YY/T 1232-2014gamma-Glutamyl transpeptidase test reagent kit (GPNA substrate method) (English PDF)
γ-谷氨酰基转移酶测定试剂(盒)(GPNA底物法)
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1232-2014 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
June 17, 2014
Implementation date
July 1, 2015
Scope
YY/T 1232-2014 is the English-translated version of γ-谷氨酰基转移酶测定试剂(盒)(GPNA底物法).
YY/T 1232-2014 is the Chinese pharmaceutical industry standard for the gamma-glutamyl transpeptidase test reagent kit based on the GPNA substrate method.
It specifies the requirements, the test methods, the labels, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision, the interference and the stability of the reagents. Issued on 17 June 2014, it has applied since 1 July 2015.
Found what you need?
Document preview — YY/T 1232-2014
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Requirement4
- 4 Test Method6
- 5 Labeling and Instructions for Use9
- 6 Packaging, Transport and Storage15
- Bibliography17
Foreword
This Standard is drafted according to the rules specified in GB/T 1.1-2009.
1 Scope
This Standard specifies the technical requirements, test method, labeling, instructions for use, packaging, transport and storage of gamma-Glutamyl transpeptadase test reagent kit (Method of GPNA).
This Standard applies to the gamma-glutamyl transpeptadase test reagent (kit) that is used to carry out quantitative determination of the gamma-glutamyl transpeptadase activity in serum and plasma, with the method of L-gamma glutamyl-3-carboxyl-paranitroaniline. It includes manual reagent and reagents used in semi-automatic and full-automatic biochemical analyzer.
This Standard does not apply to the dry gamma-glutamyl transpeptadase test reagent (kit).
2 Normative References
The following document is indispensable for the application of this Document. For dated references, only dated edition applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB 3100 The international system of units and its application
GB/T 21415-2008 In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials YY /T 0316 Medical devices - Application of risk management to medical devices
YY/T 0466.1 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1. General requirements
YY/T 0638 In Vitro Diagnostic Medical Devices. Measurement of Quantities in Biological Samples. Metrological Traceability of Values Assigned to Enzymatic Concentration of Calibrators and Control Materials
3 Requirement
3.1 Appearance
It shall meet the normal appearance requirements specified by the manufacturer.
e) The microbiological status of the product shall be specified when applicable;
f) Unless the use of the reagent (kit) is easy and obvious; otherwise it shall provide the instructions for use.
- Corresponding notices or symbol shall be present to prompt the user that it shall read the instructions for use carefully before using the reagent (kit);
- The language used in the instructions for use shall be understood by the expected user.
g) It shall focus on prompting the user of the reagent (kit)'s important changes and the position of related information;
h) Words or symbols shall be used to warn the users about the existing hazards and risks. The requirements of YY/T 0316 are applicable;
i) The name, letter, number, symbol, color and graphics of each component of the reagent (kit) shall be marked in the same way;
j) The instructions for use of the reagent may be in the forms such as leaflet that is inserted into packaging; on the surface of the outer container; in the user's manual; or integrated into the operation manual with the instructions for use of instruments or analysis system;
k) The instructions for use of the reagent may be in electronic version; li) Part of the instructions for use may be provided in coded form, and explained in the operation manual of the system;
m) If the detailed instructions for use is not enclosed with the reagent (kit), the manufacturer shall ensure that the users can get the correct version of the instructions for use;
n) The labels of external and internal packaging shall include the specified information. The language and/or symbols used shall be easy to understand; Example: Printing quality, font and font size.
o) The marking, labeling and instructions for use accompanying with the reagent (kit) shall include at least the before-use safe handling and storage.
5 Labeling and Instructions for Use
5.2 Marking and labels of external packaging
The marking and labels of external packaging shall include the following information.
a) Manufacturer's name and address;
Note 1. the address may include the province, city, county and town, when applicable;
Note 2.
a) Manufacturer's name and full address or contact information;
b) Marking and reagent (kit) name. Additional marking information shall be provided if the reagent (kit) can not be identified uniquely based on the name;
c) Intended use. The intended use (including measured amount) and use restrictions shall be described. The medical indications shall also be described, when applicable;
d) Principle of measurement procedure. The principle of measurement procedure shall be specified, including the reaction type (such as chemical, microbial, or immunochemical reaction), indicator or monitoring system, and/or other appropriate details;
e) Traceability of calibrators and accuracy control materials.
- If applicable, metrological traceability of values assigned to calibrators and control materials shall be specified, including the available reference materials and/or higher level of reference measurement procedures;
Note: GB/T 21415-2008 specifies the traceability requirements for the reference materials and/or reference measurement procedures.
- Relevant scientific document or other available reference measurement procedures or reference material document shall be provided, when applicable.
f) Component. The nature and amount or concentration of reactive components shall be provided. Relevant information of other components affecting measurement procedures shall be provided;
g) Additional equipment required.
- All the special equipment that are not listed by the manufacturer but that are required to ensure the safety and effectiveness of reagent (kit) shall be provided;
- Marking information of these equipment and connection methods for normal use shall be provided.
h) Reagent preparation. All preparation steps of the reagent shall be described;
i) Storage and service life after the first unpacking.
- If the storage conditions and period of validity after the first packing are different from those which shall be given by the reagent label, they shall be stated in the instructions for use;
- The storage conditions and stability of the working reagent shall be given, when applicable.
j) Warning and preventive measures.
o) Explanation of results.
- The acceptance and rejection criteria of the test results shall be stated. For instance, if a special result is obtained, then it shall indicate whether an additional test is required;
- The meaning of the test results obtained shall be explained.
p) Performance characteristics.
- The characteristics of analysis performance related to the intended use shall be described;
- The characteristics of diagnostic performance related to the intended use shall be described.
q) If applicable, descriptions of the reference interval, reference population, and related references shall be provided;
r) The restrictions of inspection procedure shall be specified, including.
- Information of related clinical known interfering substances;
- Information of the tests for known inappropriate samples and the potential consequences;
- Factor and environment that may affect the results; if applicable, the possibility of carrying over contamination.
s) The references shall be given.
6 Packaging, Transport and Storage
6.1 Packaging
The packaging shall meet the following requirements.
a) The reagent (kit) packaging shall be free of natural and mechanical damage;
b) If applicable, the operation manual and product certificate shall be enclosed within the packaging.
6.2 Transport
The product shall be transported in accordance with the conditions stipulated in the contract.
6.3 Storage
Bibliography 1. GB/T 3358.1-2009 Statistics - Vocabulary and symbols - Part 1. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 3100SI units and recommendations for the use of their multiples and of certain other units
- GB/T 21415-2008In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
- YY/T 0638In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials
Similar standards
How to Download YY/T 1232-2014
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 3100-1993 — SI units and recommendations for the use of their multiples and of certain other units
GB/T 21415-2008 — In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
YY/T 0466.1-2009 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements