the biomarker manufacturer.

It should be noted that the biological indicator is not intended to indicate that the sterilized item is indeed sterile. But to test a known sterilization process and the effectiveness of the sterilization equipment used, which is based on the principle of assessing microbial lethality at a sterility assurance level. This Class studies should be conducted by trained and appropriate personnel.

1 Scope

This standard provides for the selection, use, and judgment of biological indicators when applied to the development, validation, and routine monitoring of sterilization processes.

This standard applies to biological indicators of existing national standards.

Note 1: See ISO 11138 for an example.

Note 2: The general information provided in this standard also applies to processes and biological indicators not mentioned in the current national standards. Such as new and developing sterilization process.

This standard does not consider processes that physically remove microorganisms only, such as filtration.

This standard does not apply to the use of various combination processes, such as washing and sterilizing or pipe flushing and steaming. This standard does not apply to liquid sterilization processes.

2 Normative references

The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this document. For undated references, the latest edition (including all amendments) applies to this document.

GB/T 24628-2009 Sterilization biological and chemical indicator test equipment for healthcare products (ISO 18472:2006, IDT)

ISO 11135-1 Sterilization of ethylene oxide for health care products - Part 1. Development, validation and routine Control requirements (Sterilizationofhealthcareproducts-Ethyleneoxide-Part 1. Requirements fordevelop- Ment,validationandroutinecontrolofasterilizationprocessformedicaldevices)

ISO 11138-1:2006 Medicinal products - Sterilization of biological indicators - Part 1 products-Biologicalindicators-Part 1: Generalrequirements)

ISO 11138-2 Sterilization of biological indicators for health care products - Part 2. Biological indicators for the sterilization of ethylene oxide (Sterili-zationofhealthcareproducts-Biologicalindicators-Part 2: Biologicalindicatorsforethyleneoxide Sterilizationprocesses)

ISO 11138-3 Sterilization of biological indicators for health care products - Part 3 ofhealthcareproducts-Biologicalindicators-Part 3: Biologicalindicatorsformoistheatsterilization Processes)

ISO 11138-4 Biomedical products - Sterilization of biological indicators - Part 4. Sterilization for dry heat sterilization ofhealthcareproducts-Biologicalindicators-Part 4: Biologicalindicatorsfordryheatsterilization Processes)

ISO 11138-5 Biomedical products - Sterilization of biological indicators - Part 5. Biological indicators for steril (Sterilizationofhealthcareproducts-Biologicalindicators-Part 5. Biologicalindicatorsforlow-tem- Peraturesteamandformaldehydesterilizationprocesses)

ISO 11737-1 Methods of sterilizing microbiology of medical devices - Part 1 (Sterilizationofmedicaldevices-Microbiologicalmethods-Part 1: Determinationofapopulationof Microorganisms products)

ISO 14937 Characteristics of Sterilization Factors for Healthcare Products and Development, Validation and Routine Control of Medical Device Sterilization Processes General requirements (Sterilizationofhealthcareproducts-Generalrequirementsfor characterizationofasterili- Zingagentandthedevelopment,validationandroutinecontrolofasterilizationprocessformedicalde- Vices)

ISO 17665-1 Sterilization of moisturizing products for health care products - Part 1. Development, validation and routine control Requirements (Sterilization of health care products - Moistheat-Part 1. Requirements for the development, Validationandroutinecontrolofasterilizationprocessformedicaldevices)

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Accreditation

Formally indicate that the conformity assessment body has a third-party certificate for the ability to perform specific conformity assessment work.

Note 1: See ISO /IEC 17011 [3].

Note 2: Accreditation itself does not confirm that the laboratory has the qualification to approve a product, but when the approval and certification authority decides whether to adopt a specific laboratory When it comes to data about its own activities, recognition is related to this.

3.2 Aseptic technique

To rule out the conditions and methods used for microbial contamination.

3.3 Bioburden

The total number of viable microorganisms on the surface and/or inside of a product and/or sterile barrier system. [ISO /T S11139, Definition 2.2]

3.4 Biological indicator

A test system containing viable microorganisms with defined resistance to a defined sterilization process. [ISO /T S11139, Definition 2.3]

3.5 D value; D10 value

The time or dose required to inactivate 90% of the total number of test microorganisms under specified conditions. [ISO /T S11139, definition 2.11]

3.6 Duration; holding time

During a sterilization phase, the sterilization variables in the sterilizer and all locations within the load are maintained within the specified limits.

3.7 Bacteria carrier; inoculated carrier

A vector that has been stained with the indicated number of test microorganisms.

