YY 0600.4-2013Lung ventilators for medical use—Particular requirements for basic safety and essential performance - Part 4: operator-powered resuscitators (English PDF)
医用呼吸机基本安全和主要性能专用要求 第4部分:人工复苏器
This is a limited preview
Buy now to download the full PDF (22 pages)
Need quotation for your management?
Get a formal quotation for YY 0600.4-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY 0600.4-2013 is the English-translated version of 医用呼吸机基本安全和主要性能专用要求 第4部分:人工复苏器.
YY 0600.4-2013 is the Chinese pharmaceutical industry standard that sets the particular requirements for the basic safety and essential performance of the operator-powered resuscitator, the hand-operated bag with which lung ventilation is given.
It is the fourth part of the YY 0600 series and applies to portable resuscitators for patients of every age, those for infants and children being identified by the weight range they suit and the approximate age that goes with it.
It covers the delivered volume and pressure, the valves and the connectors, the resistance and the dead space, the oxygen enrichment, the operation at extreme temperatures and the marking.
Electrically and pneumatically powered resuscitators are outside its scope. Issued on 21 October 2013, it has applied since 1 October 2014.
Found what you need?
Document preview — YY 0600.4-2013
National Standard of the People's Republic of China
- ICS
- 11.040.10
- Classification
- C 46
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 connector
- 4.1 Patient connection port connector
- 4.2 * Exhale connector 3 for breathing gas
- 4.3 Mask connector
- 4.4 * Airbag valve connector
- 4.5 Airbag inlet valve connector
- 4.6 Threaded Gas Filter Connector
- 4.7 Oxygen fittings and manometer fittings
- 5 Operating requirements
- 5.1 General
- 5.2 Disassembly and assembly
- 5.3 Patient valve function after vomit contamination
- 5.4 Mechanical shock
- 5.5 Immersion in water
- 5.6 Airbag inflation valve
- 5.7 Construction materials
- 6 Ventilation requirements
- 6.1 Supplemental oxygen and delivered oxygen concentration
- 6.2 Expiratory impedance
- 6.3 Inspiratory impedance
- 6.4 Patient valve failure
- 6.5 Patient valve leak-forward leak
- 6.6 Resuscitator Dead Space and Repeated Breathing
- 6.7 Ventilation performance
- 7 Storage and handling conditions
- 7.1 Storage
- 7.2 Operating conditions
- 8 Requirements for sterile-packed resuscitators or components
- 8.1 Aseptic Assurance
- 8.2 Aseptic packaging of resuscitators or components
- 9 Mark
- 9.1 General
- 9.2 Indication of operating conditions
- 9.3 Indication of pressure limit system settings
- 10 Information to be provided by the manufacturer in the operation and maintenance instructions
- 10.1 General
- 10.2 Content
- Appendix A (Normative) Test Method
- Appendix B (Informative) Basic Principles
- References
Foreword
The general title of YY 0600 is `` Special Requirements for Basic Safety and Main Performance of Medical Ventilators '', which consists of the following parts.
--- Part 1. Household respiratory support equipment;
--- Part 2. Home ventilator for ventilator dependent patients;
--- Part 3. Ventilator for emergency and transport;
--- Part 4. artificial resuscitator;
--- Part 5. Pneumatic emergency resuscitator. This section is Part 4 of YY 0600.
This section is drafted in accordance with the rules given in GB/T 1.1-2009.
This part is equivalent to the international standard ISO 10651-4.:2002 "Medical ventilator Part 4. Specific requirements for artificial resuscitators" (English Version), and made the following editorial changes.
--- Modify the standard name to "Special Requirements for Basic Safety and Main Performance of Medical Ventilators-Part 4. Artificial Resuscitators";
--- The international standards quoted in ISO 10651-4.:2002 are correspondingly adopted as national standards and industry standards.
Use these national standards and industry standards as the normative use; if there is no corresponding adoption as national standards and industry standards, the The cited international standards are used as norms.
1 Scope
This part of YY 0600 specifies special requirements for portable artificial resuscitators applicable to all ages, for Pulmonary ventilation is provided in separate personnel. For infants and children, artificial resuscitators are identified based on their weight range and their corresponding approximate age.
Electric resuscitators and pneumatic resuscitators are not included in the scope of this section.
