YY/T 1430-2016Standard test method for small punch testing of ultra-high molecular weight polyethylene used in surgical implants (English PDF)
外科植入物用超高分子量聚乙烯小冲孔试验方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1430-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
January 26, 2016
Implementation date
January 1, 2017
Scope
YY/T 1430-2016 is the English-translated version of 外科植入物用超高分子量聚乙烯小冲孔试验方法.
YY/T 1430-2016 is the Chinese standard "Standard test method for small punch testing of ultra-high molecular weight polyethylene used in surgical implants".
Its scope clause reads: This standard is a small disc test (thickness 0.5mm, diameter 6.4mm) small punching test method to determine the ultra-high molecular The mechanical properties of polyethylene (UHMWPE).
This standard applies to characterization of plunger extrusion or compression molding after UHMWPE surgery The properties of the material are also suitable for the evaluation of radiation cross-linked and sterilized after the same process to manufacture the implant, but also for the recovery of the human body (Taken from the body) of the corresponding implants were tested.
The small punch test parameters, that is, peak load, ultimate displacement, ultimate load and loss of work, can provide UHMWPE in the multi-axis Measurement of yield strength, ultimate strength, ductility and toughness under loading conditions.
Due to UHMWPE under uniaxial and multi-axis loading conditions Mechanical properties are not the same, small punching test GB/T 19701.2 Chinese medicine grade UHMWPE uniaxial tensile test conducted a mechanical test Test technology.
In addition to the study tools for implantation analysis, the small punch test can also be used in laboratory screening tests to evaluate the new UH- MWPE material, such as gamma irradiation or electron beam irradiation.
The test method is also applicable to accelerated aging of UHMWPE materials (Eg, ASTMF2003), and can specify the mechanical properties of different UHMWPE samples after oxidative aging Down grade.
Small punch test can be used for other polymers, including polymethyl methacrylate (PMMA) bone cement, polyacetal and high density polyethylene (HDPE). However, other polymer non-UHMWPE materials, small punching test is not within the scope of this standard test method.
International units (SI) are used in this standard. This standard does not attempt to address all the security issues involved, even those related to its use. establish Appropriate safety and health practices, as well as the applicability of regulatory restrictions before application, are the responsibility of the user of this standard. Its clauses include normative reference documents; terms and definitions.
Found what you need?
Document preview — YY/T 1430-2016
National Standard of the People's Republic of China
- ICS
- 11.040.40
- Classification
- C 35
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- 1 Scope
- 2 normative reference document
- 3 Terms and definitions
- 4 meaning and application
- 5 equipment
- 6 Test sample
- 8 test speed
- 9 Condition
- 10 Procedure
- 11 Calculate
- 12 Report
1 Scope
This standard is a small disc test (thickness 0.5mm, diameter 6.4mm) small punching test method to determine the ultra-high molecular
The mechanical properties of polyethylene (UHMWPE). This standard applies to characterization of plunger extrusion or compression molding after UHMWPE surgery
The properties of the material are also suitable for the evaluation of radiation cross-linked and sterilized after the same process to manufacture the implant, but also for the recovery of the human body (Taken from the body) of the corresponding implants were tested.
The small punch test parameters, that is, peak load, ultimate displacement, ultimate load and loss of work, can provide UHMWPE in the multi-axis
Measurement of yield strength, ultimate strength, ductility and toughness under loading conditions. Due to UHMWPE under uniaxial and multi-axis loading conditions
Mechanical properties are not the same, small punching test GB/T 19701.2 Chinese medicine grade UHMWPE uniaxial tensile test conducted a mechanical test
Test technology.
In addition to the study tools for implantation analysis, the small punch test can also be used in laboratory screening tests to evaluate the new UH-
MWPE material, such as gamma irradiation or electron beam irradiation. The test method is also applicable to accelerated aging of UHMWPE materials (Eg, ASTMF2003), and can specify the mechanical properties of different UHMWPE samples after oxidative aging
Down grade.
Small punch test can be used for other polymers, including polymethyl methacrylate (PMMA) bone cement, polyacetal and high density polyethylene (HDPE). However, other polymer non-UHMWPE materials, small punching test is not within the scope of this standard test method.
International units (SI) are used in this standard.
This standard does not attempt to address all the security issues involved, even those related to its use. establish
Appropriate safety and health practices, as well as the applicability of regulatory restrictions before application, are the responsibility of the user of this standard.
2 normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this article
Pieces. For undated references, the latest edition (including all modifications) applies to this document.
GB/T 16825.1 Tests for static uniaxial testing machines - Part 1. Testing and testing of force systems for tension and/or pressure testing machines (GB/T 16825.1-2008, ISO 7500-1..2004, IDT)
GB/T 19701.2 Surgical implants - Ultrahigh molecular weight polyethylene - Part 2. Molded plastics (GB/T 19701.2-2005,
ISO 5834-2..1998, IDT)
YY/T 0651.1 Surgical implants for wear of total hip prostheses Part 1. Load and displacement parameters for wear testing machines and
Related test environmental conditions (YY/T 0651.1-2016, ISO 14242-1..2012, IDT)
YY/T 1426.1 Surgical implants of total knee prosthesis wear - Part 1. Load - controlled wear test machine load and
Displacement parameters and associated test environmental conditions (YY/T 1426.1-2016, ISO 14243-1..2009, IDT)
3 terms and definitions
The following terms and definitions apply to this document.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 16825.1Metallic materials - Calibration and verification of static uniaxial testing machines - Part 1: Tension/compression testing machines - Calibration and verification of the force-measuring system
- GB/T 19701.2Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 2: Moulded forms
- YY/T 0651.1Implants for surgery—Wear of total hip-joint prostheses - Part 1: Loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test
- YY/T 1426.1Implants for surgery - Wear of total knee-joint prostheses - Part 1: Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test
Similar standards
How to Download YY/T 1430-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 16825.1-2022 — Metallic materials - Calibration and verification of static uniaxial testing machines - Part 1: Tension/compression testing machines - Calibration and verification of the force-measuring system
GB/T 19701.2-2024 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 2: Moulded forms
YY/T 0651.1-2008 — Implant for surgery—Wear of total hip-joint prostheses - Part 1: Loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test
Secure payment via Stripe
Payments accepted
YY/T 1430-2016
$135.00