GB/T 11892-1989Water quality. Determination of permanganate index (English PDF)
This is a limited preview
Buy now to download the full PDF (8 pages)
Need quotation for your management?
Get a formal quotation for GB/T 11892-1989 in 30 minutes.
Issued by
National Environmental Protection Authority
Level / Type
National · Recommended
Issue date
December 25, 1989
Implementation date
July 1, 1990
Scope
GB/T 11892-1989 (Water quality. Determination of permanganate index) is available as an English-translated PDF.
GB/T 11892-1989 — This Standard is applicable to the determination of drinking water, water-source water and ground water. The determination scope is 0.5 ~ 4.5 mg/L. For heavy-polluted water, it shall take few water sample for determination after dilution.
This Standard is not applicable to the determination of the amount of organic pollution in industrial waste. If the determination is required, potassium dichromate can be used to determine the chemical oxygen demand.
The inorganic reducing substances in sample such as , S2- and Fe2+ can be determined. When chloride ion concentration is greater than 300 mg/L, it shall use the method for the determination of oxidation in alkaline media.
Found what you need?
Document preview — GB/T 11892-1989
National Standard of the People's Republic of China
Issued by: National Environmental Protection Authority
Contents
- 1 Subject and scope3
- 2 Definition3
- 3 Principle4
- 4 Reagents4
- 5 Apparatus5
- 6 Sample storage5
- 7 Analysis steps5
- 8 Expression of results6
- 9 Precision7
- Annex A (Supplement) Alkaline potassium permanganate oxidation method8
- Additional information8
1 Subject and scope
1.1 Subject
This Standard specified the method for determination of permanganate index in water.
1.2 Scope
This Standard is applicable to the determination of drinking water, water-source water and ground water. The determination scope is 0.5 ~ 4.5 mg/L. For heavy-polluted water, it shall take few water sample for determination after dilution.
This Standard is not applicable to the determination of the amount of organic pollution in industrial waste. If the determination is required, potassium dichromate can be used to determine the chemical oxygen demand.
The inorganic reducing substances in sample such as , S2- and Fe2+ can be determined. When chloride ion concentration is greater than 300 mg/L, it shall use the method for the determination of oxidation in alkaline media.
2 Definition
The permanganate index is a common indicator of the contamination of organic and inorganic oxidizable substances in water. It is defined as. under certain conditions, the equivalent amount of oxygen calculated by the potassium permanganate content consumed by using certain organic and inorganic reducing substances in potassium permanganate oxidizes water sample.
The permanganate index can not be used as an indicator of theoretical oxygen demand or total organic matter content because under certain conditions, many organic matters can only be partially oxidized; volatile organic matter is not included in the measured value.
4.8 Potassium permanganate standard solution, concentration C3 (1/5
KMnO4) of about 0.01 mol/L. pipet 100 mL of potassium permanganate standard stock solution (4.7) in a 1000 mL volumetric flask; use water to dilute to the scale; well mix. This solution can be stored in the dark for several months. Use the same day to calibrate its concentration.
5 Apparatus
5.1 Commonly used laboratory instruments and the following instruments.
5.2 Water bath or equivalent heating device. with enough volume and power.
5.3 Acid burette, 25 mL.
NOTE The new glassware must be cleaned with acidic potassium permanganate solution.
6 Sample storage
After sampling, add sulfuric acid (4.3) to make sample pH 1~2 and analyze as soon as possible. If the storage time exceeds 6h, it needs storing in a dark place at 0~5°C. Must not exceed 2 days.
7 Analysis steps
7.1 Pipet 100.0 mL of sample that has been thoroughly shaken and well
mixed (or pipet appropriate amount for several times; use water to dilute to 100 mL). Place in a 250 mL conical flask. Add 5 ± 0.5 mL of sulfuric acid (4.3).
Use a burette to add 10.00 mL of potassium permanganate standard solution (4.8). Shake well. Place the conical flask in a boiling water bath for 30 min (starts timing when water bath boils).
7.2 After removal, use a burette to add 10 mL of sodium oxalate solution
(4.6) till the solution turns colorless. When it is still heated, use potassium permanganate solution (4.8) to titrate till pink just appears and maintain for 30s. Record the volume of potassium permanganate solution consumed.
7.3 Blank test. use 100 mL of water to replace sample; determine according
to 7.1, 7.2; record the volume of potassium permanganate solution dropped (4.8).
7.4 Add 10.00 mL of sodium oxalate solution (4.6) into the solution that has
been titrated by the blank test (7.3). If necessary, heat the solution to 80°C.
Use potassium permanganate solution (4.8) to carry out the continuous titration till pink just appears and maintain for 30s. Record the volume of ......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 8 pages — is available in the English PDF.
How to Download GB/T 11892-1989
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 7463-2026 — Surface active agents - Determination of the power to disperse calcium soap - Acidimetric method (Modifled Schoenfeldt method)
GB/T 8336-2026 — Special threads gauges for gas cylinders
GB/T 48115-2026 — Technical requirements for GNSS radio occultation detection payload
Secure payment via Stripe
Payments accepted
GB/T 11892-1989
$65.00