YY 1741-2021Antithrombin III testing kit (English PDF)
抗凝血酶III测定试剂盒
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY 1741-2021 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
March 9, 2021
Implementation date
April 1, 2023
Scope
YY 1741-2021 is the English-translated version of 抗凝血酶III测定试剂盒.
YY 1741-2021 is the Chinese pharmaceutical industry standard for the antithrombin III assay kit.
It sets the requirements, the test methods, the labelling and instructions for use and the rules for packaging, transport and storage, covering the appearance, the traceability of the calibrator, the accuracy, the linear range, the detection limit, the precision and the stability of the reagents.
It applies to kits that measure the heparin cofactor activity of antithrombin III by the chromogenic substrate method, and to kits such as the turbidimetric ones that measure its mass concentration. Issued on 9 March 2021, it applies from 1 April 2023.
Found what you need?
Document preview — YY 1741-2021
National Standard of the People's Republic of China
- Classification
- C 44
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
Introduction
In 1954, Seegers et al. first divided antithrombin into four types. Since then, the name antithrombin III has been used. International Health Organization (WHO) established the first-generation international reference product (72/1) in 1978, using the name antithrombin III. Made by British National Biology The second-generation international reference product (93/578) prepared by the National Institute for Quality Control (NIBSC) and established by the WHO in 1994, and the third-generation national reference product in 2010 The international reference (08/258) no longer uses the name "antithrombin III", but uses the name "antithrombin". Currently, International Thrombosis and Hemostasis The term for the meeting (ISTH) is also antithrombin. The names of products that have been listed on the domestic market are divided into two categories. antithrombin III determination kit and Antithrombin determination kit. Although the terminology and international reference products have been unified internationally, in view of the fact that the names of international products According to the actual registration situation of domestic products, antithrombin III is still used in the name of this standard.
The international unit of plasma antithrombin is established through a multi-center joint study based on the assignment of normal mixed plasma from each laboratory of. The pooled normal plasma of each laboratory is defined as 1IU, and the public mean is assigned to the standard. Currently, many coagulation diagnostic products are manufactured Manufacturers use the "%" value in their calibration plasma, where 100% is equivalent to 1IU, and use the value of "%" in their calibration plasma and/or diagnostic kits.
1 Scope
This standard specifies the requirements, test methods, labels and instructions for use, packaging, transportation and storage of antithrombin III assay kits.
This standard is applicable to the kit for the determination of antithrombin III based on the chromogenic substrate method of heparin cofactor activity (hereinafter referred to as the kit).
It is suitable for kits for measuring the mass concentration of antithrombin III such as turbidimetric method.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version is applicable to this file. For undated reference documents, the latest version (including all amendments) is applicable to this document.
GB/T 29791.2 Information provided by manufacturers of in vitro diagnostic medical devices (labeling) Part 2: In vitro diagnostic reagents for professional use
3 Requirements
3.1 Appearance
The appearance should meet the following requirements.
a) The components of the kit should be complete and complete, without leakage, and the label should be clear and easy to identify;
b) The liquid reagent should be a homogeneous solution;
c) The lyophilized product will be a homogeneous solution after reconstitution.
3.2 Quantity
The amount of liquid reagents should not be less than the labeled value.
3.3 Linearity
The linearity of the kit covers at least 20%~140%. In the linear interval, the slope of the linear regression equation is in the range of 1±0.05, and the correlation The coefficient r>=0.980.
3.4 Repeatability
The repeatability test uses samples with high, medium and low concentration levels. The coefficient of variation (CV) for high-concentration samples should be <=10%, and medium-concentration samples The coefficient of variation (CV) should be <=10%, and the coefficient of variation (CV) of low-concentration samples should be <=15%.
3.5 Accuracy
Select one of the following substances to repeat the test 3 times, and calculate the relative deviations respectively, which should not exceed ±15.0%.
a) Use recognized reference materials or certified reference materials (CRM);
b) Use corporate reference materials traceable to recognized reference materials or certified reference materials.
3.6 Detection limit
The detection limit should not be greater than 20%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
How to Download YY 1741-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 29791.2-2013 — In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 2: In vitro diagnostic reagents for professional use
YY 0017-2026 — Implants for osteosynthesis - Metallic bone plates
YY 0018-2026 — Implants for osteosynthesis - Metallic bone screws
YY 1741-2021
$105.00