Note: The basic principles of this section are listed in Appendix B. Article numbers with principle descriptions are marked with "*".
2 Normative references
The following documents are essential for the application of this document. For dated references, only the dated version applies to this document. For undated references, the latest version (including all amendments) applies to this document.
GB/T 4999-2003 Anesthesia breathing equipment term (ISO 4135.:2001, IDT) A valve that opens when the pressure in the compression unit of the resuscitator is lower than atmospheric pressure and inflates the compression unit with gas at ambient air pressure.
YY/T 0615.1-2007 Requirements for marked "sterile" medical devices Part 1. Requirements for final sterilized medical devices (EN 556-1.:2001, IDT) A valve that opens when the pressure in the compression unit of the resuscitator is lower than atmospheric pressure and inflates the compression unit through the compressed air source. The valve is unmanned Trigger.
YY 0801.1 Terminals for medical gas piping systems-Part 1. Terminals for compressed medical gases and vacuum (YY 0801.1-2010, ISO 9170-1.:2008, IDT) A component on an artificial resuscitator, such as an air bag or bellows, can send out a certain volume of gas when it is squeezed by the operator.
YY 1040.1 Conical joints for anesthesia and breathing equipment-Part 1. Taper heads and sleeves 1..1996, IDT) A connector at the patient connection port that is directly connected to a mask or a suitable mating airway device.
EN 148-1 Threads for respiratory protective masks-Part 1. Standard threaded connections (Respiratoryprotective) devices-Threadsforfacepieces-Part 1. Standardthreadconnection.) A valve in the respiratory system that allows gas to enter the lungs in the inspiratory phase and the atmosphere in the expiratory phase.
EN 868-1 Packaging materials and systems for sterilization of medical articles. Part 1. General requirements and test methods materialsandsystemsformedicaldeviceswhicharetobesterilized-Part 1. Generalrequirementsand testmethods.)
EN 1041 Information provided by medical device manufacturers (Informationsupplied by themanufacturerwithmedical devices.) prEN13544-2..2000 Respiratory therapy equipment-Part 2. Specific requirements for tubing and fittings (Respiratorytherapyequip-ment-Part 2. Specificationsfortubingandconnectors.)
3 Terms and definitions
The terms and definitions defined in GB/T 4999-2003 and the following apply to this document.
Note: Part of the definition comes from GB/T 4999, but for the sake of convenience, this section also includes them; GB/T 4999 gives the general equipment He defined that this section has been slightly modified because it is specific to resuscitators.
3.1 Reverseleakage
No outflow from the exhalation port, but back to the exhaled breath volume of the resuscitator.
3.5 Delivered oxygen concentration
The average oxygen concentration of the gas delivered by the resuscitator.
3.6 Delivered volume, Vdel
In the inspiratory phase, the volume of gas output from the resuscitator through the patient connection port is expressed in milliliters.
3.7 Forwardleakage
The volume of gas that is delivered from the resuscitator in the inspiratory phase and is not delivered to the patient via the patient end, but leaks to the atmosphere.
3.8 Minutevolume
Volume of gas inhaled or exhaled by the patient's lungs per minute.
3.9 Artificial resuscitator
Resuscitation equipment for lung ventilation is achieved by the operator pressing the compression unit on the device.
Note: hereinafter referred to as "resuscitator".
3.10 Patient connection port
The patient-side interface on the resuscitator, through which the patient inhales and exhales gas.
3.13 Pressure limiting system
Device for limiting maximum delivery pressure
3.14 Resuscitator dead space
The volume of the exhaled gas from the previous exhalation in the next inspiratory phase. (VD, app)
3.15 Tidal volume
The volume of gas that enters during the inspiratory phase, or leaves the patient during the expiratory phase, or simulates the lungs, is expressed in milliliters.
3.16 Ventilation cycle
The ventilation cycle contains the inspiratory and expiratory phases of breathing.
4 Connector
4.2 * Exhale connector for breathing gas If equipped with an exhalation port connector, it should be one of the following.
a) 30mm outer conical joint according to YY 1040.1;
b) Permanent or special connectors that are not compatible with YY 1040.1 and YY 0801.1;
It cannot be connected to the lumen of any breathing accessory.
