YY/T 1551.3-2017Air filters for medical infusion and transfusion equipments - Part 3: Integrity test methods (English PDF)
输液、输血器具用空气过滤器 第3部分:完整性试验方法
This is a limited preview
Buy now to download the full PDF (22 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1551.3-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
February 28, 2017
Implementation date
January 1, 2018
Scope
YY/T 1551.3-2017 is the English-translated version of 输液、输血器具用空气过滤器 第3部分:完整性试验方法.
YY/T 1551.3-2017 is the Chinese pharmaceutical industry standard that gives the integrity tests for the air filter of an infusion or transfusion set: the bubble point pressure test and the water breakthrough or penetration test.
It is the third part of the YY/T 1551 series and sets out the apparatus, the wetting of the membrane, the way the pressure is raised and the point read, and the interpretation, so that a non-destructive test can stand for the retention of the filter. Issued on 28 February 2017, it has applied since 1 January 2018.
Found what you need?
Document preview — YY/T 1551.3-2017
National Standard of the People's Republic of China
- ICS
- 11.040.20
- Classification
- C 31
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Terms and definitions
- 3 State conditioning and test environment
- 4 Bubble point pressure test
Foreword
YY/T 1551 "infusion, blood transfusion device with air filter" consists of the following components.
--- Part 1. Aerosol bacterial retention test method;
--- Part 2. Liquid bacterial retention test method;
--- Part 3. Integrity test methods. This part of YY/T 1551 Part 3. This section drafted in accordance with GB/T 1.1-2009 given rules.
Introduction
For direct confirmation of infusion, blood transfusion devices with air filters to prevent the environment into the transfusion of microorganisms, blood transfusion system, a common method Liquid bacterial retention test and aerosol bacterial retention test two. The former is the first air filter in the filter by a special treatment, and then press Similar to the bacteriostasis-type liquid filter bacterial retention test method with a diameter of 0.3µm ~ 0.4µm, length 0.6µm ~ 1.0µm defects Pseudomonas bacteria challenge test, the specific test methods see the second part of this standard; the latter is the average particle diameter of about 3µm microbial aerosol simulation of clinical conditions of use challenge test, the specific test methods, see YY/T 1551 Part 1.
Air filters that pass the aerosol bacterial entrapment test are thought to effectively filter out airborne microbial aerosols. For nominal membrane pores The air filter with a diameter of 0.22 µm must meet the more stringent bacterial entrapment test requirements.
As the bacterial retention test method is more complicated, and the requirements of the test conditions and personnel to operate high, the general does not apply to the air filter Routine control test. Conventional control tests may use non-destructive physical integrity tests that have established correlation with the bacterial retention test Methods, such as bubble point pressure method. The air filter physical integrity test carried out on the premise that the air filter can withstand the corresponding bacteria interception test A correlation filter membrane bubble pressure was established. Bubble-point pressure, established on the basis of studies related to bacterial entrapment tests, can be used as infusion, transfusion Air bubble filter with integrity check bubble pressure. If the air filter integrity check bubble point pressure is lower than the membrane bubble point pressure, still with fine Bacteria retention test to establish correlation.
At present, infusion, blood transfusion devices with air filters are mainly air intake devices, liquid filter on the vent and other infusion, blood transfusion devices On the exhaust and other types. This standard mainly to the intake device, liquid filter on the vent as an example to establish air filter integrity experiment method.
1 Scope
YI/T 1551 provisions of this part of the infusion, blood transfusion device with air filter integrity test methods, including bubble point pressure test and water Breakthrough/penetration test.
2 Terms and definitions
The following terms and definitions apply to this document.
2.1 Outside
The air filter of the infusion set is exposed to the side of the air environment when in use.
2.2 Inside
The other side of the infusion set's air filter, opposite the outside.
3 State conditioning and test environment
The sample should be conditioned for 24 h at (23 ± 2) ° C ambient temperature and (50 ± 5)% relative humidity and conditioned under conditions of conditioning test.
NOTE Sterilization of the filter will have some effect on the integrity of the filter, and the integrity test should be performed on the sterilized product.
4 Bubble point pressure test
4.1 Principle
Air filters wetted with a suitable wetting fluid (see Appendix A), the filter membrane of which is wetted, immersed in a test medium (water or other moistened Wet liquid), a predetermined gas pressure (bubble point depression limit value) is applied from the inside thereof, and whether or not an air flow passes is observed from the outside.
4.2 Wetting
Apply a suitable amount of a suitable wetting fluid (see A.1.1) to both sides of the air filter with a syringe to allow the filter membrane to fully wet.
NOTE When using water as the bubble point pressure test medium, a syringe pre-filled with purified water is used to inject a suitable amount of air from the inside of the air filter Purified water Rinse the filter lumen and filter membrane (over the membrane rinse) to fully remove the wetting fluid. See A.1.2.
4.3 Test layout
The air filter after 4.2 wetting arrangement test system according to FIG. The cavity volume of the test system should be as small as possible to stabilize the pressure. Also avoid pipe twisting in the test system.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 22 pages — is available in the English PDF.
Similar standards
How to Download YY/T 1551.3-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 1551.2-2017 — Air filters for medical infusion and transfusion equipments - Part 2: Liquid bacterial retention test method
YY/T 1551.1-2017 — Air filters for medical infusion and transfusion equipments - Part 1: Aerosol bacterial retention test method
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
Secure payment via Stripe
Payments accepted
YY/T 1551.3-2017
$145.00