Valid

GB/T 10588-2006Specifications for mould growth testing chambers (English PDF)

长霉试验箱技术条件

Open the GB/T 10588-2006 preview as PDF

Preview — first pages of GB/T 10588-2006 (full document: 13 pages)

This is a limited preview

Buy now to download the full PDF (13 pages)

Issued by

AQSIQ; SAC

Level / Type

National · Recommended

Issue date

April 3, 2006

Implementation date

October 1, 2006

Scope

GB/T 10588-2006 is the English-translated version of 长霉试验箱技术条件.

China's national specification for the mould growth testing chambers used to run fungus tests on electrical and electronic products, on components and on materials. Mould is one of the quiet destroyers of equipment sent into a warm and humid climate: the fungi feed on the organic constituents of coatings, insulation, adhesives and plastics, they creep across insulating surfaces and open leakage paths, and they hold moisture against metal long after the air has dried, so a cabinet that was never wetted comes back corroded and electrically unsound. Testing for it means holding a chamber at 29 degrees C and above 90 per cent relative humidity for weeks on end, which is a harder thing to build than it sounds, because the organisms will only grow inside a narrow band of conditions and a single night of drift spoils a month of work. The standard therefore fixes what the chamber must achieve - the test temperature and its tolerance, the temperature gradient and fluctuation, the humidity floor and the air speed over the specimens - and then the construction that makes those figures repeatable: corrosion-resistant internal walls that can be cleaned, condensate drained away rather than left to drip on the samples and never reused for humidification, heaters whose radiant heat does not fall on the specimen, an observation window with lighting, and reliable sealing of door, ducts and refrigeration circuit. Safety gets its own clause, since the machine combines mains voltage with saturated air: insulation resistance hot and cold, a proper protective earth, over-temperature and over-current protection with alarms, and a noise limit. The test methods that follow say how each figure is verified, down to the position and the number of the measuring points inside the working space and the instruments allowed to read them, so that two laboratories testing the same chamber reach the same verdict. It replaces GB/T 10588-1989 and belongs to the series of specifications for environmental test equipment that also covers damp heat, salt mist, high and low temperature and low air pressure chambers.

Document preview — GB/T 10588-2006

National Standard of the People's Republic of China

ICS
19.040
Classification
N 04
Replacing
GB/T 10588-1989

Issued by: General Administration of Quality Supervision, Inspection and Quarantine; Standardization Administration of the PRC

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Conditions of use
  • 4.1 Ambient conditions
  • 4.2 Power supply conditions
  • 4.3 Water supply conditions
  • 4.4 Test load conditions
  • 5 Technical requirements
  • 5.1 Performance of the product
  • 5.2 Requirements for the structure and the appearance of the product
  • 5.3 Safety and environmental protection requirements
  • 6 Test methods
  • 6.1 Principal measuring instruments and devices
  • 6.2 Test conditions
  • 6.3 Method for testing temperature and humidity
  • 7 Inspection rules
  • 8 Marking, packaging and storage
  • Annex A (informative) Method for the identification of suspicious data
  • Annex B (informative) Evaluation of the uncertainty of measurement of temperature deviation

Foreword

This standard is one of the series of specifications for environmental testing equipment. The series is made up of GB/T 10586-2006 damp heat test chambers, GB/T 10587-2006 salt mist test chambers, GB/T 10588-2006 mould growth test chambers, GB/T 10589-1989 low temperature chambers, GB/T 10590-2006 high and low temperature / low air pressure test chambers, GB/T 10591-1989 high temperature / low air pressure test chambers, GB/T 10592-1989 high and low temperature test chambers and GB/T 11158-2006 high temperature test chambers.

From the date of its implementation this standard replaces GB/T 10588-1989 Specifications for mould growth testing chambers.

