GB 9706.236-2021Medical electrical equipment - Part 2-36: Particular requirements for the basic safety and essential performance of equipment for extracorporeally induced lithotripsy (English PDF)

Also coversGB9706.236-2021

Open the GB 9706.236-2021 preview as PDF

Preview — first pages of GB 9706.236-2021

This is a limited preview

Buy now to download the full PDF

Issued by

State Administration for Market Regulation, China National Standardization Administration

Level / Type

National · Mandatory

Issue date

December 1, 2021

Implementation date

May 1, 2023

Scope

GB 9706.236-2021 (Medical electrical equipment - Part 2-36: Particular requirements for the basic safety and essential performance of equipment for extracorporeally induced lithotripsy) is available as an English-translated PDF.

GB 9706.236-2021 — replace. This document specifies the basic safety and performance of in vitro lithotripsy equipment. This document is applicable to the in vitro induced lithotripsy equipment defined in.201.3.206, including other in vitro induced focused pressure pulses. Other medical application equipment, hereinafter referred to as ME equipment. The applicability of this document is limited to components directly related to lithotripsy. For example (not limited to Yu), pressure pulse generator, patient support equipment and their related imaging and monitoring equipment. Other equipment such as those used in patient treatment The program-controlled computer, X-ray and ultrasound equipment are not included in this document. This document does not apply to. ---Ultrasonic physical therapy equipment intended for physical therapy; ---Ultrasonic equipment (HITU) and other treatment equipment designed for high-intensity therapy as described in Appendix AA.

Document preview — GB 9706.236-2021

National Standard of the People's Republic of China

Classification
C 39

Issued by: State Administration for Market Regulation, China National Standardization Administration

Contents

  • 201.1 Scope, purpose and related standards1
  • 201.2 Normative references2
  • 201.3 Terms and Definitions2
  • 201.4 General requirements3
  • 201.5 General requirements for ME equipment testing4
  • 201.6 Classification of ME equipment and ME systems4
  • 201.7 ME equipment identification, marking and documentation4
  • 201.8 Protection of ME equipment against electric shock hazard5
  • 201.9 Protection of ME equipment and ME system against mechanical hazards5
  • 201.10 Protection against unwanted or excessive radiation hazards (sources)6
  • 201.11 Protection against over-temperature and other hazards (sources)6
  • 201.12 Accuracy of controllers and instruments and protection of dangerous outputs6
  • 201.13 Dangerous conditions and fault states of ME equipment7
  • 201.14 Programmable Medical Electrical System (PEMS)7
  • 201.15 Structure of ME equipment7
  • 201.16 ME System7
  • 201.17 Electromagnetic compatibility of ME equipment and ME systems7
  • 202 *Electromagnetic Compatibility---Requirements and Tests8
  • Appendix9
  • Appendix AA (informative) Special guide and principle explanation10
  • Appendix BB (informative) Coordinate definition, focus and target position11
  • Reference12

Foreword

This document is in accordance with the provisions of GB/T 1.1-2020 "Guidelines for Standardization Work Part 1.Structure and Drafting Rules of Standardization Documents"

Drafting.

This document is part 2-36 of GB 9706 "Medical Electrical Equipment". GB 9706 has released the following parts.

---Part 1.General requirements for basic safety and basic performance;

---Part 1-3.General requirements for basic safety and basic performance Parallel standard. Radiation protection of diagnostic X-ray equipment;

---Part 2-1.Special requirements for the basic safety and basic performance of electron accelerators with energy from 1MeV to 50MeV;

---Part 2-2.Special requirements for the basic safety and basic performance of high-frequency surgical equipment and high-frequency accessories;

---Part 2-3.Special requirements for basic safety and basic performance of shortwave treatment equipment;

---Part 2-5.Special requirements for the basic safety and basic performance of ultrasonic physical therapy equipment;

---Part 2-6.Special requirements for the basic safety and basic performance of microwave therapy equipment;

---Part 2-8.Special requirements for the basic safety and basic performance of therapeutic X-ray equipment with an energy of 10kV to 1MV;

---Part 2-11.Special requirements for the basic safety and basic performance of gamma beam therapy equipment;

