YY 0885-2013Medical electrical equipment - Part 2: Particular requirements for the safety, including essential performance of ambulatory electrocardiographic systems (English PDF)
医用电气设备 第2部分:动态心电图系统安全和基本性能专用要求
This is a limited preview
Buy now to download the full PDF (24 pages)
Need quotation for your management?
Get a formal quotation for YY 0885-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY 0885-2013 is the English-translated version of 医用电气设备 第2部分:动态心电图系统安全和基本性能专用要求.
YY 0885-2013 is the Chinese pharmaceutical industry standard for the safety and the essential performance of the ambulatory electrocardiographic system, the recorder and analysis software with which the heart is monitored over a day or more.
It applies together with the general standard for medical electrical equipment, whose clauses it modifies one by one, and covers the identification, marking and documents, the input power and the batteries, the protection against electric shock, the accuracy of the recorded and reproduced signal, the detection and reporting of events and the environmental conditions of use. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY 0885-2013
National Standard of the People's Republic of China
- ICS
- 11.040.55
- Classification
- C 39
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Introduction The first chapter Overview
- 1 Scope and purpose
- 2 Terms and definitions
- 5 Category
- 6 Identification, marking and documents
- 10 4 environmental conditions Third of the risk of electric shock protection
- 20 Dielectric strength
- 21 Mechanical strength
- 36 Electromagnetic compatibility *
- 7 accuracy and prevent the risk of output data of the eighth chapter of the work
- 7 50 Operating data accuracy Chapter 9. abnormal operation and fault conditions; environmental testing
- 56 components and assembly
- Annex L (normative) Normative documents
- Annex AA (informative) Theoretical and guidance Conduction emission test apparatus of FIG. 101 (see this standard 36.201.1)
- 102 radiated emission and immunity test apparatus of FIG. (See this standard 36.201.1 and 36.201.2)
- 103 input dynamic range of the test signal (see 51.5.1 of this standard)
- 106 pacemaker pulse tolerance test circuit (see this standard 51.5.11)
Foreword
Medical electrical equipment standard series of standards, the series standard consists of two parts.
--- Part 1. General requirements for medical electrical equipment safety;
--- Part 2. Safety requirements for medical electrical equipment.
The specific standard is "Medical electrical equipment - Part 2-47. Part 2 Medical electrical equipment. Holter system security and groups This special performance requirements. " This is a specific standard GB 9706.1-2007 "Medical Electrical Equipment Part 1. General requirements for safety" of repair Changes and additions to the specific standards take precedence over generic standards.
This dedicated drafting standards in accordance with GB/T 1.1-2009 given rules.
The use of specific standard redrafted law, amend the International Electrotechnical Commission standard IEC 60601-2-47:2001 "Medical Electrical Equipment Part 2-47. Holter system safety and essential performance requirements "(in English). This standard and IEC 60601-2-47:2001 main differences are.
--- IEC 60601-2-47:2001 in Chapter 36 is juxtaposed standard IEC 60601-1-2:1993 complements our current and electromagnetic Compatibility YY 0505-2011 "Medical electrical equipment - Part 1-2. General requirements for safety - Collateral standard. Electromagnetic and Capacity requirements and testing "identical with IEC 60601-1-2:2004. Thus, in this specific standard-setting process, technology standardization Operation Committee believes that it is necessary to refer to the latest parallel standards, updates to electromagnetic compatibility requirements.
--- Other international standards quoted in the standard, if it has accordingly into our standards, places of reference standards prevail.
--- Remove IEC 60601-2-47:2001 Introduction, according to China's industry standards rewritten.
1 Scope and purpose
Except as follows, common standards in this chapter apply.
1.1 Scope
Addition.
This standard specifies the special safety requirements applicable to specific definitions of 2.101 Holter system.
The following types of systems within the scope of this standard.
Gives basic when a) continuous ECG recording and analysis system, and the system for a complete re-analysis of ECG The same results. The system can be recorded and stored electrocardiogram, and then in a separate unit for analysis or record And analysis at the same time. This standard does not involve the use of the type of storage media;
b) to provide continuous analysis and partial or limited record, but not a complete re-ECG analysis system.
Meet any of these types of systems are applicable to the standard safety requirements.
If Holter ECG analysis system provides automated, the applicable minimum performance requirements of this standard of measurement and analysis functions.
By the GB 10793-2000 and GB 9706.25-2005 covered medical electrical equipment not within the scope of this standard.
This standard does not apply to the ECG can not be continuous recording and analysis system (for example, "intermittent event recorders").
1.2 Purpose
replace.
The specific standard is intended to develop specific requirements for Holter system safety and essential performance.
1.3 Specific Standard
Addition. The specific reference standard GB 9706.1-2007.
It said in this specific standards in Part 1 of the "General Standard" or "General requirements." The specific standards of each chapter, and all chapters of the number corresponding to the General Standard, General Standard for changes in the terms expressed by the following provisions. "Replace" indicates that common standards chapter or completely replaced by the terms of this specific standard provision. "Supplement" means this specific standard provisions to increase the general standard. "Modify" means to press the special provisions of this Standard to the General Standard or chapter shall be modified.
General Standard for additional terms or FIG numbered starting from 101, additional annexes are lettered as AA, BB, etc., increased column items As aa), bb) and the like.
The term "this standard" is used to refer to the General Standard and this Particular Standard collectively.
When universal standard articles, chapters and bars without the corresponding articles, chapters and clauses in this specific standard, even may not be relevant, and no change to Be applicable.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 24 pages — is available in the English PDF.
Similar standards
How to Download YY 0885-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY 0017-2026 — Implants for osteosynthesis - Metallic bone plates
YY 0018-2026 — Implants for osteosynthesis - Metallic bone screws
YY 0118-2026 — Joint replacement implants - Hip joint prostheses - Fundamental requirements
Secure payment via Stripe
Payments accepted
YY 0885-2013
$475.00