The main changes with respect to GB/T 10588-1989 are the following: a) a clause on terms and definitions has been added, taking its content from the relevant parts of IEC 60068-3-5 and IEC 60068-3-6; b) the concept of temperature fluctuation has been brought into line with IEC 60068-3-5 and the requirement changed to 1 degree C (see 5.1.3); c) the recording of temperature data has been brought into line with IEC 60068-3-5 and changed to one reading per minute (see 6.3); d) the range of ambient atmospheric pressure among the conditions of use has been widened (4.1); e) requirements have been placed on the quality of the cooling water (4.3 a)); f) phase-failure protection of the supply has been changed to over-current protection (see 5.3.3); g) a requirement for electrical insulation strength has been added (see 5.3.1); h) the testing of temperature and humidity performance is now carried out under no-load conditions (see 6.2.2); i) information on the method of evaluating the uncertainty of measurement of temperature deviation and on its application has been added (see Annex B).

Annex A and Annex B of this standard are informative.

This standard was proposed by the China Machinery Industry Federation and is under the jurisdiction of the Institute of Comprehensive Technical and Economic Research on Instruments of the Machinery Industry.

The standard was issued on 3 April 2006 and has been in force since 1 October 2006.

1 Scope

China's national specification for the mould growth testing chambers used to run fungus tests on electrical and electronic products, on components and on materials. Mould is one of the quiet destroyers of equipment sent into a warm and humid climate: the fungi feed on the organic constituents of coatings, insulation, adhesives and plastics, they creep across insulating surfaces and open leakage paths, and they hold moisture against metal long after the air has dried, so a cabinet that was never wetted comes back corroded and electrically unsound. Testing for it means holding a chamber at 29 degrees C and above 90 per cent relative humidity for weeks on end, which is a harder thing to build than it sounds, because the organisms will only grow inside a narrow band of conditions and a single night of drift spoils a month of work. The standard therefore fixes what the chamber must achieve - the test temperature and its tolerance, the temperature gradient and fluctuation, the humidity floor and the air speed over the specimens - and then the construction that makes those figures repeatable: corrosion-resistant internal walls that can be cleaned, condensate drained away rather than left to drip on the samples and never reused for humidification, heaters whose radiant heat does not fall on the specimen, an observation window with lighting, and reliable sealing of door, ducts and refrigeration circuit. Safety gets its own clause, since the machine combines mains voltage with saturated air: insulation resistance hot and cold, a proper protective earth, over-temperature and over-current protection with alarms, and a noise limit. The test methods that follow say how each figure is verified, down to the position and the number of the measuring points inside the working space and the instruments allowed to read them, so that two laboratories testing the same chamber reach the same verdict. It replaces GB/T 10588-1989 and belongs to the series of specifications for environmental test equipment that also covers damp heat, salt mist, high and low temperature and low air pressure chambers.

This standard specifies the terms and definitions, the conditions of use, the technical requirements, the test methods, the inspection rules and the marking, packaging and storage of mould growth testing chambers (hereinafter referred to as test chambers).

This standard applies to test chambers used to carry out mould growth tests on electrical and electronic products, on other products, on parts and components and on materials.

2 Normative references

The provisions of the following documents become provisions of this standard through reference in it. For dated references, subsequent amendments (excluding corrections) or revisions do not apply to this standard; however, the parties to agreements based on this standard are encouraged to investigate whether the most recent editions may be applied. For undated references, the latest edition applies.

GB/T 191-2000 Packaging - Pictorial marking for handling of goods (eqv ISO 780:1997)

GB/T 6999-1986 Relative humidity conversion tables for environmental testing

GB/T 14048.1-2000 Low-voltage switchgear and controlgear - General rules (eqv IEC 60947-1:1999)

JB/T 9512-1999 Climatic environmental testing equipment and test chambers - Determination of the noise sound power level

JJF 1059-1999 Evaluation and expression of uncertainty in measurement

3 Terms and definitions

The following terms and definitions apply to this standard.

3.1 test chamber - an enclosed cabinet or space, some part of which is able to satisfy the specified test conditions.

3.2 temperature setpoint - the desired temperature set by means of the control device of the test chamber.

3.3 achieved temperature - after stabilization, the temperature at any point within the working space of the test chamber.

