YY/T 1216-2013Alpha-fetoprotein quantitative labelling immunoassay kit (English PDF)
甲胎蛋白定量标记免疫分析试剂盒
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1216-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 1216-2013 is the English-translated version of 甲胎蛋白定量标记免疫分析试剂盒.
YY/T 1216-2013 is the Chinese standard "Alpha-fetoprotein quantitative labelling immunoassay kit". Its scope clause reads: This Standard specifies the classification, requirements, test method, marks, labels, operating instructions, packaging, transportation, and storage of the alpha-fetoprotein quantitative labelling immunoassay kit.
This Standard is applicable to the quantitative detection of alpha-fetoprotein labelling immunoassay kit (hereinafter referred to as AFP kit).
It includes AFP immunoassay kit of quantitative detection by using labelling methods such as enzyme labelling, (electrical) chemiluminescent labelling, (time resolution) fluorescence labelling AS capture antibody; and using microplates, pipes, magnetic particles, microbeads, plastic beads and others AS the carrier coated antibody.
This Standard does not apply to: a) Colloidal gold labelled AFP test strip; b) Various types of radio-immunity or IRMA reagent kit labelled with 125 I and other radioactive isotopes. Its clauses include classifications; test methods.
Found what you need?
Document preview — YY/T 1216-2013
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Classifications4
- 4 Requirements5
- 5 Test methods6
- 6 Marks, labels and operating instructions8
- 7 Packaging, transportation and storage9
- Bibliography10
1 Scope
This Standard specifies the classification, requirements, test method, marks, labels, operating instructions, packaging, transportation, and storage of the alpha-fetoprotein quantitative labelling immunoassay kit.
This Standard is applicable to the quantitative detection of alpha-fetoprotein labelling immunoassay kit (hereinafter referred to as AFP kit). It includes AFP immunoassay kit of quantitative detection by using labelling methods such as enzyme labelling, (electrical) chemiluminescent labelling, (time resolution) fluorescence labelling AS capture antibody;
and using microplates, pipes, magnetic particles, microbeads, plastic beads and others
AS the carrier coated antibody.
This Standard does not apply to:
a) Colloidal gold labelled AFP test strip;
b) Various types of radio-immunity or IRMA reagent kit labelled with
125
I and other radioactive isotopes.
2 Normative references
The articles contained in the following documents have become part of this Standard when they are quoted herein. For the dated documents so quoted, all the modifications (Including all corrections) or revisions made thereafter shall be applicable to this document.
YY/T 0466.1-2009 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
3 Classifications
The AFP kits can be divided into ELISA reagent kit, chemiluminescent reagent kit, time resolution fluorescence reagent kit etc., according to the various labelling methods. It can be divided into different kinds of AFP kits with carriers such as microplates, pipes, magnetic particles, microbeads and plastic beads, according to the various solid-phase carriers. The AFP kits can be divided into manual operation method and automatic […]
4.6 Specificity
DETECT at least 10 normal serums. The detection results shall not be higher than 20.0 ng
/ mL.
4.7 Measured value of quality control material
Within the linear range of the kit, SET 2 or 3 quality control materials with different concentrations. The measured results shall be within the allowable interval of the measured value of quality control products.
4.8 Stability
The following methods may be selected:
4.8.1 Stability at the end of validity
The reagent kit is preserved until the end of the validity under a specified condition. The measured results shall comply with the provisions of 4.1, 4.2, 4.3, 4.4, 4.5.1, 4.5.2, 4.6 and 4.7.
4.8.2 Thermal stability
According to the period of validity of the kit, the kit is usually placed under the condition of
37°C for some time (usually 3 days - 7 days). The inspection results shall comply with the provisions of 4.1, 4.2, 4.3, 4.4, 4.5.1, 4.5.2, 4.6, and 4.7.
Note 1: The thermal stability cannot be used to derive the period of validity of products, unless it uses the derivation equation that is established based on a large number of stability study data;
Note 2: One of the above methods can be selected according to the product characteristics. However, the selected method shall be capable of verifying the stability of the product, so as to ensure that the performance of the product can comply with the standard requirements within the period of validity.
5 Test methods
5.1 Appearance
USE visual inspection method. Visually INSPECT under natural light and bright place. It shall comply with the provision of 4.1.
5.2 Minimum detection limit
DETERMINE the signal values of zero-value calibrator or sample diluent for not less than
10 times. CALCULATE the mean ( ) and standard deviation (SD). And CALCULATE the value of ( + 2SD). The sample concentration corresponding to that value is the minimum detection limit of the kit. It shall comply with the provision of 4.2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Similar standards
How to Download YY/T 1216-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 1216-2013
$125.00