YY/T 0987.4-2016Magnetic resonance compatibility for surgical implant - Part 4: Test method for radio-frequency heat (English PDF)
外科植入物磁共振兼容性 第4部分:射频致热试验方法
This is a limited preview
Buy now to download the full PDF (21 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0987.4-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 23, 2016
Implementation date
January 1, 2017
Scope
YY/T 0987.4-2016 is the English-translated version of 外科植入物磁共振兼容性 第4部分:射频致热试验方法.
YY/T 0987.4-2016 is the Chinese pharmaceutical industry standard that gives the test method for the radiofrequency heating of a passive implant and of the tissue around it during magnetic resonance imaging.
It is the fourth part of the YY/T 0987 series and sets out the phantom and its gel, the placing of the implant and of the temperature probes, the RF exposure and its duration, and the measurement and reporting of the temperature rise. Issued on 23 March 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 0987.4-2016
National Standard of the People's Republic of China
- ICS
- 11.040.40
- Classification
- C 35
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Preface
- 1 Scope
- 2 normative reference document
- 3 Terms and definitions
- 4 Overview of test methods
- 5 meaning and application
- 7 Test sample
- 8 Step
- 9 calorimetric measurement of saline solution body (body mold) average SAR value
- 10 Report
- Appendix A (informative) Principle of test method References
Foreword
YY/T 0987 "Surgical Implants Magnetic Resonance Compatibility" is divided into the following sections.
- Part 1. Safety markings;
- Part 2. Magnetic flux force test method; Part 3. Image artifact evaluation method;
- Part 4. Radiofrequency test method;
- Part 5. Magneto - induced torque test method. This part is part 4 of YY/T 0987.
This part is drafted in accordance with the rules given in GB/T 1.1-2009.
This section uses the re-drafting method for reference to ASTM F2182-2011a "Magnetic Resonance Environment in Passive Implants and Surrounding Tissues Radio Frequency Heat test method "preparation.
The technical differences between this section and ASTM CF2182-2011a are as follows.
--- on the normative reference documents, this part has made a technical differences in the adjustment to meet China's technical conditions, adjust the situation The situation is reflected in Chapter 2, "Normative references", as follows.
- Replacement with ASTM Y2503-08 with YY/T 0987.1;
YY/T 0987.2 instead of ASTM F2052-06;
YY/T 0987.3 instead of ASTM F2119-07;
- Replacement with ASTM Y2213-06 with YY/T 0987.5; Replace Y 601-2-33..2002 with YY 0319-2008;
--- remove the content of polyacrylic acid partial sodium salt and hydroxyethyl cellulose brand;
- Remove Chapter 11, Chapter 12 and Appendix X1.10 of ASTM F2182-11a.
1 Scope
This part of YY/T 0987 specifies tests for radiofrequency (RF) heating on passive implants or surrounding tissue by magnetic resonance imaging (MRI) method.
This part of the test method is to determine whether the presence of medical devices in the MR imaging process will lead to injury to the patient one of the test methods.
Other safety issues that need to be addressed include magnetic flux and torque and the functional realization of various configurations of medical devices in the MR environment.
At a given specific absorptivity (SAR), the temperature rise caused by radio frequency depends on the frequency of the radio frequency, which in turn depends on the Magnetic field strength. The test methods in this section focus on the 1.5T or 3T cylindrical MR system, the MR system for other static magnetic field strengths Or magnet design, this part of the test method can be modified to evaluate the RF caused by the temperature rise.
This part of the test hypothesis is carried out when the implant is completely in the body. For other implant conditions (such as external fixation equipment, puncture needle, guide Tube or tethered equipment such as ablation probe), the need to modify the test method.
This part of the test method is applicable to a whole body magnetic resonance device with a whole body radio frequency transmitting coil, which is defined in YY 0319-2008 2.2.103, the definition of the whole body radio frequency transmitter coil see 2.2.100. Assume that the RF coil is an orthogonal excitation coil.
This part uses the international unit value as the standard, this part does not contain other units of measurement.
2 Normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this document. For undated references, the latest edition (including all modifications) applies to this document.
Medical electrical equipment - Part 2-33. Particular requirements for the safety of magnetic resonance devices for medical diagnostics (IEC 60601-2 - 33..2002, IDT)
YY/T 0987.1 Surgical implants - Magnetic resonance compatibility - Part 1. Safety signs
YY/T 0987.2 Magnetic resonance compatibility of surgical implants - Part 2. Magnetic susceptibility test method
YY/T 0987.3 Magnetic resonance compatibility of surgical implants - Part 3. Image artifact evaluation method
YY/T 0987.5 Surgical implants - Magnetic resonance compatibility - Part 5. Magnetic torques test method NEMAMS8-2008 magnetic resonance imaging system specific absorption rate characteristics (Characterizationofthespecificabsorp-tionrate
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Salt gelatin geledsaline
Sodium chloride and polyacrylic acid or sodium chloride and hydroxyethyl cellulose were dissolved in water in this part of the test medium.
3.2 Implant
Medical means an object, device, or instrument that is intended to remain in the body for the purpose of diagnosis, repair, or other treatment.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.
Referenced standards
Normative references
- YY/T 0987.1Implants for Surgery - Magnetic Resonance Compatibility - Part 1: Safety Marking
- YY/T 0987.2Magnetic resonance compatibility for surgical implant - Part 2: Test method for magnetostrictive displacement
- YY/T 0987.3Magnetic resonance compatibility for surgical implant - Part 3: Assessment method for image artifact
- YY/T 0987.5Magnetic resonance compatibility for surgical implants - Part 5: Test method for magnetically induced torque
Similar standards
How to Download YY/T 0987.4-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0987.1-2016 — Implants for Surgery - Magnetic Resonance Compatibility - Part 1: Safety Marking
YY/T 0987.2-2016 — Magnetic resonance compatibility for surgical implant - Part 2: Test method for magnetostrictive displacement
YY/T 0987.3-2016 — Magnetic resonance compatibility for surgical implant - Part 3: Assessment method for image artifact
Secure payment via Stripe