YY/T 0997-2015Elbow, knee joints passive motion equipment (English PDF)
肘、膝关节被动运动设备
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0997-2015 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 2, 2015
Implementation date
January 1, 2016
Scope
YY/T 0997-2015 is the English-translated version of 肘、膝关节被动运动设备.
YY/T 0997-2015 is the Chinese pharmaceutical industry standard for the passive motion equipment of the elbow and the knee, the machine that moves the joint through a set range while the patient stays passive.
It specifies the terms and definitions, the requirements and the test methods, covering the range of movement and its accuracy, the speed and the cycle, the force the machine may apply and its limit, the emergency stop and the release of the limb, the strength of the frame and the electrical safety. Issued on 2 March 2015, it has applied since 1 January 2016.
Found what you need?
Document preview — YY/T 0997-2015
National Standard of the People's Republic of China
- ICS
- 11.040.60
- Classification
- C 42
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions GB 9706.1-2007 defined and the following
- 4 Requirements
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
The standard safety requirements for full implementation of the GB 9706.1-2007 "Medical Electrical Equipment Part 1. General requirements for safety," the content.
1 Scope
This standard specifies the elbow, knee joint passive motion device (hereinafter referred to as the device) terms and definitions, requirements, test methods.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
GB 9706.1-2007 Medical electrical equipment - Part 1. General requirements for safety
GB/T 14710-2009 medical electrical environmental requirements and test methods
GB/T 16886.1-2011 Biological evaluation of medical devices - Part 1. Evaluation of the risk management process and Test
3 Terms and Definitions
GB 9706.1-2007 defined and the following terms and definitions apply to this document.
3.1 Elbow and knee joint passive motion equipment
One kind of drive the elbow, knee joint flexion and extension in the direction of motion of the joint co-treatment or rehabilitation of electrical equipment.
This standard applies to equipment regulations.
4 Requirements
4.1 Working conditions
Unless the manufacturer otherwise specified, shall meet the requirements of GB 9706.1-2007 in Chapter 10.
4.2 Performance Requirements
4.2.1 Angle range and tolerance
4.2.1.1 Equipment shall be adjustable angle range, angular range should not exceed the values specified in Table 1.
4.2.1.2 When within an adjustable angular range, the angle is not greater than 50 °, Tolerance. ± 5 °; the angle is greater than 50 °, tolerance. ± 10%.
4.2.2 Angular velocity
The angular velocity of the equipment shall be adjusted, the maximum angular velocity shall not exceed the values specified in Table 1, tolerance. ± 20%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1-2007Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 14710-2009The environmental requirement and test methods for medical electrical equipment
- GB/T 16886.1-2011Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
How to Download YY/T 0997-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2007 — Medical electrical equipment - Part 1: General requirements for safety
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 16886.1-2011 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
YY/T 0997-2015