3.4 temperature stabilization - the condition in which the temperature at every point in the working space has reached the temperature setpoint and is held within the given tolerance.

3.5 temperature fluctuation - after stabilization, the difference between the highest and the lowest temperature reached at any one point of the working space within a given time interval.

3.6 working space - the part of the test chamber within which the specified conditions can be maintained inside the specified tolerances.

3.7 temperature gradient - after stabilization, the greatest of the differences between the mean temperatures of any two points of the working space over any time interval.

3.8 temperature rate of change - the rate of transition between two given temperatures, measured at the centre of the working space and expressed in degrees C per minute.

3.9 temperature variation in space - after stabilization, the difference, over any time interval, between the mean temperature at the centre of the working space and the mean temperature at each of the other points of the working space.

3.10 temperature extremes - after stabilization, the highest and the lowest temperatures measured within the working space.

3.11 saturation vapour pressure - at constant temperature, the water vapour pressure at which the moisture content of a given volume of air can no longer be increased.

3.12 partial vapour pressure - at constant temperature, that part of the atmospheric pressure of a given volume of air which is due to water vapour.

3.13 relative humidity (RH) - at constant temperature, the ratio of the partial vapour pressure to the saturation vapour pressure in a given volume of air, expressed as a percentage. NOTE: relative humidity is the most usual way of expressing the water vapour content of air.

3.14 humidity stabilization - the condition in which the humidity at every point of the working space has reached the humidity setpoint and is held within the given tolerance.

3.15 achieved humidity - after stabilization, the humidity at any point within the working space of the test chamber.

4.1 Ambient conditions

a) temperature: 15 degrees C to 35 degrees C; b) relative humidity: not greater than 85 per cent; c) atmospheric pressure: 80 kPa to 106 kPa; d) no severe vibration in the surroundings; e) no direct sunlight and no direct radiation from other heat sources; f) no strong air currents in the surroundings - where the surrounding air has to be moved by forced circulation, the air stream shall not blow directly onto the chamber body; g) no influence from strong electromagnetic fields; h) no high concentration of dust and no corrosive substances in the surroundings.

4.2 Power supply conditions

a) voltage: 220 V +/- 22 V or 380 V +/- 38 V; b) frequency: 50 Hz +/- 0.5 Hz.

4.3 Water supply conditions

a) Cooling water. Mains water or recirculated water satisfying the following conditions should be used: water temperature not higher than 30 degrees C; water pressure 0.1 MPa to 0.3 MPa; water quality meeting the standard for industrial water.

b) Humidification water. Where the air is humidified by a method that brings water into direct contact with it, the resistivity of the water shall not be lower than 500 ohm-metres.

4.4 Test load conditions

The load in the test chamber shall satisfy all of the following conditions at the same time: a) the total mass of the load placed in each cubic metre of working chamber volume shall not exceed 80 kg; b) the total volume of the load shall not be greater than 1/5 of the volume of the working chamber; c) on any cross-section perpendicular to the prevailing direction of the air flow, the sum of the areas of the load shall not be greater than 1/3 of the cross-sectional area of the working chamber at that place, and the load shall not be so placed as to obstruct the flow of air.

5.1 Performance of the product

5.1.1 The test temperature and its permissible deviation shall be 29 degrees C +/- 1 degree C.

5.1.2 The temperature gradient shall be not greater than 1 degree C.

5.1.3 The temperature fluctuation shall be not greater than 1 degree C.

5.1.4 The relative humidity inside the test chamber shall be not lower than 90 per cent.

5.1.5 The air speed inside the test chamber shall be not greater than 1 m/s.

5.2 Requirements for the structure and the appearance of the product

5.2.1 The inner walls of the working chamber shall be made of corrosion-resistant material and the surface shall be easy to clean.

5.2.2 Condensate shall not be allowed to drip inside the working space. It shall be drained away continuously, and it shall not be used as humidification water unless it has been treated.

5.2.3 The radiant heat of the heating elements shall not act directly on the test specimens.

5.2.4 The working chamber shall be provided with an observation window and with a lighting device.

5.2.5 The chamber door, the ventilation ducts and the refrigeration system shall be reliably sealed, and there shall be no leakage of air, of water or of oil.

