YY 0901-2013 is the English-translated version of 紫外治疗设备.
YY 0901-2013 is the Chinese pharmaceutical industry standard for ultraviolet therapy equipment used in medical practice.
It specifies the terms and definitions, the classification, the requirements, the test methods, the inspection rules, the signs and the manual and the rules for packaging, transport and storage, covering the spectral range and the irradiance delivered at the treatment distance and their accuracy, the lamp and its ageing, the timer, the protection of the eyes and of the untreated skin, and the electrical safety. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY 0901-2013
National Standard of the People's Republic of China
- ICS
- 11.040.60
- Classification
- C 42
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions GB 9706.1-2007 defined and the following
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
The standard safety requirements for full implementation of the GB 9706.1-2007 "Medical Electrical Equipment Part 1. General requirements for safety," the regulation Set the standard GB 9706.1-2007 the relevant provisions were amended and supplemented.
1 Scope
This standard specifies the terms and definitions of UV treatment equipment, classification, requirements, test methods, inspection rules, signs and manual, package Loading, transport and storage.
UV treatment device for use (hereinafter referred to as the device) of the requirements of this standard applies 3.1 defined in medical practice. This standard does not apply to the following devices.
--- Instruments and materials for irradiating ultraviolet disinfection or sterilization equipment only;
--- Curing light;
--- Ultraviolet laser equipment;
--- UV-sensitive treatment equipment;
--- Ultraviolet blood irradiation apparatus.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 191 Packaging - Pictorial signs
GB 9706.1-2007 Medical electrical equipment - Part 1. General requirements for safety
GB/T 14710 medical electrical environmental requirements and test methods
GB/T 16886.1 Biological evaluation of medical devices - Part 1. Evaluation of the risk management process and Test Symbols - Part 1 YY/T 0466.1 medical equipment used with medical device labels, labeling and information. General requirements
3 Terms and Definitions
GB 9706.1-2007 defined and the following terms and definitions apply to this document.
3.1 Ultraviolet therapy equipment
Use effective wavelength at 200nm ~ 400nm ultraviolet radiation on the human body radiation therapy equipment.
Note: In general, ultraviolet light is divided into three bands. UVA (400nm ~ 320nm), UVB (320nm ~ 275nm), UVC (275nm ~ 200nm).
3.2 Effective irradiated area enablediluminatedarea
Area for irradiation treatment specified by the manufacturer, usually including the irradiated area, distance, angle, shape and other information.
3.3 UV radiation dose
UV radiation energy by unit area on the photograph, which is equal to the product of the intensity of UV radiation and irradiation time, in units of J/cm2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 9706.1-2007Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
How to Download YY 0901-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2007 — Medical electrical equipment - Part 1: General requirements for safety
GB/T 14710-1993 — The environmental requirements and test methods for medical electrical equipment
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
YY 0901-2013
$165.00