YY/T 0820-2010Dental cartridge syringes (English PDF)
牙科筒式注射器
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0820-2010 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
December 27, 2010
Implementation date
June 1, 2012
Scope
YY/T 0820-2010 is the English-translated version of 牙科筒式注射器.
YY/T 0820-2010 is the Chinese pharmaceutical industry standard for the dental cartridge syringe, both the aspirating and the non-aspirating kind, through which a local anaesthetic is given from a cartridge.
It sets the requirements for such syringes and the methods by which they are tested, covering the dimensions and the fit of the cartridge and of the needle, the aspiration where it applies, the force needed to inject, the tightness of the assembly, the corrosion resistance and the sterilization of the syringe. Issued on 27 December 2010, it applies from 1 June 2012.
Found what you need?
Document preview — YY/T 0820-2010
National Standard of the People's Republic of China
- ICS
- 11.060.01
- Classification
- C 33
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 5 Requirements
Foreword
This standard is identical with ISO 9997:1999 "Dental cartridge syringes." This standard specifies the requirements for ISO metric thread sizes dental syringe barrel, and is based on information given in appendix with inch screw Grain size requirements. Appendix A of this standard is the data appendix.
1 Scope
This standard specifies the local anesthetic through a syringe inhaled, inhaled and non-inhaled from a dental syringe barrel and requirements Detection method.
This standard does not apply to the use of mechanical movements to create a high-pressure syringe barrel.
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
GB/T 193-2003 common thread diameter and pitch range (ISO 261:1998, MOD)
ISO 965-1 general purpose metric screw threads - Tolerances - Part 1. Principles and basic data
ISO 1942-3 Dental terminology - Part 3. Dental instruments
ISO 13402:1995 for dental surgical handpiece autoclave resistance requirements
ISO 11499 local anesthesia with a dental syringe
3 Terms and definitions
ISO 1942-3, ISO 11499 and established the following terms and definitions apply to this standard.
3.1 Suction aspiration
Blood and body fluids into the syringe anesthesia process.
3.2 Single primary package unitpack
Equipped with a dental cartridge syringe packaging.
3.3 Syringe cartridge
Anesthetics for loading containers.
Category 4 In accordance with the purposes of this standard, a dental cartridge syringes divided into the following types.
--- Type 1. Non inhaled;
--- Type 2. Inhaled.
--- 2a type. by pulling the piston away from the needle to generate suction;
--- 2b Type. diaphragm deflection derived via syringe suction.
5 Requirements
5.1 Summary
And ISO 11499 consistent with local anesthetic syringe.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
ISO 965-1 · ISO 1942-3 · ISO 13402:1995 · ISO 11499
Similar standards
How to Download YY/T 0820-2010
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 193-2003 — General purpose metric screw threads -- General plan
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
Secure payment via Stripe
Payments accepted
YY/T 0820-2010
$145.00