YY/T 0119.5-2014Spinal implants—Components used in the surgical fixation of the spinal skeletal system - Part 5: Test method for measure the static and fatigue bending strength of metallic screws (English PDF)
脊柱植入物 脊柱内固定系统部件 第5部分:金属脊柱螺钉静态和疲劳弯曲强度测定试验方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0119.5-2014 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
June 17, 2014
Implementation date
July 1, 2015
Scope
YY/T 0119.5-2014 is the English-translated version of 脊柱植入物 脊柱内固定系统部件 第5部分:金属脊柱螺钉静态和疲劳弯曲强度测定试验方法.
YY/T 0119.5-2014 is the Chinese pharmaceutical industry standard that gives the test method for measuring the static and fatigue bending strength of a metallic spinal screw.
It is the fifth part of the YY/T 0119 series and applies to the screws used in spinal fixation implants, setting out the test block and the embedding of the screw, the loading arrangement and the rate, the static test and the cyclic test with its runout level, and the way the results are reported. Issued on 17 June 2014, it has applied since 1 July 2015.
Found what you need?
Document preview — YY/T 0119.5-2014
National Standard of the People's Republic of China
- ICS
- 11.040.40
- Classification
- C 35
- Replacing
- YY 0119-2002
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Summary of Test Method
- 4 Significance and application
Foreword
YY/T 0119 "was within the spinal implant spinal fixation system components" is divided into five sections.
--- Part 1. General requirements;
--- Part 2. Metal screws spine;
--- Part 3. Metal spine board;
--- Part 4. Metal spinal rod;
--- Part 5. Metal screw spinal static and fatigue bending strength Test method. This section YY/T Section 50119 of. This section drafted in accordance with GB/T 1.1-2009 given rules.
Used in this section refer to redraft the law ASTM F2193-07 "standard spinal fixation system components and test methods" in Appendix A4 And X5 establishment.
1 Scope
YY/T 0119 provisions of this part of the test method for the determination of metal screws spine static and dynamic bending properties.
This section applies to the spinal fixation implants in the spine metal screws.
This section is intended to provide a means for different types of spinal screw the mechanical characterization methods. There is currently not enough pre-knowledge Use screws to know the results of specific design, it is not the purpose of this test method is used to define the level of performance of spinal screw.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
Inspection and GB/T 16825.1 of static uniaxial testing machines - Part 1 Rally and (or) pressure testing machine force measuring system calibration
GB/T 19701.2 ultra-high molecular weight polyethylene - Part 2. plastic molding ASTM E122 using specific tolerances, the sample size to assess the production batch or process characteristics mean calculation specifications (Practicefor calculatingsamplesizetoestimate, withaspecifiedtolerableerror, theaverageforacharacteristicofa lotorprocess)
ASTM E467 Axial Fatigue Test System for constant amplitude dynamic loads calibration specification (Standardpracticeforverification ofconstantamplitudedynamicforcesinanaxialfatiguetestingsystem)
ASTM E1942 cycle fatigue and fracture mechanics test data acquisition system evaluation guidelines (Guideforevaluatingdataac-quisitionsystemsusedincyclicfatigueandfracturemechanicstesting)
ASTM F1839 orthopedic instruments and tools to test using standard materials - rigid polyurethane foam (Specificaionforrigidpolyure-thanefoamforuseasastandardmaterialfortestingorthopedicdevicesandinstruments)
3 Summary of Test Method
3.1 pairs of spinal screw quasi-static cantilever bending load is applied, so that the recording member in response to applied mechanical load corresponding. By Loads - Deflection curves were used to determine the data for deriving the spinal screw flexural properties.
3.2 pairs of sinusoidal cantilever sample is applied to a given predetermined frequency spinal screw bending loads. Until one of the following three conditions Fatigue test. samples Product failure, limit or terminate the trial reaches a predetermined number of cycles (termination limit). The test results information is used to draw the moment - cycle times Curve to characterize a given moment in the spinal screw type range applied general fatigue properties. Further, in the case where the result of the appropriate Information can be used to determine the spinal screw N at a predetermined number of cycles in the value of fatigue moment.
4 Significance and application
4.1 part of the establishment of a unified method for static and dynamic bending test, and thus different types of spinal screw static and fatigue bending properties Properties were characterized and compared. This test method is used to determine the maximum bending moment in the spinal screw range conditions fatigue life, in addition, it may also be […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
ASTM E467 · ASTM E1942 · ASTM F1839
Similar standards
Editions of YY/T 0119.5
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0119.5-2014 | Spinal implants—Components used in the surgical fixation of the spinal skeletal system - Part 5: Test method for measure the static and fatigue bending strength of metallic screws | current edition | Current |
| YY 0119-2002 | Spinal implants—Components used in the surgical fixation of the spinal skeletal system - Part 5: Test method for measure the static and fatigue bending strength of metallic screws | previous edition | Superseded |
This page sells the current edition, YY/T 0119.5-2014. Earlier editions are listed for reference only.
How to Download YY/T 0119.5-2014
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 19701.2-2024 — Implants for surgery - Ultra-high-molecular-weight polyethylene - Part 2: Moulded forms
YY/T 0119.4-2014 — Spinal implants—Components used in the surgical fixation of the spinal skeletal system - Part 4: Metallic spinal rods
YY/T 0119.3-2014 — Spinal implants—Components used in the surgical fixation of the spinal skeletal system - Part 3: Metallic spinal plates
Secure payment via Stripe
Payments accepted
YY/T 0119.5-2014
$210.00