YY/T 1203-2013Sodium assay kit(Enzymic method) (English PDF)
钠测定试剂盒(酶法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1203-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 1203-2013 is the English-translated version of 钠测定试剂盒(酶法).
YY/T 1203-2013 is the Chinese pharmaceutical industry standard for the sodium assay kit based on an enzymatic method.
It specifies the requirements, the test methods, the marking, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the blank absorbance, the traceability of the calibrator, the accuracy, the linear range, the precision, the interference and the stability, so that the quality of such a kit can be controlled. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY/T 1203-2013
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
1 Scope
This standard specifies the sodium assay kit (enzymatic) requirements, test methods, marking, labeling and instructions for use, packaging, transport and storage.
This standard applies to soda assay kit (enzymatic) quality control.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
Symbols - Part 1 YY/T 0466.1 medical equipment used with medical device labels, labeling and information. General requirements
3 Requirements
3.1 Appearance
Compatible with the manufacturer's normal appearance requirements.
3.2 PACKING
Loading amount of the liquid reagent to be less than the labeled amount.
3.3 Reagent blank
3.3.1 Reagent blank absorbance
Reagent blank absorbance <=1.0 (wavelength 405nm, light trails 1.0cm).
3.3.2 Reagent blank absorbance change rate
Reagent blank absorbance change rate <=0.500/min (wavelength 405nm, light trails 1.0cm).
3.4 Linear Interval
In 90mmol/L ~ 160mmol/L range, the linear correlation coefficient (r) shall not be less than 0.9900.
3.5 Accuracy
You can use the following methods.
a) relative deviation. serum assay provides reference material or reference values of the reference method, the deviation measured value and the value declared in within ± 15.0%;
b) recovering the experiment. The determination of calibrator, kits recovery should be within 90% to 110%;
c) comparative experiments. no calibrator to comparison method for determining, r>=0.9900.
3.6 Sensitivity Analysis
When reagent (kit) test analyte absorbance difference (deltaA) or change in absorbance should be consistent with manufacturer's given range.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1203-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0466.1-2009 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2016 — Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY/T 0466.1-2023 — Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
YY/T 1203-2013
$145.00