---Part 2-12.Special requirements for the basic safety and basic performance of intensive care ventilators;

---Part 2-13.Special requirements for the basic safety and basic performance of the anesthesia workstation;

---Part 2-16.Specific requirements for the basic safety and basic performance of hemodialysis, hemodiafiltration and hemofiltration equipment;

---Part 2-17.Special requirements for the basic safety and basic performance of automatic control brachytherapy after-installation equipment;

---Part 2-18.Specific requirements for the basic safety and basic performance of endoscopic equipment;

---Part 2-19.Special requirements for basic safety and basic performance of infant incubators;

---Part 2-24.Special requirements for basic safety and basic performance of infusion pumps and infusion controllers;

---Part 2-25.Special requirements for basic safety and basic performance of electrocardiograph;

---Part 2-26.Special requirements for the basic safety and basic performance of EEG machines;

---Part 2-27.Special requirements for basic safety and basic performance of ECG monitoring equipment;

---Part 2-28.Special requirements for the basic safety and basic performance of medical diagnostic X-ray tube components;

---Part 2-29.Specific requirements for the basic safety and basic performance of radiotherapy simulators;

---Part 2-36.Special requirements for the basic safety and basic performance of in vitro lithotripsy equipment;

---Part 2-37.Special requirements for the basic safety and basic performance of ultrasonic diagnostic and monitoring equipment;

---Part 2-39.Special requirements for the basic safety and basic performance of peritoneal dialysis equipment;

---Part 2-43.Special requirements for the basic safety and basic performance of interventional X-ray equipment;

---Part 2-44.Special requirements for the basic safety and basic performance of X-ray computed tomography equipment;

---Part 2-45.The basic safety and basic performance of mammography equipment and mammography stereotaxic devices

Require;

---Part 2-54.Special requirements for the basic safety and basic performance of X-ray photography and fluoroscopy equipment;

---Part 2-60.Special requirements for basic safety and basic performance of dental equipment;

---Part 2-63.Special requirements for the basic safety and basic performance of extraoral imaging dental X-ray machines;

---Part 2-65.Specific requirements for the basic safety and basic performance of dental X-ray machines for intraoral imaging.

This document replaces GB 9706.22-2003 "Medical Electrical Equipment Part 2.Special Requirements for the Safety of In Vitro Induced Lithotripsy Equipment", and

Compared with GB 9706.22-2003, in addition to structural adjustments and editorial changes, the main technical changes are as follows.

--- Added the term "energy flow density", "single pulse energy", "stone crushing equipment" (see.201.3.201,.201.32,.201.3.206);

---Added basic performance requirements (see.201.4.3.11);

---Added the minimum, typical and maximum output energy settings of the energy flow density, including the technical parameters of the time integration limit [see

201.7.9.3.101d)].

This document uses the redrafting method to modify and adopt IEC 60601-2-36.2014 "Medical Electrical Equipment Part 2-36.In Vitro Induced Fragmentation

Special requirements for basic safety and basic performance of stone equipment.

The technical differences between this document and IEC 60601-2-36.2014 and the reasons are as follows.

---Regarding normative reference documents, this document has made adjustments with technical differences to adapt to my country's technical conditions and adjustments.

The situation is collectively reflected in.201.2 "Normative Reference Documents", and the specific adjustments are as follows.

* Replace IEC 60601-1.2005 with GB 9706.1-2020, which is modified to adopt international standards;

* Replace IEC 60601-1-2.2007 with YY9706.102-2021 which is modified to adopt international standards;

* The IEC 60601-2-5.2009, which is not normatively cited in the text, is transferred to the reference.

The following editorial changes have been made to this document.

---The term index is deleted.

Introduction

Medical electrical equipment safety standards are also known as the 9706 series of standards, which consist of general standards, parallel standards, special standards, guidelines and interpretations.

---General standards. Medical electrical equipment should be generally applicable to safety standards, that is, equipment that meets the definition of medical electrical equipment should meet this

Basic standard requirements.

---Parallel standards. Medical electrical equipment should be generally applicable to safety standards, but in most cases it is limited to certain specific functions or features.

Sexual equipment only needs to meet the requirements of such standards.