Note 1: See ISO 11138-1.

Note 2: The test microorganism is a microorganism used for the production of the carrier for the infection.

3.8 Installation identification; IQ

The process of obtaining evidence and documenting evidence that the equipment has been provided and installed as required by the technical specifications. [ISO /T S11139, definition 2.22]

3.9 Inoculation

A specified number of specific microbial entities are added to the interior or surface of the article.

3.10 Log reduction value; LR

The decrease in the number of viable microorganisms is expressed in units of lg.

3.11 Operational authentication

Obtain evidence and document the process of proving that the installed equipment is in the intended range when the equipment is used in accordance with the equipment operating procedures Running inside. [ISO /T S11139, definition 2.27]

3.12 Performance qualification

Obtain evidence and document the process of proving that equipment that has been installed and operated according to the operating procedures will continue to be stable according to predetermined criteria Produce products that meet the requirements of product specifications. [ISO /T S11139, definition 2.30]

3.13 Process challenge device

A device that constitutes a specific resistance to the sterilization process and is used to evaluate the effectiveness of the sterilization process. [ISO /T S11139, definition 2.33]

3.14 Process challenge site

It is possible to simulate a sterilization factor in a "most unfavorable state" position in a batch of items to be sterilized.

3.15 Process parameter

The specified value of the process variable.

Note: The specified values for the sterilization process include process parameters and their tolerances. [ISO /T S11139, definition 2.34]

3.16 Process variable

Conditions within the scope of the sterilization process can change the effectiveness of the sterilization. Examples. time, temperature, pressure, concentration, humidity, wavelength. [ISO /T S11139, definition 2.35]

3.17 Reference microorganism

A microbial strain obtained by a recognized strain collection institution. [ISO /T S11139, definition 2.39]

3.18 Resistance meter

Test equipment designed to produce a combination of physical and/or chemical variables during the sterilization process.

Note 1: According to ISO 11138-1, definition 3.15 and GB/T 24628-2009, definition 3.11.

Note 2: The biometric indicator resistance tester (BIER) is also referred to.

3.19 Spore log reduction value

spore-log-reduction; SLR The logarithm of the initial number of spores, N0, minus the logarithm of the final number of spores, NF.

3.20 Sterile

No living microorganisms. [ISO /T S11139, definition 2.43]

3.21 Sterility assurance level

The probability of a single living microorganism present on the product after sterilization.

Note: The term "SAL" is expressed in terms of magnitude and is usually 10-6 or 10-3. When applying this magnitude to the sterility assurance level, the 10-6 SAL has a lower The value thus provides a greater level of sterility assurance than the 10-3 SAL. [ISO /T S11139, definition 2.46]

3.22 Sterilization

A confirmed process for making the product free of viable microorganisms.

Note: During sterilization, the nature of microbial inactivation is exponential; thus the survival of microorganisms on a single product can be expressed in terms of probability. although This probability can be reduced to very low, but it is impossible to drop to zero, see "Aseptic Assurance Level". [ISO /T S11139, definition 2.47]

3.23 Sterilization cycle development

The process of determining the appropriate process parameters such that a known product or class of products achieves the desired ratings and label claims.

3.24 Sterilization cycle confirmation

Used to obtain, document, and describe results documenting procedures and to establish a sustainable production of products that meet expected technical specifications. process.

3.25 Supplier

The organization or individual that provided the product.

Examples: Manufacturers, wholesalers, retailers of products, or providers of vendors, services, or information.

Note 1: The supplier can be internal or external to the organization.

Note 2: The supplier sometimes becomes a "contractor" in the case of a contract. [ISO 9000, Definition 3.3.6]

3.26 Survival-killing interval; survival-kill window

When sterilized under the specified conditions, the biological indicator transitions from all living microorganisms (survival time) to all killed microorganisms (killing) The degree of exposure of the time). [ISO 11138-1, definition 3.18]

3.27 Third party

An individual or institution that is recognized as independent of the parties involved in the issues involved.

Note 1: See ISO /IEC Guide 2 [1].

Note 2: The parties concerned are usually the supplier (first party) and the demander (second party).

3.28 User

Individuals and institutions that use biological indicators for their intended use.

Note 1: See ISO 9000 [4].

Note 2: The customer is the customer of the product provided by the supplier. In the case of the contract, the user may be referred to as the "buyer". The user can be a customer, a consignee or a buyer. user It can be internal or external to the organization and represents the "second party".

3.29 Z value

The value of the exposure temperature required to vary the value of D by an order of magnitude during thermal sterilization.

Note: See ISO 11138-3 and ISO 11138-4.