4.4 * Airbag Inflation Valve Connector If the airbag inflation valve is equipped with a conical joint, it should be a special 32mm inner conical design and have a good fit with the gauge shown in Figure A.1. Cooperation, the error is within the allowable range.
4.1 Patient connection port connector
The patient connector of the resuscitator shall be a 15 mm inner cone and 22 mm outer cone joint in accordance with ISO 5356-1.
4.3 Mask connector
If the resuscitator is equipped with a mask, the mask should have a 22mm inner conical joint, or a 15mm outer conical joint, which should conform to YY 1040.1 The specified connectors match.
4.5 Airbag inlet valve connector
The airbag inlet valve connector should not match the connector in accordance with YY 1040.1. Airbag intake valves should be designed to minimize Possibility of misconnection of the plug valve's breathing attachment.
4.6 Threaded Gas Filter Fitting
If the resuscitator can be fitted with a threaded gas filter fitting, it should comply with EN 148-1.
4.7 Oxygen fittings and pressure gauge fittings
If equipped with an oxygen pipe joint, it shall comply with prEN13544-2..2000. Manometer fittings (if equipped) should not The tubing of the connector is matched.
5 Operating requirements
5.2 * Disassembly and assembly Resuscitators intended to be disassembled by the user, if disassembled for cleaning purposes, should be designed so that there is a risk of incorrect assembly between the various components lowest.
The manufacturer shall recommend a method of performance inspection performed after reassembly [see 10.2d)].
5.3 * Patient valve function after vomit contamination After testing in accordance with A.4.3, the resuscitator shall meet the requirements of 6.2, 6.4, 6.7.1 and 6.7.2.
5.6 * Airbag inflation valve There should be no manually operated device on the airbag inflation valve used with the resuscitator.
5.1 General
When the resuscitator is operated by a single person, all performance requirements in this section should be met.
Note: The structure of the valve chamber is preferably designed so that its mechanical operation can be observed by the operator, such as using a transparent housing. Observe the mechanism of the patient valve to help operate Avoid incorrect operation.
5.4 Mechanical shock
5.4.1 * Drop test At room temperature, after the drop test according to A.4.4, the resuscitator should meet the requirements of 6.2, 6.4, 6.7.1.
5.5 Immersion
After immersion in water according to the method of A.4.5, the resuscitator shall meet the requirements of 6.2, 6.4, 6.7.1 and 6.7.2.
5.7 Construction materials
All air-conducting parts shall be made of materials that have been declared by the manufacturer to be used in artificial resuscitators, taking into account their physical and chemical characteristics Manufacturing.
6 Ventilation requirements
6.1 * Supplemental oxygen and oxygen delivered According to the requirements of its corresponding classification (see 6.7.1), when tested by the method of A.4.6, the resuscitator is connected to the flow rate no higher than 15L/min oxygen source should provide at least 35% oxygen by volume; and the ability to provide 85% oxygen by volume (See note). The manufacturer should declare that it is conveyed at some representative flow rate (such as 2L/min, 4L/min, 6L/min, 8L/min, etc.) Range of oxygen concentration.
Note: The requirement of 85% volume fraction can be met by using accessories.
6.2 * expiratory impedance Without using positive end-expiratory pressure, test according to the method described in A.4.7, and the pressure generated at the connection port of the patient should not exceed 0.5kPa (~5cmH2O). [See also 10.2c) 8)].
6.3 * Inspiratory impedance According to the method described in A.4.8, the pressure generated at the connection port of the patient should not exceed 0.5kPa (~5cmH2O) at atmospheric pressure. [See also 10.2c) 8)].
6.4 * Patient valve failure Tested according to the method described in A.4.9, when the maximum flow rate of 30L/min is added according to the manufacturer's instructions, it should not cause more than Unexpected positive end expiratory pressure of 0.6 kPa (~6 cmH2O).
6.5 * Patient Valve Leak-Forward Leak When the resuscitator has a design feature of forward leakage, it should be explained in the instruction manual.
6.6 * Resuscitator Dead Space and Repeated Breathing Test according to the method described in A.4.10. According to the requirements of its corresponding classification, the dead space of the resuscitator should not exceed a minimum of 5mL (110%). Conveying capacity (see 6.7.1). During spontaneous breathing, excessive repeated breathing should not occur.
