YY/T 1216-2013Alpha-fetoprotein quantitative labelling immunoassay kit (English PDF)

甲胎蛋白定量标记免疫分析试剂盒

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1216-2013 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

October 21, 2013

Implementation date

October 1, 2014

Scope

YY/T 1216-2013 is the English-translated version of 甲胎蛋白定量标记免疫分析试剂盒.

YY/T 1216-2013 is the Chinese standard "Alpha-fetoprotein quantitative labelling immunoassay kit". Its scope clause reads: This Standard specifies the classification, requirements, test method, marks, labels, operating instructions, packaging, transportation, and storage of the alpha-fetoprotein quantitative labelling immunoassay kit.

This Standard is applicable to the quantitative detection of alpha-fetoprotein labelling immunoassay kit (hereinafter referred to as AFP kit).

It includes AFP immunoassay kit of quantitative detection by using labelling methods such as enzyme labelling, (electrical) chemiluminescent labelling, (time resolution) fluorescence labelling AS capture antibody; and using microplates, pipes, magnetic particles, microbeads, plastic beads and others AS the carrier coated antibody.

This Standard does not apply to: a) Colloidal gold labelled AFP test strip; b) Various types of radio-immunity or IRMA reagent kit labelled with 125 I and other radioactive isotopes. Its clauses include classifications; test methods.

Found what you need?

Document preview — YY/T 1216-2013

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Classifications4
  • 4 Requirements5
  • 5 Test methods6
  • 6 Marks, labels and operating instructions8
  • 7 Packaging, transportation and storage9
  • Bibliography10

1 Scope

This Standard specifies the classification, requirements, test method, marks, labels, operating instructions, packaging, transportation, and storage of the alpha-fetoprotein quantitative labelling immunoassay kit.

This Standard is applicable to the quantitative detection of alpha-fetoprotein labelling immunoassay kit (hereinafter referred to as AFP kit). It includes AFP immunoassay kit of quantitative detection by using labelling methods such as enzyme labelling, (electrical) chemiluminescent labelling, (time resolution) fluorescence labelling AS capture antibody;

and using microplates, pipes, magnetic particles, microbeads, plastic beads and others

AS the carrier coated antibody.

This Standard does not apply to:

a) Colloidal gold labelled AFP test strip;

b) Various types of radio-immunity or IRMA reagent kit labelled with

125

I and other radioactive isotopes.

2 Normative references

The articles contained in the following documents have become part of this Standard when they are quoted herein. For the dated documents so quoted, all the modifications (Including all corrections) or revisions made thereafter shall be applicable to this document.

YY/T 0466.1-2009 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements

3 Classifications

The AFP kits can be divided into ELISA reagent kit, chemiluminescent reagent kit, time resolution fluorescence reagent kit etc., according to the various labelling methods. It can be divided into different kinds of AFP kits with carriers such as microplates, pipes, magnetic particles, microbeads and plastic beads, according to the various solid-phase carriers. The AFP kits can be divided into manual operation method and automatic […]

4.6 Specificity

DETECT at least 10 normal serums. The detection results shall not be higher than 20.0 ng

/ mL.

4.7 Measured value of quality control material

Within the linear range of the kit, SET 2 or 3 quality control materials with different concentrations. The measured results shall be within the allowable interval of the measured value of quality control products.

4.8 Stability

The following methods may be selected:

4.8.1 Stability at the end of validity

The reagent kit is preserved until the end of the validity under a specified condition. The measured results shall comply with the provisions of 4.1, 4.2, 4.3, 4.4, 4.5.1, 4.5.2, 4.6 and 4.7.

4.8.2 Thermal stability

According to the period of validity of the kit, the kit is usually placed under the condition of

37°C for some time (usually 3 days - 7 days). The inspection results shall comply with the provisions of 4.1, 4.2, 4.3, 4.4, 4.5.1, 4.5.2, 4.6, and 4.7.

Note 1: The thermal stability cannot be used to derive the period of validity of products, unless it uses the derivation equation that is established based on a large number of stability study data;

Note 2: One of the above methods can be selected according to the product characteristics. However, the selected method shall be capable of verifying the stability of the product, so as to ensure that the performance of the product can comply with the standard requirements within the period of validity.

5 Test methods

5.1 Appearance

USE visual inspection method. Visually INSPECT under natural light and bright place. It shall comply with the provision of 4.1.

5.2 Minimum detection limit

DETERMINE the signal values of zero-value calibrator or sample diluent for not less than

10 times. CALCULATE the mean ( ) and standard deviation (SD). And CALCULATE the value of ( + 2SD). The sample concentration corresponding to that value is the minimum detection limit of the kit. It shall comply with the provision of 4.2.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Similar standards

How to Download YY/T 1216-2013

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1216-2013

$125.00