---Special standards. safety standards that a certain type of medical electrical equipment should apply, and not all medical electrical equipment have special

standard.

---Guidelines and explanations. application guidelines and explanations for the relevant requirements in the standards involved.

In the GB 9706 series of standards, in addition to the published standards and this document, the standards that have been included in the plan are as follows.

---Part 2-4.Special requirements for basic safety and basic performance of cardiac defibrillators. The purpose is to establish the basic

Special requirements for safety and basic performance.

---Part 2-22.Special requirements for basic safety and basic performance of laser equipment for surgery, plastic surgery, treatment and diagnosis. The purpose is

Establish special requirements for basic safety and basic performance of laser equipment for surgery, plastic surgery, treatment and diagnosis.

---Part 2-66.Special requirements for basic safety and basic performance of hearing equipment and hearing equipment system. The purpose is to build listening

Specific requirements for basic safety and basic performance of equipment and hearing equipment systems.

This document focuses on the basic safety and basic performance of in vitro lithotripsy equipment.

This document modifies and supplements GB 9706.1-2020 "Medical Electrical Equipment Part 1.General Safety and Basic Performance

Require".

In this document, the asterisk (*) is used as the first character of the title or the beginning of the paragraph or table title, which means that there is a related item in Appendix AA.

Dedicated guides and principle explanations related to the project. The relevant content of Appendix AA is not only helpful for the correct use of this document, but also can be added in a timely manner.

The process of revising standards due to changes in clinical practice or technological development.

Medical Electrical Equipment Part 2-36.In Vitro Initiation

Specific requirements for basic safety and basic performance of stone crushing equipment

201.1 Scope, purpose and related standards

In addition to the following, Chapter 1 of the general standard applies.

201.1.1 Scope

replace.

This document specifies the basic safety and performance of in vitro lithotripsy equipment.

This document is applicable to the in vitro induced lithotripsy equipment defined in.201.3.206, including other in vitro induced focused pressure pulses.

Other medical application equipment, hereinafter referred to as ME equipment. The applicability of this document is limited to components directly related to lithotripsy. For example (not limited to

Yu), pressure pulse generator, patient support equipment and their related imaging and monitoring equipment. Other equipment such as those used in patient treatment

The program-controlled computer, X-ray and ultrasound equipment are not included in this document.

This document does not apply to.

---Ultrasonic physical therapy equipment intended for physical therapy;

---Ultrasonic equipment (HITU) and other treatment equipment designed for high-intensity therapy as described in Appendix AA.

201.1.2 *Purpose

replace.

The purpose of this document is to establish the basic safety and basic performance requirements of in vitro lithotripsy equipment, including according to the definition of.201.3.206

Other medical application equipment for in vitro induced focused pressure pulse therapy.

201.1.3 Parallel standards

Supplement.

This document refers to Chapter 2 of the General Standard and applicable parallel standards listed in.201.2 of this document.

YY9706.102-2021 is applicable after Chapter 202 is revised. GB 9706.103 and YY/T 9706.110 are not applicable.

201.1.4 Specific standards

replace.

In the GB 9706 series, special standards can be modified, replaced or deleted according to the professional ME equipment considered, general standards and parallel standards

The requirements contained in the standards. Other basic safety and basic performance requirements can be added.

The requirements of specific standards take precedence over the requirements of general standards.

In this document, GB 9706.1-2020 is referred to as the general standard. Collateral standards are indicated by their respective document numbers.

The numbering of chapters and articles in this document is prefixed with "201" to correspond to the general standard (for example,.201.1 in this document corresponds to Article

Chapter 1), or add the prefix "20x" to correspond to the applicable collateral standard, where x is the last digit of the standard number corresponding to the collateral standard (for example,

202 in this document corresponds to the content in the parallel standard YY9706.102).

Standardize changes to the text of the general standard by using the following words.

"Replacement" means that the chapters and articles of the general standards or applicable parallel standards are completely replaced by the contents of this document.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.

How to Buy GB 9706.236-2021

  1. 1Add to cart. Click the "Buy GB 9706.236-2021" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB 9706.236-2021

$275.00

$235.00for partners