YY/T 1485-2016Dentistry—Dental implant unit (English PDF)
牙科学牙科种植机
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1485-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
July 29, 2016
Implementation date
June 1, 2017
Scope
YY/T 1485-2016 is the English-translated version of 牙科学牙科种植机.
YY/T 1485-2016 is the Chinese pharmaceutical industry standard for the dental implant unit, the motor and control unit used with a handpiece to prepare the site and place a dental implant.
It specifies the requirements, the test methods and the instructions for use, covering the speed and the torque delivered and their accuracy, the reduction handpiece and the coupling, the irrigation of the site, the programmes and the display, the sterilization of the parts that are sterilized and the electrical safety. Issued on 29 July 2016, it has applied since 1 June 2017.
Found what you need?
Document preview — YY/T 1485-2016
National Standard of the People's Republic of China
- Classification
- C 33
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative reference documents
- 3 Terms and definitions
- 4 Requirements
Foreword
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
Introduction
The dental implants specified in this standard are intended to provide a source of power for mobile phones used for implant treatment.
The dental implants specified in this standard may be used in conjunction with mobile phones.
1 Scope
This standard specifies the requirements for dental implants, test methods and instructions.
This standard applies to dental implants (as defined in 3.1). The dental implants are used in conjunction with mobile phones for dental implantation for dental implantation treatment.
This standard does not cover the requirements of mobile phones used for implant treatment.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
GB 4208-2008 enclosure protection class (IP code) (IEC 60529.:2001, IDT)
GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety (GB 9706.1-2007, IEC 60601-1. 1988, IDT)
GB/T 9937 (all parts) Oral vocabulary [ISO 1942 (all parts)] Graphical symbols for dental equipment (ISO 9687.:1993, IDT) YY/T 0628-2008
YY 1012-2004 Dental phone coupling size (ISO 3964. 1982, IDT)
YY 1045.2 Dental handsets Part 2. Straight phones and curved handsets (YY 1045.2-2010, ISO 7785-2.:1995, IDT) General technical conditions for medical foot switch YY 1057-2016
YY/T 1400-2016 Dentistry Dental equipment Surface materials - Determination of chemical disinfectants (ISO 21530.:2004, IDT)
3 Terms and definitions
GB 9706.1, GB/T 9937 Definitions and the following terms and definitions apply to this document.
3.1 Dental implant machine
dentalimplantunit A device that provides driving force for mobile phones used for implantation therapy, consisting of at least a host, a planter motor, and a foot switch.
4 Requirements
4.1 Operation control device
4.1.1 Overview
The operation control device should be designed and positioned to prevent its accident from starting.
Operation control devices and graphic symbols indicating performance shall comply with the requirements of YY/T 0628-2008.
4.1.2 Speed
The dental plant should have a speed setting function.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 9937Dentistry - Vocabulary
- YY 1012-2004Dental handpieces - Coupling dimensions
- YY 1045.2Dental handpieces - Part 2: Straight and geared angle handpieces
- YY/T 1400-2016Dentistry - Materials used for dental equipment surfaces - Determination of resistance to chemical disinfectants
GB 4208-2008
Similar standards
How to Download YY/T 1485-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB/T 9937.2-2008 — Dental vocabulary -- Part 2: Dental materials
GB/T 9937-2020 — Dentistry - Vocabulary
Secure payment via Stripe
Payments accepted
YY/T 1485-2016
$85.00