5.2.6 A specimen rack shall be provided on which the specimens can be placed or from which they can be suspended.

5.2.7 The external coating or plating shall be even and smooth and of uniform colour, and shall be free from exposed substrate, blisters, peeling and scratch marks.

5.3 Safety and environmental protection requirements

5.3.1 The insulation resistance between the terminals and the metal enclosure of the chamber shall be above 2 megohms in the cold condition and above 1 megohm in the hot condition, measured with a 500 V insulation tester of accuracy class 1.0; the equipment shall also withstand a dielectric strength test at 1 500 V a.c., 50 Hz, applied for 5 s.

5.3.2 The protective earthing terminal shall be well connected electrically to the enclosure of the test chamber and shall allow a convenient and firm earth connection; it shall comply with 7.1.9 of GB/T 14048.1-2000.

5.3.3 Over-temperature and over-current protection and alarm devices shall be provided.

5.3.4 The noise of the complete machine shall be not higher than 75 dB(A).

6.1 Principal measuring instruments and devices

6.1.1 Anemometer. The sensing range of the anemometer shall be not less than 0.05 m/s.

6.1.2 Thermometer. A temperature measuring system made up of platinum resistance thermometers, thermocouples or similar temperature sensors shall be used, satisfying the following requirements: time constant of the sensor 20 s to 40 s; expanded uncertainty of the temperature measuring system (k = 2) not greater than 0.4 degrees C.

6.1.3 Hygrometer. A wet and dry bulb thermometer or a humidity measuring system made up of other sensors may be used. The expanded uncertainty of the humidity measuring system (k = 2) shall be not greater than 1/3 of the tolerance of the humidity being measured.

6.2 Test conditions

6.2.1 The test conditions shall satisfy the requirements of 4.1, 4.2 and 4.3.

6.2.2 The tests shall be carried out under no-load conditions.

6.3 Method for testing temperature and humidity

6.3.1.1 Three horizontal measuring planes, referred to as the upper, middle and lower planes, shall be established inside the working chamber of the test chamber. The distance from the upper plane to the ceiling of the working chamber shall be 1/10 of the height of the working chamber; the middle plane shall pass through the geometric centre of the working chamber; and the lower plane shall lie 10 mm above the lowest specimen rack. NOTE: where the working chamber has a sloping or pointed roof, the ceiling is taken as the imaginary horizontal plane through the line of intersection between the slope and the vertical wall.

6.3.1.2 The measuring points lie on those three planes. The centre point is at the geometric centre of the working chamber, and the distance from each of the remaining measuring points to the wall of the working chamber is 1/10 of the corresponding side length (Figure 1). For a test chamber whose working chamber volume is not greater than 1 cubic metre, that distance shall be not less than 50 mm.

6.3.1.3 The number of measuring points is related to the size of the working chamber as follows: a) where the volume of the working chamber is not greater than 2 cubic metres there shall be 9 temperature measuring points and 3 relative humidity measuring points, arranged as in Figure 1; b) where the volume is greater than 2 cubic metres there shall be 15 temperature measuring points and 4 relative humidity measuring points, arranged as in Figure 2; c) where the volume is greater than 50 cubic metres the number of temperature and humidity measuring points may be increased as appropriate.

6.3.2.1 The test procedure begins by bringing the temperature in the working space of the test chamber to 29 degrees C and the relative humidity to 93 per cent.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 13 pages — is available in the English PDF.

Referenced standards

Normative references

GB/T 14048.1-2000 Low-voltage switchgear and controlgear - General rules (eqv IEC 60947-1:1999) · JB/T 9512-1999 Climatic environmental testing equipment and test chambers - Determination of noise sound power level · JJF 1059-1999 Evaluation and expression of uncertainty in measurement

How to Buy GB/T 10588-2006

  1. 1Add to cart. Click the "Buy GB/T 10588-2006" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
13 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 10588-2006

$230.00

$195.00for partners