Valid

YY/T 1704.1-2020Cervical dilator for single use - Part 1: Progressive dilators (English PDF)

一次性使用宫颈扩张器 第1部分:渐进式

Preview PDF

This is a limited preview

Buy now to download the full PDF (21 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1704.1-2020 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

February 21, 2020

Implementation date

January 1, 2021

Scope

YY/T 1704.1-2020 is the English-translated version of 一次性使用宫颈扩张器 第1部分:渐进式.

YY/T 1704.1-2020 is the Chinese pharmaceutical industry standard for the single-use gradual cervical dilator, the set of tapered dilators used one after another to open the cervix.

It is the first part of the YY/T 1704 series and sets the classification, the requirements, the test methods, the marks, packaging and instruction manual, and the rules for transport, storage and the sterile shelf life, covering the dimensions and the taper of each size, the surface finish, the strength and stiffness of the dilator, the materials and their biological evaluation, and sterility. It was issued on 21 February 2020 and has applied since 1 January 2021.

Found what you need?

Document preview — YY/T 1704.1-2020

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Classification4
  • 4 Requirements6
  • 5 Test methods7
  • 6 Marks, packaging and instruction manual9
  • 7 Transport, storage9

1 Scope

This Part of YY/T 1704 specifies the classification, requirements, test methods, marks, packaging and instruction manual, transport, storage and sterilization period for gradual cervical dilator for single use (hereinafter referred to as the dilator).

This Part is applicable to cervical dilator for single use. This product is used for dilatation of the cervix in obstetrics and gynecology and family planning departments.

This Part is not applicable to the cervical dilator made of metal.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 14233.1-2008, Test methods for infusion transfusion injection equipment for medical use - Part 1: Chemical analysis methods

GB/T 16886.1, Biological evaluation of medical devices - Part 1: Evaluation and testing

YY/T 0171, Surgical instruments - Packaging, Marking and Instructions Pharmacopoeia of the People's Republic of China (2015 Edition) Part IV

3 Classification

3.1 Type

3.1.1 The dilator is the expansion section within 60mm of the ring line from the head end. It is usually composed of 3 gradual units. Each unit is about 20mm long.

4 Requirements

4.1 Appearance

4.1.1 The head of the dilator shall be round and smooth, and no fibers shall be pulled out.

4.1.2 The marks of the dilator shall be clear and obvious.

4.1.3 The overall dilator shall be smooth and uniform in color, and there shall be no cracks, burrs, plastic flow, or defects.

4.2 Dimensions

The gradual diameter range d1 and d2 of the dilator, the tilt angle theta of the tip shall meet the requirements of Table 1.

4.3 Physical properties

4.3.1 The dilator shall be able to withstand 50N axial tension without breaking.

4.3.2 Apply a force of 15N to the head of the dilator, and there shall be no obvious deformation or fracture.

NOTE: Obvious deformation can be understood as the tip tilt angle theta<=0° after applying force.

4.4 Chemical properties

4.4.1 Heavy metals

The total content of heavy metals in the dilator test solution shall not exceed 1µg/mL.

4.4.2 PH

Compare the dilator test solution with the blank solution. The difference in pH shall not exceed 1.5.

4.4.3 UV absorbance

The absorbance of the dilator test solution in the wavelength range of 250nm~320nm shall not exceed 0.1.

4.5 Sterilization

The dilator shall be sterilized by a confirmed sterilization process, and the sterilized dilator shall be sterile.

NOTE: N is the loading force.

Figure 2 -- Schematic diagram of bending performance test of the dilator

5 Test methods

5.4 Chemical properties

5.4.1 Test solution preparation

Take sample. Add water at a ratio: 0.2g of sample to 1mL of water. Keep constant temperature at (37±1)°C for 8h. Separate the sample from water. Cool to room temperature as the test solution. Take the same volume of water and place it in a glass container. Prepare blank control solution in the same way.

5.4.2 Heavy metals

Test according to the method specified in 5.6.1 of GB/T 14233.1-2008, in accordance with the requirements of 4.4.1.

5.4.3 PH

Test according to the method specified in 5.4.1 of GB/T 14233.1-2008, in accordance with the requirements of 4.4.2.

5.4.4 UV absorbance

Test according to the method specified in 5.7 of GB/T 14233.1-2008, in accordance with the requirements of 4.4.3.

5.5 Sterilization

The inspection is carried out in accordance with the method stipulated in the fourth part 1101 sterile inspection method of the "Pharmacopoeia of the People's Republic of China" (Edition 2015), in accordance with the requirements of 4.5.

NOTE: Each sterilization batch undergoes an effective monitoring process to make the product sterile.

5.6 Residual ethylene oxide

Test according to the method in Clause 9 of GB/T 14233.1-2008, in accordance with the requirements of 4.6.

5.7 Biological evaluation

Evaluate according to the corresponding method of biological evaluation, in accordance with the requirements of 4.7.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1704.1-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
21 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1704.1-2020

$105.00