YY/T 1221-2013Cardiac troponin I (cTnI) diagnostic kit (colloidal gold method) (English PDF)
心肌肌钙蛋白1诊断试剂(盒)(胶体金法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1221-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 1221-2013 is the English-translated version of 心肌肌钙蛋白1诊断试剂(盒)(胶体金法).
YY/T 1221-2013 is the Chinese pharmaceutical industry standard for the diagnostic kit that detects cardiac troponin I by the colloidal gold method, the marker used in myocardial injury.
It specifies the terms and definitions, the requirements, the test methods, the inspection and judgement, the marking, the label, the instruction manual and the rules for packaging, transport and storage, covering the appearance of the strip, the detection limit, the agreement with reference materials, the specificity, the repeatability and the stability. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY/T 1221-2013
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Terms and Definitions4
- 4 Requirements5
- 5 Test Methods6
- 6 Inspection and Judgment7
- 7 Marking, Label and Instruction Manual8
- 8 Transportation and Storage9
- Bibliography11
Foreword
This Standard was drafted as per the rules specified in GB/T 1.1-2009.
Please note some contents of this document may involve patents. The issuing agency of this document shall not assume the responsibility to identify these patents.
1 Scope
This Standard specifies the terms and definitions, requirements, test methods, inspection and judgment, marking, label, instruction manual, packaging, transportation and storage of cardiac troponin I diagnostic kit (colloid gold method).
This Standard is applicable to the cardiac troponin I diagnostic kit (colloid gold method).
This reagent is used to qualitatively detect the cardiac troponin I (cTnI) in human serum or plasma.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) is applicable to this document.
GB/T 191 Packaging - Pictorial Marking for Handling of Goods
3 Terms and Definitions
For the purpose of this document, the following terms and definitions apply.
3.1 Colloid gold technique
A labeling technique that uses colloidal gold particles as a tracer or chromogenic substance for antigen-antibody reaction or other molecular detection.
3.2 Lowest detection limit
The lowest measured value in a sample that can be declared different from zero with a certain probability.
3.3 Analytical specificity
The measurement program only measures the ability to be measured.
3.5 Repeatability
Under the same measurement conditions, the consistency between the results obtained by consecutive multiple measurements on the same measured object.
4 Requirements
4.1 Physical properties
4.1.1 Appearance
It shall meet the normal appearance requirements specified by the manufacturer.
4.2 Accuracy
Test the positive control materials of cardiac troponin I (cTnI), all the results shall be positive.
4.3 Repeatability
Take 10 pieces of test papers of the same batch number to test the positive control materials of cardiac troponin I (cTnI) with the same concentration; the reaction results shall be consistent, the color shall be uniform, and all shall be positive.
4.5 Lowest detection limit
Test the sensitivity control materials of the cardiac troponin I (cTnI), which shall be no higher than 1.0ng/mL.
4.6 Stability
Take the test papers of the same batch number and place them at 37°C for 20d. The result shall meet the requirements of 4.2~4.5.
5 Test Methods
5.1 Basis for result judgment (see Figure 1)
5.1.1 Positive. The color of the color band of the test line (T) is the same as or deeper than the control line (C).
5.2 Physical properties
5.2.1 Appearance
Randomly select 1 piece of test paper and visually observe it under natural light. It shall meet the requirements of 4.1.1.
5.3 Accuracy
Test the positive control materials of the cardiac troponin I (cTnI); operate according to the instruction manual; and the results shall meet the requirements of 4.2.
5.4 Repeatability
Test the positive control materials of the cardiac troponin I (cTnI) with the same concentration; operate according to the instruction manual; and the result shall meet the requirements of 4.3.
5.7 Stability
Take the test papers of the same batch number and place them at 37°C for 20d; and test according to the methods shown in 5.3 ~ 5.6. The result shall meet the requirements of 4.6.
6 Inspection and Judgment
6.1 Reagent (kit) inspection type
Reagent (kit) inspection is divided into exit-factory inspection and type inspection.
6.3 Type inspection
Type inspection items are all items; and each batch of products shall be all-item inspected according to this Standard. Type inspection shall be carried out in one of the following situations.
7 Marking, Label and Instruction Manual
7.1 The outer packaging box of the reagent (kit) shall have at least the following contents.
7.2 The transportation and packaging of the reagent kit shall have the following contents.
7.3 The instruction manual shall have the following contents. […]
8 Transportation and Storage
8.1 Packaging
8.1.1 The reagent may be transported at the room temperature for a short period of time. Corresponding protective measures shall be enhanced in Winter or Summer, avoid becoming icy or at high temperature for a long time.
8.2 Storage
8.2.1 The reagent shall be stored in the environment at 4°C~30°C; no freezing storage is allowed.
8.2.2 Avoid heavy stress during the storage period. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
How to Download YY/T 1221-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 191-2025 — Graphical symbols marking for handling and storage of packages
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
Secure payment via Stripe
Payments accepted
YY/T 1221-2013
$145.00