4 General

4.1 This standard provides guidance on biological indicators, which are generally applicable to any sterilization process, including those not specified in the relevant standards. New sterilization process.

4.2 The use of biological indicators is usually in the program documentation and/or instructions.

Note: The use of quality management systems such as ISO 13485 [7] generally meets the requirements of this clause.

4.3 Biological indicators should generally be combined with physical and/or chemical measurements to demonstrate the effectiveness of the sterilization process. When in the process of sterilization Physical and/or chemical variables are outside the specified limits, and the cause of the sterilizer's inability to obtain its process parameters should be evaluated and the problem Corrected. Systems and/or procedures should be established to evaluate any deviation from the periodic process limits and the completeness of the reasons for accepting any deviations Piece.

4.4 A qualified biological indicator includes carrier material, packaging, and certain microbial components that can be disposed of without the need for special container equipment. Health Long conditions should be properly documented and should be described simply and appropriately to avoid user misunderstandings as much as possible.

4.5 Internationally, there is no formal approval system for the sale and use of biological indicators under certain conditions or conditions. but In some countries, regulatory agencies have special procedures for the selection and use of biological indicators and for the identification and management of sterile or sterilized products. Requirements.

4.6 Biological indicators represent a microbial challenge in a sterilization process and are used to verify whether a sterilization process can inactivate those that are The bacterial process has microorganisms that determine resistance. The test microorganisms used in biological indicators have resistance to sterilization, often exceeding the usual The resistance of bio-loaded microorganisms, however, there are also some microorganisms that may exhibit resistance beyond the test microorganisms. a suitable creature The challenge of the display to a particular sterilization process should exceed the actual bioburden on the instrument to challenge this sterilization process. The challenge is microbes.

The number and combination of resistance. If there is reason to believe that the item to be treated may receive some kind of microbial contamination with special resistance, According to the bioburden situation, it is required to extend the sterilization process time.

4.7 Biological indicators are not used in sterilization processes other than those specified by the manufacturer on their product labels, biological indicator species and strain selection Based on the manufacturer's resistance to specific sterilization processes.

The user should ensure that the biological indicator has been confirmed and is suitable for the specific sterilization conditions used. For this reason, it may be necessary to provide some ratio labels More details are listed above. When the biological indicator is used to exceed the reference condition, the user may request that the relevant biological indicator may be issued.

Information on the raw reaction, such as suboptimal humidity condition effects when the biological indicator is used in the ethylene oxide sterilization process. Non-standard sterilization When biological indicators are used in the technology, the biological indicators should be fully described for their resistance to specific sterilization processes, and to a wide range of microorganisms.

Comparisons, including comparisons with various harmful microorganisms and vectors that may form part of the bioburden of the product. Biological indicator and The correspondence of process parameters should be demonstrated clearly.

4.8 Sterilize the product to ensure that the type of biological indicator is suitable for confirmation and/or routine monitoring of a particular sterilization process.

4.9 Manufacturer's recommendations for the use and storage of biological indicators should always be followed. Otherwise, the perfection of the biological indicator cannot be guaranteed.

If the infected carrier is removed from the inner packaging of the biological indicator by the user, its resistance characteristics may change. Therefore, should be from the manufacturer Obtain information about the extent of this change, otherwise the user should determine the specific change in resistance. The user should specify what the dye carrier has. The performance characteristics are suitable for adoption.

4.10 Biological indicators should not be used if they are in the expiration date specified by the manufacturer.

4.11 Users who confirm and/or routinely sterilize with biological indicators should be well trained in the use of biological indicators. Extinguish The time between the completion of the bacterial process and the biomarker test shall be adjusted as specified in 8.2.4. Sterile technology should be used to sterilize The treated microorganisms are transferred to a suitable resuscitation medium.

4.12 ISO 11138 specifies the requirements for the description of biological indicators provided by the manufacturer. The form of the manual can be a label, a small package Clip pages or general technical specifications, sent along with biological indicators. The standard also includes minimum requirements for resistance characteristics. Test strip given Both the parts and methods are reference methods.

4.13 Biometrics users come from a variety of industrial sectors, private companies and health care institutions. Usually, these users are not required to enter Conducting resistance tests for biological indicators, but there are different requirements for their respective quality assurance systems, including for sellers and/or manufacturing Business review (see 6.2.2).

4.14 If necessary, the user shall verify the resistance characteristics, which is a method of replacing the audit and/or the auxiliary audit.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 50 pages — is available in the English PDF.

Referenced standards

How to Download GB/T 19972-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
50 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 19972-2018

$590.00