YY/T 1793-2021Bacterial endotoxin detection kit (English PDF)
细菌内毒素测定试剂盒
This is a limited preview
Buy now to download the full PDF (18 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1793-2021 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
September 6, 2021
Implementation date
March 1, 2023
Scope
YY/T 1793-2021 is the English-translated version of 细菌内毒素测定试剂盒.
YY/T 1793-2021 is the Chinese pharmaceutical industry standard for the kit that measures bacterial endotoxin by the limulus amoebocyte lysate method read photometrically, the endotoxin being the lipopolysaccharide complex of the outer membrane of Gram-negative bacteria.
It sets the requirements, the test methods, the labels and marks, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the sensitivity of the lysate, the linearity and accuracy of the curve, precision, the recovery of endotoxin added to the sample and the stability.
It applies to kits for the quantitative determination of endotoxin in human serum and plasma. Issued on 6 September 2021, it has applied since 1 March 2023.
Found what you need?
Document preview — YY/T 1793-2021
National Standard of the People's Republic of China
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the requirements, test methods, labeling, labeling, instructions for use, packaging, transportation and storage of bacterial endotoxin assay kits and so on.
This standard applies to the quantitative detection of bacterial endotoxins produced by Gram-negative bacteria in human serum and plasma based on the principle of Limulus reagent spectrophotometry. test kit. The chemical composition of bacterial endotoxin is phospholipid polysaccharide-protein complex, and lipopolysaccharide is the component of bacterial endotoxin.
This standard is also applicable to the quantitative detection of Gram-negative bacteria lipopolysaccharide in human serum and plasma based on the principle of Limulus reagent spectrophotometry kit.
2 Normative references
The following documents are essential for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 29791.2 Information provided by manufacturers of in vitro diagnostic medical devices (labeling) Part 2: In vitro diagnostic reagents for professional use
3 Requirements
3.1 Appearance
Appropriate appearance requirements should be specified according to the packaging characteristics of the product. Generally, the composition and properties of each component of the kit should be included; inner and outer packaging, labels requirements for clarity etc.
a) The kit should have complete components, the inner and outer packaging should be complete, and the labels should be clear;
b) There is no leakage of liquid reagents; the lyophilized components are loose bodies, and the liquid is clear after reconstitution (no particles visible to the naked eye, no precipitation).
3.2 Loading capacity
Should meet one of the following requirements.
a) For single-serve liquid reagents that do not need to be extracted and tested by direct sample addition, the deviation of the net content of the liquid reagents should not exceed ±5%;
b) For multiple doses of liquid reagents, the net content of the liquid reagents should not be less than the indicated value.
3.3 Accuracy
Should meet one of the following requirements.
a) Relative deviation. should not exceed ±20%;
b) Recovery rate. should be in the range of 80%~120%.
3.4 Linear
The manufacturer should specify the linear interval of the kit and meet the following requirements.
a) The upper limit of the linear range of the kit should not be lower than 0.16EU/mL, and the lower limit should not be higher than 0.04EU/mL;
b) The linear correlation coefficient |r| shall not be less than 0.980.
Note. If the enterprise adopts the mass unit, the conversion relationship between the mass unit and the EU unit shall be clearly indicated.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1793-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 29791.2-2013 — In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 2: In vitro diagnostic reagents for professional use
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 1793-2021
$150.00