YY/T 1523-2017Bicarbonate assay kit(PEPC enzymatic method) (English PDF)
二氧化碳测定试剂盒(PEPC酶法)
This is a limited preview
Buy now to download the full PDF (19 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1523-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 28, 2017
Implementation date
April 1, 2018
Scope
YY/T 1523-2017 is the English-translated version of 二氧化碳测定试剂盒(PEPC酶法).
YY/T 1523-2017 is the Chinese pharmaceutical industry standard for the bicarbonate assay kit based on the PEPC enzymatic method.
It sets the requirements, the test methods, the label and instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, analytical sensitivity, linearity, precision, accuracy and stability.
It applies to kits that measure bicarbonate quantitatively in human samples. Issued on 28 March 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 1523-2017
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Requirements4
- 4 Test Methods6
- 5 Label and Use Instructions9
- 6 Package, Transportation and Storage10
- Bibliography11
1 Scope
This Standard specifies the requirements, test methods, label and use instructions, package, transportation and storage of bicarbonate assay kit.
This Standard is applicable to the reagent kit performing quantitative detection (hereinafter referred to as kit) against the carbon dioxide in the human serum or plasma by phosphoenolpyruvate carboxylase (PEPC) enzymatic method, including the reagent used by manual, semi-automatic, fully automatic biochemistry analyzer.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.
GB/T 29791.2 In Vitro Diagnostic Medical Devices - Information Supplied by the Manufacturer (Labelling) - Part 2. In Vitro Diagnostic Reagents for Professional
3 Requirements
3.1 Appearance
The appearance of kit shall meet the following requirements.
a) The components of kit shall be complete and intact; there is no liquid leakage;
b) The text and symbols of package label shall be clear.
3.2 Loading amount
It shall be no less than the marked value.
3.3 Reagent blank absorbance
certain period of time after the expiration date; test the reagent blank absorbance, analytical sensitivity, linear, repeatability, accuracy, which shall meet the requirements of 3.3, 3.4, 3.5, 3.6.1, 3.7.
3.8.3 Thermal stability test
Test the reagent blank absorbance, analytical sensitivity, linear, repeatability, accuracy, which shall meet the requirements of 3.3, 3.4, 3.5, 3.6.1, 3.7.
NOTE 1. Thermal stability cannot be used to derive the valid date of the product; unless it is from a derivation formula based on a large number of stability studies.
NOTE 2. Generally, a product with a valid period of 1 year and no more than 1 month is selected; the valid period shall be half a year; and so on. However, if it exceeds the specified time, when the product meets the requirements, it is also acceptable.
NOTE 3. According to the product characteristics, any combination of the methods specified in 3.8.2, 3.8.3 can be selected; but the selected method shall be able to verify the stability of the product to ensure that the product performance meets the standard requirements with the shelf life period.
4 Test Methods
4.1 Basic requirements of instrument
Spectrophotometer or biochemical analyzer, the wavelength range shall meet the requirements for reagent use; biochemical analyzer shall be equipped with thermostat (temperature value shall be within ±0.3°C of the setting value; fluctuation shall be no greater than ±0.2°C); the absorbance measurement resolution shall be 0.001 above.
4.2 Appearance
Take visual examination with normal or corrected visual acuity under natural light, the results shall meet the requirements of 3.1.
4.3 Loading amount
Use general measuring gauge to measure; it shall meet the requirements of 3.2.
4.4 Reagent blank absorbance
Use reagent kit to test the blank sample; record the absorbance value at the dominant wavelength of the reading point specified in the reagent kit parameters; which shall meet the requirements of 3.3.
NOTE. The blank sample can be pure water sample, saline, zero calibration solution, and the Xi - measurement concentration; T - calibrated concentration of reference material.
4.8.3 Recovery test
Add certain volume of standard solution or pure product to human sample (the volume ratio between standard solution and human sample shall not generate matrix change; after adding the standard solution, the total concentration of the sample shall be within the linear range of the kit); each concentration shall be test for 3 times; calculate the average value; calculate the recovery rate as per the Formula (5), which shall meet the requirements of 3.7b). Where. R - recovery rate; V - volume of the added standard solution; V0 - volume of human sample;
C - mean concentration of detection after human sample is added standard solution; C0 - mean concentration of detection of human sample; Cs - concentration of standard solution.
4.9 Stability
4.9.1 Shelf life stability
Take the sample after the expiration date, test as per the method specified in 4.4, 4.5, 4.6, 4.7.1, 4.8, which shall meet the requirements of 3.8.2.
4.9.2 Thermal stability test
Take the sample within the shelf life of product, according to the manufacturer claimed thermal stability conditions, test as per the methods specified in 4.4, 4.5, 4.6, 4.7.1, 4.8, which shall meet the requirements of 3.8.3.
5 Label and Use Instructions
It shall meet the requirements of GB/T 29791.2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1523-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 29791.2-2013 — In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 2: In vitro diagnostic reagents for professional use
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
Secure payment via Stripe
Payments accepted
YY/T 1523-2017
$115.00