YY/T 1510-2017Medical plasma virus inactivation cabinet (English PDF)
医用血浆病毒灭活箱
This is a limited preview
Buy now to download the full PDF (21 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1510-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 28, 2017
Implementation date
April 1, 2018
Scope
YY/T 1510-2017 is the English-translated version of 医用血浆病毒灭活箱.
YY/T 1510-2017 is the Chinese pharmaceutical industry standard for the medical plasma virus inactivation cabinet, the equipment in which plasma is treated, typically with methylene blue and light, to inactivate viruses before transfusion.
It sets the general requirements, the test methods, the inspection rules, the marking, the instructions for use and the packaging, covering the illumination and its uniformity over the plasma bag, the temperature control, the timer, the inactivation performance, the alarms and the electrical safety. Issued on 28 March 2017, it has applied since 1 April 2018.
Found what you need?
Document preview — YY/T 1510-2017
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction4
- 1 Scope5
- 2 Normative References5
- 3 Terms and Definitions5
- 4 Requirements6
- 5 Test Methods7
- 6 Inspection Rules8
- 7 Marking, Instruction for Use9
- 8 Packaging10
Introduction
This Standard provides basic technical guidance to medical plasma virus inactivated device of different light sources, design structures and specification models.
Medical plasma virus inactivated device is a device of preparing clinically infused virus inactivated plasma for blood collection and supply institutions, and clinical medical institutions. It is used cooperatively with methylene blue virus inactivated device.
Medical plasma virus inactivated device is mainly applicable to photochemical method for plasma virus inactivation technology. In other words, after adding an appropriate amount of virus inactivating agent---methylene blue to plasma and placing it in the device, plasma can receive certain amount of illumination at a certain temperature.
Under the illumination, sway the plasma, so that the plasma that contains methylene blue can receive uniform and effective illumination energy, which will reach the effect of virus inactivation. Medical plasma virus inactivated device is an exclusive supporting device that guarantees safe, effective and reliable plasma virus inactivation through the photochemical method.
Medical plasma virus inactivated device provides appropriate illumination conditions to plasma virus inactivation through the photochemical method. Effective illumination conditions for virus inactivation depend on the variety of light source and the intensity of illumination that is emitted from a specific waveband, and illumination energy that plasma receives. There are two commonly seen light sources of virus inactivated device in the market. florescence and LED, and the duration of illumination also varies with the intensity of illumination. In consideration of these, this Standard specifies that illumination parameters stipulated by manufacturers shall be satisfied in tests with methods provided by the manufacturers. However, satisfying these parameters does not signify that it can reach an ideal effect of virus inactivation. Manufacturers shall take the responsibility of verifying the validity of these parameters and confirming the effect of virus inactivation in accordance with relevant national-level stipulations.
Furthermore, manufacturers of medical plasma virus inactivated device shall confirm that there is no obvious adverse influence on blood component quality and function and verify infusion safety after virus inactivation. Thus, after virus inactivation, blood component can comply with relevant national-level stipulations. Medical Plasma Virus Inactivated Device
1 Scope
This Standard specifies the general requirements, test methods, inspection rules, marking, instruction for use and packaging of medical plasma virus inactivated device (hereinafter referred to as virus inactivated device).
This Standard is applicable to medical plasma virus inactivated device that is matched with virus inactivated device that is stipulated in YY 0765.1.
2 Normative References
The following documents are indispensable to the application of this Standard. In terms of references with a specified date, only versions with a specified date are applicable to this Standard. The latest version (including all the modifications) of references without a specified date is also applicable to this Standard.
GB 4793.1 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1. General Requirements
GB/T 14710 Environmental Requirement and Test Methods for Medical Electrical Equipment
GB/T 18268.1 Electrical Equipment for Measurement, Control and Laboratory Use - EMC Requirements - Part 1. General Requirements
YY/T 0466.1 Medical Devices - Symbols to be Used with Medical Device Labels, Labelling and Information to be Supplied - Part 1. General Requirements
YY 0765.1 Sets for Inactivation of Viruses in Blood and Blood Components for Single Use. Part 1. Sets for Virus Photodynamic Inactivation with Methylene Blue
3 Terms and Definitions
The following terms and definitions are applicable to this Standard.
3.1 Stable Operating State
Stable operating state means after virus inactivated device is activated, indicated temperature reaches 2 °C ~ 8 °C; illumination intensity reaches the range explicitly instructed by manufacturers.
5 Test Methods
5.3 Illumination Intensity
Conduct the test in accordance with methods stipulated by manufacturers.
5.4 Smoothness and Reliability of Load Shelf
Under the maximum load, conduct visual inspection of whether the load shelf in the virus inactivated device can operate smoothly and reliably; whether the extraction is flexible; whether there are obvious sway and distortion.
5.5 Alarm Temperature
When the practical simulated temperature and illumination intensity exceed the limitation, use sound level meter to measure alarm sound level at a distance of 1 m from the front panel and 1 m from the ground.
5.6 Time Setting
Use a timer to measure the practical operating hours of virus inactivated device; calculate the deviation from the previously set time.
6 Inspection Rules
6.1 Inspection Type
Inspection of virus inactivated device shall be divided into exit-factory inspection and model inspection.
6.2 Exit-factory Inspection
Exit-factory inspection items of each virus inactivated device (including process inspection and/or ultimate inspection) shall at least include 4.2 ~ 4.9.
6.3 Model Inspection
6.3.1 Model inspection shall be conducted under any of the following circumstances.
a) When new products are put into production;
b) When there are significant changes in structure, critical components and technology;
c) When production is resumed after production suspension and rectification;
d) When there are significant differences between the result of exit-factory inspection result and the last model inspection;
e) When it is stipulated by contract or requested by management department.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 4793.1Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 18268.1Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
- YY 0765.1Single-use materials for virus inactivation of blood and blood components - Part 1: Methylene blue virus inactivation materials
Similar standards
How to Download YY/T 1510-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 18268.1-2025 — Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements