Scope
YY/T 1450-2016 is the English-translated version of 载脂蛋白A-I测定试剂(盒).
YY/T 1450-2016 is the Chinese pharmaceutical industry standard for the apolipoprotein A-I test reagent.
It specifies the requirements, the test methods, the labels, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the traceability of the calibrator, the accuracy, the linear range, the detection limit, the precision and the stability.
It applies to kits that determine apolipoprotein A-I quantitatively in human serum or plasma by immunoturbidimetry read in transmission. Issued on 26 January 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 1450-2016
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative reference documents
- 3 Request
Foreword
This standard is drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard specifies the requirements, test methods, labeling, labeling and instructions for use of apolipoprotein AI assays (cartons), packaging, transport And storage requirements.
This standard is applicable to the use of immunoturbidimetry (transmission method) on human serum or plasma apolipoprotein AI in the quantitative detection of reagents (Box), including manual and semi-automatic, automatic biochemical analyzer used on the reagent.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
YY/T 0466.1 Symbols for medical devices for labeling, labeling and providing information on medical devices - Part 1. General requirements
3 Request
3.1 Appearance
Should specify the normal appearance requirements.
3.2 Capacity
Should not be less than the marked value.
3.3 Reagent Blank Absorbance
Should specify the reagent blank absorbance requirements.
3.4 Analysis of sensitivity
Test the concentration of 1.00g/L of the sample, the absorbance difference (deltaA) should be not less than 0.10.
Note: The test wavelength is 340nm, other test wavelengths or through the sample pre-dilution test can be adjusted accordingly.
3.5 Linearity
The linear range of reagent (box) should be covered [0.40,2.20] g/L.
a) The linear correlation coefficient (r) should be not less than 0.990;
b) [0.40,2.20] g/L interval, the linear deviation should not exceed ± 10%.
3.6 Precision
3.6.1 Repeatability
Repeated test (1.00 ± 0.50) g/L of human serum or control substance, the results of the coefficient of variation CV (%) should not be greater than 3%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1450-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0466.1-2009 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2016 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2023 — Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
YY/T 1450-2016
$139.00