Scope

GB/T 18882.2-2017 is the English-translated version of 离子型稀土矿混合稀土氧化物化学分析方法 第2部分:三氧化二铝量的测定.

China's national method for determining aluminium oxide in the mixed rare earth oxide produced from ion-adsorption rare earth ore, by two alternative routes: inductively coupled plasma atomic emission spectrometry and titration. Ion-adsorption deposits are the weathered granite clays of southern China, and they are the world's main source of the heavy rare earths - the elements that no other deposit type supplies in quantity. In these ores the rare earths are not locked in a mineral lattice but held on clay surfaces, from which they are stripped by leaching with a salt solution and then precipitated as a mixed oxide. That is what makes them valuable and also what makes them analytically awkward: the same leach that takes the rare earths off the clay also takes aluminium off it, and aluminium follows the rare earths through precipitation. Aluminium oxide in the product is therefore the characteristic contaminant of this ore type, and it matters commercially because it is inert dilution in a material sold on rare earth content, and technically because it interferes with the solvent extraction circuit that separates the individual elements downstream. The standard gives two methods for the same determination so that a laboratory can choose according to its equipment and the level present. The ICP-OES route dissolves the sample in a mixed acid attack including hydrofluoric and perchloric acid and measures aluminium against matrix-matched calibration, which is faster and better at low levels; the titration route is the classical alternative, independent of an emission instrument. Issued on 14 October 2017 and in force since 1 May 2018, it replaces GB/T 18882.2-2008.

Document preview — GB/T 18882.2-2017

National Standard of the People's Republic of China

ICS
77.120.99
Classification
H 14
Replacing
GB/T 18882.2-2008

Issued by: General Administration of Quality Supervision, Inspection and Quarantine; Standardization Administration of the PRC

Contents

  • 1 Scope
  • 2 Method 1: Inductively coupled plasma atomic emission
  • 2.2.1 Hydrochloric acid (rho =
  • 2.2.2 Hydrofluoric acid (rho =
  • 2.2.3 Perchloric acid (rho =
  • 2.2.4 Nitric acid (rho =
  • 2.3 Instruments and equipment
  • 2.3.1 Inductively coupled plasma atomic emission spectrometer, resolution <
  • 2.4 Test sample
  • 2.5 Procedure
  • 3.2.1 Hydrochloric acid (rho =
  • 3.2.2 Hydrofluoric acid (rho =
  • 3.2.4 Perchloric acid (rho =
  • 3.2.5 Nitric acid (rho =
  • 3.2.11 EDTA (Ethylene Diamine Tetraacetic Acid) solution (about
  • 3.2.12 Cresol red indicator (2 g/L): Weigh
  • 3.2.15 Xylenol orange indicator (4 g/L): Weigh
  • 3.2.18 Zinc sulfate solution (about
  • 3.2.19.1 Preparation: Weigh
  • 3.2.19.2 Calibration: Pipette
  • 3.3 Test sample
  • 3.4 Procedure
  • 3.4.4 Preparation of analytical solution

1 Scope

China's national method for determining aluminium oxide in the mixed rare earth oxide produced from ion-adsorption rare earth ore, by two alternative routes: inductively coupled plasma atomic emission spectrometry and titration. Ion-adsorption deposits are the weathered granite clays of southern China, and they are the world's main source of the heavy rare earths - the elements that no other deposit type supplies in quantity. In these ores the rare earths are not locked in a mineral lattice but held on clay surfaces, from which they are stripped by leaching with a salt solution and then precipitated as a mixed oxide. That is what makes them valuable and also what makes them analytically awkward: the same leach that takes the rare earths off the clay also takes aluminium off it, and aluminium follows the rare earths through precipitation. Aluminium oxide in the product is therefore the characteristic contaminant of this ore type, and it matters commercially because it is inert dilution in a material sold on rare earth content, and technically because it interferes with the solvent extraction circuit that separates the individual elements downstream. The standard gives two methods for the same determination so that a laboratory can choose according to its equipment and the level present. The ICP-OES route dissolves the sample in a mixed acid attack including hydrofluoric and perchloric acid and measures aluminium against matrix-matched calibration, which is faster and better at low levels; the titration route is the classical alternative, independent of an emission instrument. Issued on 14 October 2017 and in force since 1 May 2018, it replaces GB/T 18882.2-2008.

This Part of GB/T 18882 specifies the determination method of aluminum oxide content in mixed rare earth oxide of ion-absorpted type rare earth ore. This Part applies to the determination of aluminum oxide content in mixed rare earth oxide of ion-absorpted type rare earth ore, including two methods: Method

2 Method 1: Inductively coupled plasma atomic emission

spectrometry (ICP-OES)

2.1 Principle of the method Decompose the test sample by nitric acid and hydrofluoric acid; after the perchloric acid has smoked out, dissolve it in hydrochloric acid to make it clear; use approximate matrix matching to eliminate the rare earth matrix interference; directly excite by argon plasma light source; perform the spectrum measurement.

