YY/T 1304.2-2015Detection system of time-resolved fluoroimmunoassay - Part 2: Quantitative reagent(kit) for time-resolved fluoroimmunoassay (English PDF)
时间分辨荧光免疫检测系统 第2部分:时间分辨荧光免疫分析定量测定试剂(盒)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1304.2-2015 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 2, 2015
Implementation date
January 1, 2016
Scope
YY/T 1304.2-2015 is the English-translated version of 时间分辨荧光免疫检测系统 第2部分:时间分辨荧光免疫分析定量测定试剂(盒).
YY/T 1304.2-2015 is the Chinese pharmaceutical industry standard for the quantitative reagent kit used in time-resolved fluoroimmunoassay.
It is the second part of the YY/T 1304 series and specifies the terms and definitions, the requirements, the test methods, the identification, the labelling, the instruction manual and the rules for packaging, transport and storage, covering the appearance, the traceability of the calibrator, the accuracy, the linear range, the detection limit, the precision and the stability. Issued on 2 March 2015, it has applied since 1 January 2016.
Found what you need?
Document preview — YY/T 1304.2-2015
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Requirements5
- 5 Test method7
- 6 Identifications, labels, user instructions9
- 7 Packaging, transportation and storage11
1 Scope
This part of YY/T 1304 stipulates the terms and definitions, requirements and test methods, identification, labeling and instruction manuals, packaging, transportation and storage of quantitative reagent (kit) for time-resolved fluoroimmunoassay (hereinafter referred to as kits).
This part applies to single-label time-resolved fluoroimmunoassay.
This part does not apply to dual-label and multi-label time-resolved fluoroimmunoassay.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB/T 21415 In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Time-resolved fluoroimmunoassay; TRFIA
The analytical techniques which combine time-resolved fluorescence and immunoassay, use rare earth ions and their chelates as tracers, to label antigens or antibodies which undergo a series of immunological reactions with the analyte, to determine the fluorescence intensity of the final product, thereby obtaining the concentration of the analyte.
Note 2: When there is a certified reference substance (including national reference substance or international reference substance) in the test substance, use the item a) for testing.
Note 3: When there is no certified reference substance (including national reference substance or international reference substance) in the test substance, use the item b) for testing.
4 Requirements
b) Add the corresponding substances of known concentration to the blood matrix or other body fluid components, the recovery rate shall be in the range of 85% ~ 115%.
Note 1: The concentration level of the certified reference material should be close to the product's critical value or medically determined level.
4.7 Specificity
It shall specify the specificity requirements.
4.8 Stability
The following methods can be selected for detection:
a) Expiration date stability: The manufacturer shall specify the expiration date of the kit. Take a batch of kits within a certain period of time after the expiration date, to test the items 4.2 ~ 4.4, which shall meet the corresponding requirements.
b) Thermal stability: After placing a batch of kits within the validity period at 37 °C for 7 days, test the items 4.2 ~ 4.4, which shall meet the corresponding requirements.
Note 1: According to product characteristics, any combination of a) and b) methods can be selected, but the selected method shall be able to verify the stability of the product, to ensure that the product performance meets the corresponding requirements within the valid period.
Note 2: In general, when the validity period of the kit is specified as 12 months, it should select the kit with an expiration period of not more than 1 month for stability test; when the validity period of the kit is specified as 6 months, it should select the kit with an expiration period of not more than half a month for stability test; and so on. However, it may also use the expiration date verified in the stability report provided by the manufacturer.
Note 3: Thermal stability cannot be used to derive the expiration date of the kit, unless a derivation formula based on a large amount of stability study data is used. CV - Coefficient of variation; SD - Standard deviation;
- Average value.
5 Test method
5.5 Differences between batches
Use 3 batches of reagents (kits) to test the same sample, respectively; repeat 10 tests for each. Calculate the average value and the standard deviation SD of the 30 measured concentration values. According to equation (1), obtain the coefficient of variation CV. The results shall meet the requirements of 4.5.
5.6 Accuracy
One of the following methods can be selected for accuracy:
a) Use a certain concentration of certified reference material as a sample.
Perform the testing according to the steps in the instructions. After 3 measurements, the result is recorded as Mn; the relative deviation Bn is calculated according to equation (2). If the three results all meet the requirements, it is judged as qualified. If the result is of two or more than two tests fail to meet requirements, it is judged as unqualified. If one of the results does not meet the requirements, the test shall be repeated 20 times in succession, meanwhile the relative deviation shall be calculated according to equation (2), respectively. If the results of more than or equal to 19 tests comply with requirements, the accuracy shall meet the requirements.
Where: Bn - Relative deviation (n = 1, 2, 3,..., 20); Mn - Measured value (n = 1, 2, 3,..., 20); T - Theoretical value.
b) Add a high level of analyte A with a known concentration to a low concentration of serum B (or other body fluid components). The volume ratio between the added analyte A and serum B (or other body fluid components) is not greater than 1:9; repeat the test 3 times; take the average value; calculate the recovery rate according to formula (3). The result shall meet the above requirements.
a) The name and address of the manufacturer;
b) Product name and specifications;
c) Storage conditions;
d) Production batch number and validity period;
e) Medical device registration certificate number and product standard number.
6 Identifications, labels, user instructions
6.2 User instruction of the kit
It shall include at least the following:
a) Product name;
b) Packaging specifications;
c) Intended use;
d) Applicable instruments;
e) Testing principle;
f) Main components;
g) Storage conditions and expiration date;
h) Sample requirements;
i) Testing method;
j) Interpretation of testing results;
k) Reference value (reference range);
l) Limitations of Testing methods;
m) Product performance indicators;
n) Precautions;
o) Manufacturer's name and address;
p) License number of medical device manufacturer (only applicable to domestic medical device manufacturers);
q) Medical device registration certificate number; […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1304.2-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 21415-2008 — In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
GB/T 21415-2025 — In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples
YY/T 1304.1-2015 — Detection system of time-resolved fluoroimmunoassay - Part 1: Semi-auto time-resolved fluoroimmunoassay analyser
Secure payment via Stripe
Payments accepted
YY/T 1304.2-2015
$115.00