Scope

YY/T 1721-2020 is the English-translated version of 游离甲状腺素测定试剂盒.

YY/T 1721-2020 is the Chinese pharmaceutical industry standard for the free thyroxine (FT4) assay kit.

It sets the classification, the requirements, the test methods, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance of the kit, the traceability of the calibrator, the accuracy against reference materials, the linearity, the detection limit, the precision, the interference from related substances and the stability. Issued on 30 June 2020, it has applied since 1 December 2021.

Found what you need?

Document preview — YY/T 1721-2020

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Classification4
  • 4 Requirements5
  • 5 Test methods7
  • 6 Labels and instructions9
  • 7 Packaging, transportation and storage9
  • Bibliography11

1 Scope

This Standard specifies the classification, requirements, test methods, labeling, instructions for use, packaging, transportation and storage of free thyroxine testing kit (hereinafter referred to as FT4 testing kit).

This Standard applies to immunoassay kits for the in vitro quantitative determination of free thyroxine (FT4) using enzyme labeling, (electro) chemiluminescence labeling, (time-resolved) fluorescence labeling, and other methods, using DAS-ELISA or competition method as the principle. This Standard does not apply to:

a) Reagents for qualitative or semi-quantitative determination of free human thyroxine labeled with colloidal gold or other methods (such as test strips, biochips, etc.);

b) Various radioimmunoassay or immunoradioactive testing kits labeled with radioisotopes such as 125I;

c) Free thyroxine calibrators and free thyroxine quality controls intended for sale individually or as part of the FT4 testing kit.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.

GB/T 21415-2008, In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials

GB/T 29791.2, In vitro diagnostic medical devices - Information supplied by the manufacturer (labelling) - Part 2: In vitro diagnostic reagents for professional use

3 Classification

According to different labeling methods, it can be divided into enzyme labeling, (electro) chemiluminescence labeling, (time-resolved) fluorescence labeling, etc. According to Within the linearity range of the FT4 testing kit, test (2 ~ 3) quality-controlled samples of different concentrations and repeat the test 10 times. The coefficient of variation (CV) shall not be greater than 15%.

4 Requirements

4.7 Batch difference

Within the linearity range of the FT4 testing kit, use three batches of testing kits to detect high and low concentration samples, and the inter-batch coefficient of variation between the three batches of testing kits shall be CV <=15%.

4.8 Analytical specificity

Manufacturers shall stipulate the analytical specificity of the FT4 testing kit and clearly explain it in technical materials such as the product manual. The results of the determination of samples like 3,5,3'-triiodothyronine with a concentration not less than 200 ng/mL, 3,3',5'-triiodothyronine of 100 ng/mL and 3,3'-diiodothyronine of 200 ng/mL shall not be higher than the limit value claimed by the FT4 testing kit.

Note: The types and concentrations of potential cross-reactive substances are not limited to the above statements.

4.9 Stability

Any of the following methods can be used for verification:

a) Valid period stability: The manufacturer shall specify the valid period of the reagent (kit). Test the accuracy, detection-limit, linearity and repeatability of the testing kit at the end of the valid period, which shall meet the requirements of 4.3 ~ 4.6;

Note: Generally, when the valid period is 1 year, choose products that are no longer than 1 month; when the valid period is half a year, choose products that are no longer than half a month, and so on. However, if the specified time is exceeded, it is acceptable if the product meets the requirements.

b) Thermal stability test: Place the testing kit within its valid period at 37 °C for a certain period of time, to test the accuracy, detection limit, linearity and repeatability, which shall meet the requirements of 4.3 ~ 4.6.

Note 1: Thermal stability tests cannot be used directly to deduce product valid period unless a derivation formula based on a large amount of stability research data is used.

Note 2: According to product characteristics, any combination of a), b) methods can be selected; but the selected method should be able to verify the stability of the product, so as to ensure that the product performance within the valid period meets the standard requirements.

5 Test methods

5.1 Appearance

Visually inspect under natural light with normal or corrected vision, which shall meet the requirements of 4.1.

5.2 Traceability

The traceability information provided by the manufacturer shall comply with the requirements of 4.2.

5.3 Linearity

Test the linearity samples specified by the manufacturer. The number of linearity samples should be no less than 5. The low value should be close to the lower limit of the linearity interval, and the high value should be close to the upper limit of the linearity interval. Repeat the measurement for each sample at least twice; take the mean of the measurement results; perform a straight-line fit with the labeled value or theoretical value of the linearity sample to obtain the linearity correlation coefficient r. The result shall meet the requirements of 4.3.

5.4 Detection-limit

Manufacturers shall provide relevant information such as the blank limit, detection limit and reference range of the FT4 testing kit. Based on the information provided by the manufacturer, test 5 low-value samples with concentrations close to the LOD. Test each sample 5 times and sort the test results by size. If the following conditions are met, it can be considered that the blank limit and detection limit provided by the manufacturer are basically reasonable and the results meet the requirements of 4.4.

a) The number of test results that are lower than the blank limit value provided by the manufacturer shall be less than or equal to 3;

b) There is no test result value higher than the lower limit of the reference interval provided by the manufacturer.

5.5 Accuracy

One of the test methods may be selected in the following order of preference:

a) Based on the linearity range of the testing kit provided by the manufacturer, use the reference material that can be used to evaluate the conventional method as the sample, reasonably set 2 ~ 3 concentrations, and test it according to the steps in the testing kit instructions. Repeat the test for each sample 3 times, and record the test results as (Xi). Calculate the relative deviation (Bi) according to Formula (1). If all three results meet the requirements of 4.5a),

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1721-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4