YY/T 1763-2021Medical electrical equipment—Medical light ion beam equipment—Performance characteristics (English PDF)
医用电气设备 医用轻离子束设备 性能特性
This is a limited preview
Buy now to download the full PDF (67 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1763-2021 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 9, 2021
Implementation date
October 1, 2022
Scope
YY/T 1763-2021 is the English-translated version of 医用电气设备 医用轻离子束设备 性能特性.
YY/T 1763-2021 is the Chinese pharmaceutical industry standard for medical light ion beam equipment, the accelerators that treat tumours with protons and other light ions.
It specifies the performance characteristics of such equipment and the test procedures by which they are verified: the beam energy and its range, the dose and dose rate delivered and their accuracy and reproducibility, the geometry and position of the beam, the patient positioning system and the safety interlocks and radiation protection that go with them.
It applies to light ion beam equipment with a specific energy between 10 MeV and 500 MeV per nucleon used for radiotherapy, and not to other ion beam equipment. Issued on 9 March 2021, it has applied since 1 October 2022.
Found what you need?
Document preview — YY/T 1763-2021
National Standard of the People's Republic of China
- Classification
- C 43
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Environmental conditions
- 4.1 Overview
- 4.2 Transportation and storage
- 4.3 Facility stability
- 5 Information provided to users
- 5.1 Overview
- 5.2 Expected parameters
- 6 Beam distribution
- 6.1 Light ion type
- 6.2 Energy
- 6.3 Beam gating
- 6.4 Frame
- 6.5 Bracket for beam limiting tube
- 6.6 Light ion beam limiting tube
- 6.7 Adjustable beam limiting devices (BLDs)
- 6.8 Isocenter
- 6.9 Horizontal expansion device (LSD)
- 6.10 Time constraints
- 6.11 Maintenance
- 7 Dose monitoring system
- 7.1 Overview
- 7.2 Standard test conditions for testing the dose monitoring system
- 7.3 Repeatability of MU output
- 7.4 Linearity of MU output
- 7.5 Off-axis response for modulated scanning beam flux monitoring
- 7.6 The effect of angular position on MU output
- 7.7 Stability of MU output
- 8 Depth dose characteristics
- 8.1 Non-range modulation field depth dose distribution
- 8.2 Range modulation technology
- 8.3 Range modulation depth dose distribution
- 8.4 Light ion range stability
- 9 Lateral distribution of light ion radiation field
- 9.1 Transverse curve of light ion field using scatterer or uniform scanning system
- 9.2 Transverse curve of light ion field using modulated scanning system
- 10 Light ion field with energy and fluence modulation (EFM)
- 10.1 User Information
- 10.2 Beam characteristics and dose system performance for delivering small doses
- 11 Irradiation time for a specified volume
- 11.1 User Information
- 11.2 Test
- 12 Instructions for the radiation field
- 12.1 Limiting tube bracket extension indicator
- 12.2 Indication of the position of multiple BLD components
- 12.3 Indication of light ion reference axis
- 12.4 Light field indicator
- 13 Patient support device
- 13.1 Overview
- 13.2 Bed surface
- 13.3 Coordinate system
- 13.4 Range of motion of the patient support device
- Appendix A (informative appendix) provides the format of performance values
- Reference
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard uses the redrafting method to amend and adopt IEC 62677:2017 "Medical Electrical Equipment Performance Characteristics of Medical Light Ion Beam Equipment".
The technical differences between this standard and IEC 62677:2017 and the reasons are as follows.
--- Regarding normative reference documents, this standard has made adjustments with technical differences to adapt to my country's technical conditions and adjustments.
The situation is collectively reflected in Chapter 2 "Normative Reference Documents", and the specific adjustments are as follows.
- Replace IEC 60601-1:2005 AMD1.2012 with GB 9706.1 modified to adopt international standards;
- Replace IEC 60580:2000 with GB/T.20012 which is equivalent to adopting international standards;
- Since there are no specific clauses cited in the main body of this standard, IEC 60601-2-1:2009, IEC 60601-2-1:2009/ AMD1.2014.