2.2 Reagents and materials Unless otherwise specified, all reagents used in this Part are analytical reagents that meet national or industry standards, and the water used is grade-II water.

2.2.4 Nitric acid (rho =

2.2.5 Hydrogen peroxide (30%).

2.2.6 Nitric acid (1+1).

2.3.1 Inductively coupled plasma atomic emission spectrometer, resolution <

0.006 nm (at 200 nm).

2.3.2 Light source: argon plasma light source.

2.4 Test sample

2.4.1 The particle size of the test sample is less than

0.074 mm.

2.4.2 Pre-dry the sample at 105 °C ~ 110 °C for 2 h; then, place it in a desiccator to cool to room temperature.

2.5 Procedure

2.5.1 Test material Weigh 0.1g of test sample (2.4), accurate to 0.000 1 g.

2.5.2 Number of determinations Independently perform two determinations; take the average value.

2.5.3 Blank test Carry out a blank test together with the test material (2.5.1).

2.5.4 Preparation of analytical solution Place the test materials (2.5.1) in a 150 mL polytetrafluoroethylene beaker; add a small amount of water; add 2 mL of nitric acid (2.2.6); heat to decompose for 3 min ~ 5 min; add 2 mL of hydrofluoric acid (2.2.2) and continue to decompose for 3 min ~ 5 min; add 3 mL of perchloric acid (2.2.3) to smoke and evaporate to nearly dry; remove and cool slightly; add 10 mL of hydrochloric acid (2.2.7) and add 3 ~ 5 drops of hydrogen peroxide (2.2.5); heat to dissolve until clear; transfer the test solution into a 100 mL volumetric flask; use water to dilute to the mark; mix well; test.

2.5.5 Series standard configuration According to Table 2, transfer the standard stock solution into six 100 mL volumetric flasks; use water to dilute to the mark; mix well; test. Unless otherwise specified, all reagents used in this Part are analytical reagents that meet national or industry standards, and the water used is grade-II water.

3.2.2 Hydrofluoric acid (rho =

3.2.3 Hydrogen peroxide (30%).

3.2.5 Nitric acid (rho =

3.2.6 Oxalic acid solution (100 g/L).

3.2.7 Sodium hydroxide solution (200 g/L).

3.2.8 Hydrochloric acid (1+1).

3.2.9 Hydrochloric acid (1+4).

3.2.10 Ammonia water (1+1).

3.2.11 EDTA (Ethylene Diamine Tetraacetic Acid) solution (about

0.05 mol/L): Weigh 20 g of EDTA and dissolve it in a small amount of water; transfer it into a 1 000 mL volumetric flask; use water to fix the volume; mix well.

3.2.12 Cresol red indicator (2 g/L): Weigh

0.2 g of cresol red and dissolve it in 100 mL of ethanol solution (1+1).

3.2.13 Phenolphthalein ethanol solution (10 g/L).

3.2.14 Acetic acid-sodium acetate buffer solution (pH 5.5): Weigh 200 g of sodium acetate and dissolve it in a small amount of water; transfer it into a 1 000 mL volumetric flask; add 10 mL of glacial acetic acid (w >= 99.5%); use water to fix the volume; mix well.

3.2.15 Xylenol orange indicator (4 g/L): Weigh

0.4 g of xylenol orange and 20 g of potassium nitrate; dissolve in a small amount of water; transfer to a 100 mL volumetric flask; use water to fix the volume; mix well. Store in a brown bottle.

3.2.16 Sodium fluoride solution (40 g/L).

3.2.17 Aluminum standard solution: Weigh 1.000 0 g of metallic aluminum (spectrally pure, remove all surface oxides before use) into a 500 mL beaker; add 50 mL of water; then, add 40 mL of hydrochloric acid (3.2.1); dissolve at low temperature until clear (add hydrochloric acid and water in between); cool. Transfer to a 1 000 mL volumetric flask; use hydrochloric acid (5+95) to dilute to the mark; mix well. This standard solution contains 1 mg of aluminum in 1 mL.

3.2.18 Zinc sulfate solution (about

0.1 mol/L): Weigh 30 g of zinc sulfate (ZnSO4·7H2O) and dissolve it in an appropriate amount of water; transfer it into a 1 000 mL volumetric flask; use water to fix the volume; mix well.

3.2.19 Zinc sulfate standard solution.

3.2.19.1 Preparation: Weigh

2.4 g of zinc sulfate (ZnSO4·7H2O) and dissolve it in a small amount of water; transfer it into a 1 000 mL volumetric flask; use water to fix the volume; mix well.

