YY/T 0842-2011Medical endoscopes—Endoscope accessories—Sheaths (English PDF)
医用内窥镜 内窥镜附件 镜鞘
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0842-2011 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
December 31, 2011
Implementation date
June 1, 2013
Scope
YY/T 0842-2011 is the English-translated version of 医用内窥镜 内窥镜附件 镜鞘.
YY/T 0842-2011 is the Chinese pharmaceutical industry standard for the sheath used with a medical endoscope, through which the endoscope is inserted.
It specifies the requirements and the test methods for it, covering the dimensions and the working length, the strength and the flexibility of the sheath, the tightness and the valves where fitted, the surface finish, the corrosion resistance and the materials and their biological evaluation. Issued on 31 December 2011, it has applied since 1 June 2013.
Found what you need?
Document preview — YY/T 0842-2011
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Requirements5
- 5 Test methods8
- 6 Inspection rules10
Foreword
This Standard is drafted in accordance with the rules given in GB/T 1.1-2009.
The Standard is established under the title of "Medical endoscopes - Endoscope accessories".
This Standard not-equivalently adopts the international standards ISO 8600-1:2005 "Optics and photonics - Medical endoscopes and endotherapy devices - Part 1.
General requirements" and ISO 8600-4:1997 "Optics and optical instruments - Medical endoscopes and certain accessories - Part 4. Determination of maximum width of insertion portion".
1 Scope
This Standard specifies the requirements and test methods of sheaths.
This Standard applies to sheaths for the insertion of medical endoscopes.
2 Normative references
The following documents are essential for the application of this document. For all dated reference documents, only the dated versions are applicable to this document.
For all undated reference documents, the latest versions (including all modifications) are applicable to this document.
GB/T 16886.1 Biological evaluation of medical devices - Part 1. Evaluation and testing within a risk management process (ISO 10993-1:2009, IDT)
YY/T 0149-2006 Medical instruments of stainless steel - Test methods of corrosion resistance (ISO 13402:1995, MOD)
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Endoscope accessory
Optional accessories that are necessary and/or suitable to be used with endoscopes, to achieve the intended use of endoscopes, or to provide convenience for the intended use of endoscopes, or to improve the intended use of endoscopes, or to increase the additional use of endoscopes.
3.2 Sheath
External inserting pipes for guiding endoscopes or endotherapy devices.
3.3 Main channel
A part of sheaths. Endoscopes or endotherapy devices are required to be inserted through this part.
3.4 Minimum main channel width
Minimum internal width of the channel of endoscopes or endotherapy devices.
3.5 Insertion portion
A part of sheaths, which can be inserted into natural channels or surgical incisions of human bodies.
3.5.1 Working length
Maximum length of sheath's insertion portion.
4 Requirements
4.1 Materials
4.1.1 Chemical composition of metal materials
Manufacturers of the metal materials that directly or indirectly contact with patients shall explicitly indicate in any possible form, wherein the metal materials shall be marked with grades and/or code as well as the requirements for materials' chemical composition. The chemical composition of metal materials shall be verified by tests.
4.1.2 Biological compatibility
Materials that directly or indirectly contact with patients shall be conducted with biological safety evaluation in accordance with the principles and requirements in GB/T Sheaths and endoscopes or scope mounts or obturators shall meet the following requirements after mounting.
a) the positioning shall be reliable and has no loosening after full insertion;
b) except for no sealing requirement claimed by the manufacturer, the sealing after full insertion shall be in a good condition. The water seeped shall not exceed more than 5 drops within 1 min when performing water permeability test.
4.4 Connection
Every connecting portion involving sheaths shall be connected by tight mounting method, the mounting area shall be no visible cracks; if it is connected by welding, the welding seam shall be no unevenness, loose weld, bead weld or apparent pitting; if it is connected by gluing, the gluing area shall be no effusion or apparent heaping glue, other parts shall be no flow trance of glue.
4.5 Outer surface quality of insertion portion
Except for special purpose, there shall be nothing that may cause injury on the insertion portion.
4.6 Water-passing and/or air-passing functions
4.6.1 Water-passing and/or air-passing ducts
If the sheath has water-passing and/or air-passing functions, there shall be no impurities in water-passing and/or air-passing ducts, and water-passing and/or air-passing processes shall be smooth.
4.6.2 Water-passing and/or air-passing connectors
If the sheath has water-passing and/or air-passing functions, the manufacturer shall provide the specification of water-passing and/or air-passing connectors. Water-passing and/or air-passing connectors of sheath shall meet the requirements of this specification or relevant standards applicable for this specification.
4.7 Tolerance (applicable for non-disposable products)
After the following tolerance tests, the sheath shall still be able to meet the requirements of 4.3.
a) High-pressure and high-temperature resistance Sheaths with high-pressure and high-temperature resistance shall be able to […]
5 Test methods
5.6 Water-passing and/or air-passing functions
5.6.1 Water-passing and/or air-passing ducts
VERIFY by visual observation and water-passing and air-passing operations.
5.6.2 Water-passing and/or air-passing connectors
MEASURE according to the method specified in the relevant standards applicable for connector specification.
5.7 Tolerance test
5.7.1 High-temperature and high-pressure resistance test
After performing 20 times of high-temperature and high-pressure resistance test (the test conditions are in accordance with the specification in manufacturer's instruction manual), the result shall still meet the requirements of 4.3.
5.7.2 Corrosion resistance test
The portion uses stainless steel material shall be tested according to the boiling water test method in YY/T 0149-2006, the result shall meet the requirements of 4.7 b). Other materials shall be processed according to the sterilization method specified by the manufacturer in the instruction manual, there shall be no corrosion, and the result shall still meet the requirements of 4.3.
6 Inspection rules
Inspection rules are made by the manufacturer according to product features. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 0842-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 16886.1-2022 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
YY/T 0149-2006 — Medical instruments of stainless steel - Test methods of corrosion resistance
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
Secure payment via Stripe
Payments accepted
YY/T 0842-2011
$125.00