YY/T 1484-2016Ophthalmic instruments—Instruments to measure axial distances in the eye (English PDF)

眼科仪器眼轴长测量仪

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1484-2016 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

July 29, 2016

Implementation date

June 1, 2017

Scope

YY/T 1484-2016 is the English-translated version of 眼科仪器眼轴长测量仪.

YY/T 1484-2016 is the Chinese pharmaceutical industry standard for ophthalmic instruments that measure axial distances in the eye, above all the axial length used in calculating an intraocular lens.

It sets the requirements the instrument must meet, the test methods, the accompanying documents and the marking, with a normative appendix giving the test body and how a measurement is evaluated.

It is a modified adoption of ISO 22665:2012: the requirements for ultrasonic axial length instruments are removed, so that the standard covers optical instruments, GB 9706.1 replaces IEC 60601-1 as the safety reference, and requirements and test methods for the repeatability of the axial length measurement are added. Issued on 29 July 2016, it has applied since 1 June 2017.

Found what you need?

Document preview — YY/T 1484-2016

National Standard of the People's Republic of China

Classification
C 40

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Preface
  • 1 Scope
  • 2 normative reference document
  • 3 Terms and definitions
  • 4 Requirements
  • 5 Test Method
  • 6 random file
  • 7 logo
  • Appendix A (Normative Appendix) Test Body and Measurement Assessment

Foreword

This standard is drafted in accordance with the rules given in GB/T 1.1-2009.

This standard uses the re-drafting method to modify the international standard ISO 22665.:2012 "ophthalmic instrument eye axis length meter" (English version).

There are technical differences between this standard and ISO 22665.:2012, and the terms involved in these differences have passed through the margin The vertical single line (|) is marked. The main technical differences between this standard and ISO 22665.:2012 are as follows.

--- on the normative reference documents, this standard has made a technical differences in the adjustment to meet China's technical conditions, adjust the situation The situation is reflected in Chapter 2, "normative reference documents", the specific adjustments are. equivalent to the use of international standards GB 9706.1 generation For IEC 60601-1;

--- Removes the requirements of ISO 22665 for the use of ultrasonic methods for axial length measuring instruments. It is clear in the scope of application that this standard applies to Optical method of axial length measuring instrument;

--- supplemented the axial length measurement repeatability requirements and test methods.

1 Scope

This standard applies to the use of optical methods to measure the length of the instrument.

This standard specifies the minimum requirements and test methods for the above instruments and systems.

2 Normative reference documents

The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.

GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety (GB 9706.1-2007, IEC 60601-1. 1988, IDT)

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Axial length

Along the axial axis, the distance between the anterior surface of the cornea and the retinal membrane (ILM) or retinal pigment epithelial cell layer (RPE).

Note. The distance between the anterior retinal (ILM) and the retina (RPE) is approximately 100 µm in the central depression and about 300 µm at the center of the fovea.

There are different ways of measuring the axial length, for example, in the ultrasound method, the distance between the anterior corneal surface and the retinal membrane (ILM) is measured;

Method, the distance between the anterior surface of the cornea and the retinal pigment epithelial cell layer (RPE) was measured.

3.2 Aphakicmode

A measuring mode and/or instrument setting for an axial length measuring instrument, which is used to measure aphakic eye (lens missing).

3.3 Group refraction rate

The ratio of the propagation rate (c0) of the light in the vacuum to the propagation rate (cg) of the light in the medium or the biological tissue is c0/cg.

3.4 Optical biometric measurement

The length of the human eye axis is measured optically.

3.5 Have a lens pattern phakicmode

A measuring mode and/or instrument setting for an axial length measuring instrument, which is used to measure the presence of a lens (lens).

4 Requirements

4.1 axial length measurement tolerance. ± 100µm.

4.2 axial length measurement repeatability. <= 33µm.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1484-2016

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4