--- Added terms and definitions "patient support device" (see 3.49).
The Chinese documents that have a consistent correspondence relationship with the normatively cited international documents in this standard are as follows.
---GB/T 18987-2015 radiotherapy equipment coordinates, movement and scale (IEC 61217:2008, IDT) This standard has made the following editorial changes.
--- In order to supplement the explanation that the bed surface includes both the bed surface and the seat surface, to avoid causing ambiguity on the bed surface, 3.40 adds Note 2;
---Position accuracy and position repeatability should be in mm, 6.7.1e) 8), 9) editing errors, modify "cm" to "mm", Consistent with Appendix A;
---Revised the editing error of 6.6 in Appendix A of IEC 62677:2017, and changed "30mm" to "300mm";
--- Deleted the "Terms and Definition Index" in IEC 62677:2017.
Introduction
This standard specifies the performance characteristics and publication methods of radiation therapy light ion beam equipment. In addition to the performance characteristics required to be announced In addition to the indicators, the corresponding test conditions and test methods are also included for the type test of light ion beam equipment. Through the performance characteristics in accordance with this standard The sexual publication method enables direct comparison between the performance data of equipment of different manufacturers.
The test methods specified in this standard are recommended and can be replaced by other equivalent and appropriate methods. However, it should be fully realized that this standard The performance characteristics of light ion beam equipment specified by the standards are closely related to its test methods and test conditions. The tests specified in this standard cannot ensure Any light ion beam equipment meets its claimed performance during its working life. Due to the differences in equipment produced by different manufacturers, the system Manufacturers can add more detailed information and special tests for each performance characteristic in their accompanying documents. This standard stipulates The performance, test methods, and performance announcement methods are as basic and universal as possible. Medical electrical equipment Medical light ion beam equipment Performance characteristics
1 Scope
This standard specifies the performance characteristics and test procedures of medical light ion beam equipment.
This standard applies to light ion beam equipment with a single nuclear energy range of 10MeV~500MeV for the purpose of radiotherapy.
This standard does not apply to ion beam equipment other than light ion.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated reference documents, the latest version (including all amendments) is applicable to this document.
GB 9706.1 Medical electrical equipment Part 1: General requirements for basic safety and basic performance (GB 9706.1-2020,
IEC 60601-1:2012, MOD)
GB/T.20012 Medical electrical equipment dose area product meter (GB/T.20012-2005, IEC 60580:2003, IDT)
IEC 60601-2-64:2014 Medical Electrical Equipment Part 2-64.Basic Safety and Basic Performance of Light Ion Beam Medical Electrical Equipment Special requirements (Medicalelectricalequipment-Part 2-64.Particularrequirementsforthebasicsafetyandes- sentialperformanceoflightionbeammedicalelectricalequipment) IEC TR60788:2004 Medical electrical equipment term definition compilation (Medicalelectricalequipment-Glossaryof definedterms)
IEC 61217:2011 Radiotherapy equipment-coordinates, movement and scale (Radiotherapy equipment-Coordinates, movementsandscales)
3 Terms and definitions
GB 9706.1, GB/T.20012, IEC 60601-2-64:2014, IEC TR60788:2004 and the following terms and definitions Applies to this document.
3.1 Acceptance test
After new equipment is installed, or after a major modification of existing equipment, tests are carried out at the user's site in order to verify compliance with the contract.
Note 1.During or after the acceptance test, immediately collect reference data as the standard for future routine tests.
Note 2.Rewrite IEC TR60788:2004, rm-70-01, modify-add "on user site".
3.2 Accompanyingdocmentation
Materials that accompany medical devices and contain user information, or medical device installation, use, or maintenance, especially information about safe use.
Note 1.The accompanying documents may be operating instructions, technical instructions, installation manuals, quick reference guides, etc.
Note 2.The accompanying documents may include auditory, visual or tactile materials, as well as multimedia-type materials.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 67 pages — is available in the English PDF.
Referenced standards
Normative references
IEC 60601-1:2012 · IEC 60601-2-64:2014 · IEC 61217:2011
Similar standards
How to Download YY/T 1763-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field