3.2.19.2 Calibration: Pipette

10.00 mL of aluminum standard solution (3.2.17) into a 250 mL conical flask; add 20 mL of LEDTA solution (3.2.11); add 1 ~ 2 drops of phenolphthalein ethanol solution (3.2.13); use sodium hydroxide solution (3.2.7) to neutralize until red appears. Use hydrochloric acid (3.2.9) to neutralize until colorless and add 1 drop in excess; add 20 mL of acetic acid-sodium acetate buffer solution (3.2.14); boil slightly at low temperature for 1 min ~ 2 min; remove and cool. Perform the following operations according to the analytical steps (3.4.4). Calculate the concentration of zinc sulfate standard solution according to Formula (2): Where: c - concentration of zinc sulfate, in moles per liter (mol/L); rho0 - mass concentration of aluminum standard solution, in milligrams per milliliter (mg/mL); V1 - volume of aluminum solution pipetted, in milliliters (mL); M - molar mass of aluminum, in grams per mole (g/mol); V2 - volume of zinc sulfate standard solution consumed, in milliliters (mL).

3.3 Test sample

3.3.1 The particle size of the test sample is less than

0.074 mm.

3.3.2 Pre-dry the sample at 105 °C ~ 110 °C for 2 h; then, place it in a desiccator to cool to room temperature.

3.4 Procedure

3.4.1 Test material Weigh 0.25g of test sample (3.3), accurate to 0.000 1 g.

3.4.2 Time of determinations Independently perform two determinations; take the average value.

3.4.3 Blank test Carry out a blank test together with the test material (3.4.1).

3.4.4 Preparation of analytical solution

3.4.4.1 Place the test material (3.4.1) in a 150 mL polytetrafluoroethylene beaker; add 5 mL of hydrochloric acid (3.2.8) and 1 mL of hydrogen peroxide (3.2.3); heat at low temperature to decompose for 3 min; add 5 mL of hydrofluoric acid (3.2.2) and continue to decompose for 5 min; add 8 mL of perchloric acid (3.2.4) until fuming and evaporate to dryness; remove and cool slightly; add 5 mL of hydrochloric acid (3.2.8) to decompose clearly; transfer to a 200 mL beaker; add 10 mL of hot oxalic acid solution (3.2.6); shake well; add 3 ~ 5 drops of cresol red indicator (3.2.12); use hydrochloric acid (3.2.9) and ammonia water (3.2.10) to adjust to orange-red (pH 1.5 ~ 2.0); keep warm at 40 °C for 30 min; cool; transfer the test solution to a 100 mL volumetric flask; dilute to the mark; mix well.

3.4.4.2 Use medium-speed quantitative filter paper to dry-filter the test solution (3.4.4.1) into a 100 mL beaker; transfer

50.00 mL of the solution into a 250 mL conical flask; add 10 mL of nitric acid (3.2.5) and 5 mL of perchloric acid (3.2.4); evaporate at low temperature until all white smoke is emitted; remove and cool; add 5 mL of hydrochloric acid (3.2.8); rinse with a small amount of water; heat slightly until boiling for a few moments to dissolve all the salts.

3.4.4.3 Remove the test solution (3.4.4.2); add 20 mL of EDTA solution (3.2.11); add 1 ~ 2 drops of phenolphthalein ethanol solution (3.2.13); use sodium hydroxide solution (3.2.7) to neutralize until red appears. Use hydrochloric acid (3.2.9) to neutralize until colorless appears and add 1 drop in excess; add 20 mL of acetic acid-sodium acetate buffer solution (3.2.14); boil slightly for 1 min ~ 2 min; remove and cool.

3.4.4.4 Add 2 drops of xylenol orange indicator (3.2.15) to the test solution (3.4.4.3); use zinc sulfate solution (3.2.18) to titrate until it is close to red; then, continue to titrate with zinc sulfate standard solution (3.2.19) until pure red appears (ignore the reading). Add 15 mL of sodium fluoride solution (3.2.16); boil slightly for 1 min ~ 2 min; remove and cool; use zinc sulfate standard solution (3.2.19) to titrate until pure red is the end point. Record the volume of zinc sulfate standard solution (3.2.19) consumed.

GB/T 15076.14-2026")| ## PDF Preview |

GB/T 18882.2-2017 (Right-click > Save as .pdf)

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 9 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of GB/T 18882.2

EditionTitleRevisionStatus
GB/T 18882.2-2017Chemical analysis methods for mixed rare earth oxide of ion-adsorption type rare earth ore - Part 2second revisionCurrent
GB/T 18882.2-2008Chemical analysis methods for mixed rare earth oxide of ion-adsorption type rare earth ore - Part 2first revisionSuperseded
GB/T 18882.2-2002Chemical analysis methods for mixed rare earth oxide of ion-adsorption type rare earth ore - Part 2first issueSuperseded

This page sells the current edition, GB/T 18882.2-2017. Earlier editions are listed for reference only.

How to Buy GB/T 18882.2-2017

  1. 1Add to cart. Click the "Buy GB/T 18882.2-2017" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
9 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 18882.2-2017

$150.00

$130.00for partners