Scope

YY/T 0891-2013 is the English-translated version of 血管造影高压注射装置专用技术条件.

YY/T 0891-2013 is the Chinese pharmaceutical industry standard that sets the particular specifications for the angiographic injector, the powered injector that delivers contrast medium into a vessel.

It specifies the terms and definitions, the composition, the requirements and the test methods, covering the injected volume and the flow rate and their accuracy, the pressure limit and its control, the rise time and the delay, the syringe and the heating of the contrast medium, the protection against air injection and the electrical safety. Issued on 21 October 2013, it has applied since 1 October 2014.

Found what you need?

Document preview — YY/T 0891-2013

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions4
  • 4 Composition6
  • 5 Requirements6
  • 6 Test methods8

Foreword

This Standard is drafted in accordance with the rules given in GB/T 1.1-2009.

1 Scope

This Standard specifies terms and definitions, composition, requirements and test methods for angiographic injectors (hereinafter referred to as "injectors"). This Standard applies to angiography injectors.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the dated edition cited applies. For undated references, the latest edition of the referenced document (including all amendments) applies.

GB 9706.1-2007 Medical electrical equipment - Part 1: General requirements for safety

GB 9706.15-2008 Medical electrical equipment - Part 1: General requirements for safety - 1.Collateral standard. Safety requirements for medical electrical systems

3 Terms and definitions

For the purpose of this document, the following terms and definitions and those defined in GB/T 10149 apply.

3.1 Angiographic injector

The injector used to rapidly inject a sufficient amount of contrast agent to the inspection site in a certain period of time, in the diagnosis or treatment of vascular intervention angiography, in order to obtain the required medical image with the use of X-ray angiography equipment.

3.2 Injector head

The part equipped with injection push mechanism and supporting syringe.

3.3 Flow rate

The dosage of the contrast agent injected per unit of time, in milliliters per second (mL/s).

3.4 Rise time

The time taken from the start of the injection to when it reaches the set flow rate, in seconds.

3.5 Maximum injection pressure

The maximum pressure that can be reached at the time of injection of an injector.

4 Composition

The injector consists of at least a push and control system, a supporting and fixing part, accessories, etc.

5 Requirements

5.1 Working conditions

5.1.1 Environmental conditions

Unless otherwise specified, the working conditions of the injector shall meet.

5.1.2 Power supply conditions

The power supply conditions of the injector shall meet.

5.2 Flow rate

The flow rate range of injectors for hearts and large vessels shall not be less than 1.0 mL/s to 30 mL/s. When the flow rate is less than or equal to 15 mL/s, the deviation shall not be greater than ±5 %; when the flow rate is greater than 15 mL/s, the deviation shall not be greater than ±10 %.

5.4 Fill rate

The fill rate of injectors shall be specified in the product standard, the deviation shall not be greater than ±5 %.

5.5 Maximum injection pressure

The maximum injection pressure of injectors for hearts and large vessels shall not be less than 6.89 MPa (1000 psi).

5.6 Pressure limit

The device shall be capable of setting the pressure limit during injection. When the pressure limit is less than or equal to 4.13 MPa (600 psi), the deviation shall not exceed ±0.34 MPa (50 psi); when the pressure limit is greater than 4.13 MPa (600 psi), the deviation shall not exceed ±0.48 MPa (70 psi).

5.9 Rise time

The setting range of the rise time should be specified in the product standard.

5.10 Rotation angle of injector head

The angle between the upward injector head and the horizontal plane shall not be less than 60°, and the angle between the downward injector head and the horizontal plane shall not be less than 10°.

5.11 Functions

The injector shall have the following functions.

5.12 Appearance

It shall meet the following requirements. […]

5.14 Security

It shall meet the requirements of GB 9706.1-2007 and GB 9706.15-2008.

6 Test methods

6.6 Pressure limit SET the pressure limit to 20 % of the maximum injection pressure (or near value) and 80 % of the maximum injection pressure (or near value) respectively; USE a tee connector to connect the syringe, the pressure gauge and the catheter; INJECT in a flow rate of 3 mL/s; READ the pressure the injector indicates when it exceeds the pressure limit and stops the injection; CALCULATE the deviation between the actual pressure and the set pressure, which shall meet the requirements of 5.5.

6.1 Working conditions

6.1.1 Environmental conditions

It shall comply with the specifications of 5.1.1.

6.1.2 Power supply conditions

The power supply conditions are as follows. […]

6.2 Flow rate

The test medium is distilled or deionized water. The injection dose is set to 100 mL, and the flow rate is set to the minimum value and 50 % of the maximum value respectively.

6.3 Injection dose

The test medium is distilled or deionized water. SET the injection dose to 2 mL (when the flow rate is the minimum value), 20 mL (when the flow rate is 50 % of the maximum value) and 40 mL (when the flow rate is the maximum value); USE a balance to weigh the distilled water and convert it to capacity; CALCULATE the injection dose deviation, which shall meet the requirements of 5.3.

6.4 Fill rate

The test medium is distilled or deionized water. USE a timer to count the time taken from the start to the end of the filling signal; DETERMINE the capacity of the filled injection agent by weighing the weight of the container before and after filling;

CALCULATE the fill rate, which shall meet the requirements of 5.4.

6.7 Injection delay

SET the delay time to 3 s; USE a timer to count the time taken from receiving the signal to starting the injection; CALCULATE the injection delay deviation, which shall meet the requirements of 5.7.

6.9 Rise time

Actual operation, which shall meet the requirements of 5.9.

6.10 Rotation angle of injector head

Actual operation, measured with a goniometer. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 0891-2013

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0891-2013

$125.00