YY/T 1764-2021Standard guide for in vitro axial, bending, and torsional durability testing of vascular stents (English PDF)
血管支架体外轴向、弯曲、扭转耐久性测试方法
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1764-2021 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
September 6, 2021
Implementation date
September 1, 2022
Scope
YY/T 1764-2021 is the English-translated version of 血管支架体外轴向、弯曲、扭转耐久性测试方法.
YY/T 1764-2021 is the Chinese pharmaceutical industry standard that guides three independent in vitro durability tests on vascular stents under cyclic deformation: axial, bending and torsional.
Each reproduces the non-radial movement an artery undergoes from musculoskeletal activity such as walking and breathing and from the beating of the heart, and the standard sets the specimens, the mounting, the amplitude and number of cycles, the medium and temperature, and the examination of the stent afterwards.
It applies to balloon-expandable and self-expanding stents of metal or alloy, including coated stents. Issued on 6 September 2021, it has applied since 1 September 2022.
Found what you need?
Document preview — YY/T 1764-2021
National Standard of the People's Republic of China
- Classification
- C 30
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Test Method Summary
- 5 Meaning and use
- 6 Specimen size, configuration and preparation
- 7 General equipment requirements
- 8 General test parameters
- 9 Test report
- Appendix A (Normative Appendix) Axial Durability of Vascular Stents
- Appendix B (Informative Appendix) Bending Durability of Vascular Stents. Axial Compression Bending
- Appendix C (Informative Appendix) Bending Durability of Vascular Stents. Mandrel Bending
- Appendix D (Informative) Bending Durability of Vascular Stents. Arc Bending Without Mandrel
- Appendix E (Normative Appendix) Torsion Durability of Vascular Stents
- Reference
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies three independent cyclic deformation durability tests related to vascular stents. axial, bending and torsional. The three tests Both model the effect of non-radial arterial motion on stent durability that may originate from musculoskeletal activities, including walking and breathing, to and cardiac exercise.
This standard applies to balloon-expandable and self-expanding stents made of metal or alloys, including coated stents (e.g., with superficial stents), single polymer stents or absorbable stents, as well as stents for the treatment of aneurysmal disease, peripheral vascular injury, or to provide vascular access Endovascular prostheses (stent grafts), and this standard applies to the testing of the above-mentioned stents (or a representative part of the stents).
This standard does not apply to the testing of alternatives, although durability testing of alternatives (such as scaled-up partial support structures) may Provide useful information.
Note 1.This standard does not cover flat, local extrusion durability or multi-mode testing. Although this standard does not cover multimodal testing, the information described in the text can Help build multimodal tests.
Note 2.For the treatment of aneurysm disease, peripheral vascular injury, or other tubular products to provide vascular access can refer to this standard.
Note 3.This standard does not provide in vivo physiological deformation conditions for vascular stents. In these durability tests, develop and justify these boundary conditions (e.g. such as literature studies, in vivo studies, cadaveric studies or modeling of stent-vascular interactions) is the responsibility of the standard user. Other than that, it may be necessary to consider Conditions include vascular calcification, vascular taper, eccentric lesions, loading offsets (eg, motion), and vascular remodeling.
Note 4.This standard does not specify the specific properties of the test product.
2 Normative references
The following documents are essential for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
YY/T 0808-2010 Standard test method for in vitro pulsation durability of vascular stents
3 Terms and Definitions
The following terms and definitions apply to this document.
3.1 Axial
Linear compression or tension along the longitudinal axis of the stent and/or simulated vessel.
3.2 Bent bending
Deformation of the longitudinal axis of the stent and/or simulated vessel to achieve a specific stent radius of curvature.
3.3 Torsional
Stenting and/or simulating twisting of the longitudinal axis of the vessel.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1764-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0808-2010 — Standard test methods for in vitro pulsatile durability testing of vascular stents
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
Secure payment via Stripe
Payments accepted
YY/T 1764-2021
$360.00