6.7 * Ventilation performance 6.7.1 * Minimum delivery capacity (Vdel)
According to the compliance, impedance, frequency, and breathing ratio given in Table 1, and test according to the method described in A.4.11, the minimum delivery capacity should conform to Table 1. Requirements.
6.7.2 * Pressure Limit
6.7.2.1 According to the method described in A.4.12, resuscitators designed for patients weighing less than 10 kg should be equipped with airway pressure limiting The pressure limit system does not exceed 4.5kPa (~45cmH2O). But it should be able to generate airway pressure of at least 3kPa (~30cmH2O).
Note: Super control device is available.
6.7.2.2 If a resuscitator designed for a patient weighing more than 10 kg is equipped with a pressure limitation system, Working pressure [see 10.2c) 9)]. Any pressure-limiting device that limits the pressure to less than 6kPa (~60cmH2O) should be equipped with a super-control device.
Home. If the override device is provided with a locking device, the override device should be designed so that the operator can Clearly identify the mode of operation (eg on or off). Compliance is checked by inspection.
7 Storage and handling conditions
7.2 * Operating conditions Test according to the method described in A.4.13, the resuscitator should be in the entire relative humidity range of 15% to 95% relative humidity, and in the following Meet the requirements of Chapter 6 in one of the temperature ranges.
7.1 Storage
Resuscitator and resuscitator kit (if available) should be arbitrarily relative between -40 °C ~ 60 °C and relative humidity 40% ~ 95% After storage under humidity, it still meets the requirements of Chapter 6 except for 6.6 (dead space).
---- 18 ° C ~ 50 ° C; or --- If given a specific operating range (see 9.2 and Chapter 10), it must be within the entire temperature range declared by the manufacturer.
8 Requirements for resuscitators or components in aseptic packaging
8.1 Aseptic guarantee
Resuscitators or components marked "sterile" should meet the requirements of EN 556-1.:2001 to ensure the sterility required to make this statement.
8.2 Aseptic packaging of resuscitators or components
Packaging should meet EN 868-1 and should effectively block the entry of microorganisms and particulate matter.
Such packaging should not be sealed once opened, otherwise it should be clearly stated that the packaging has been opened.
9 Mark
9.1 General
The marking, packaging, inserts and information provided by the manufacturer of the resuscitator or applicable parts shall comply with EN 1041.
Note: Some requirements of Chapter 9 can be met by using the corresponding symbols given in ISO 15223-1.
9.2 Indication of operating conditions
If the resuscitator cannot work normally between -18 °C ~ 50 °C specified in 7.2, there should be a warning mark on the device.
Note: Examples of warnings.
--- "For use only between °C ~ °C"; or the symbol shown in Figure 1.
9.3 Indication of pressure limit system settings
If the pressure limiting system equipped with the resuscitator is set to a fixed pressure, the nominal set pressure should be marked on the resuscitator.
10 Information to be provided by the manufacturer in the operation and maintenance instructions
10.1 General
Manufacturers are required to provide operation and maintenance instructions. These instructions should be sized and shaped to fit in the resuscitator box.
10.2 Content
In addition to the requirements of EN 1041, if applicable, the operation and maintenance instructions shall include the following.
a) warn of the danger of incorrect use of the resuscitator;
b) instructions on how to operate the resuscitator in all expected modes of operation;
c) Information detailing the following information about the resuscitator and its recommended accessories (if any).
1) Suitable weight range for resuscitator;
2) Operating environment restrictions;
3) Storage environment restrictions;
4) All contents that can be delivered by the resuscitator except air (e.g. oxygen, volatile anesthetic);
5) The oxygen concentration that can be delivered under various test conditions;
6) Range of gas capacity;
7) Dead space, backward leakage and forward leakage of the resuscitator;
8) expiratory impedance, inspiratory impedance and impedance added using any recommended accessories;
9) The end-expiratory pressure generated during normal use of the resuscitator should be stated if it is greater than 0.2kPa (~ 2cmH2O);
10) If there is a pressure limiting system and an override device, the details should be stated;
11) If the resuscitator is equipped with a control or indicating device, its accuracy should be stated;
12) external dimensions of the resuscitator and resuscitator box (if any);
13) the quality of the resuscitator and resuscitator box (if any);
d) instructions for disassembly and assembly of certain components for cleaning (including cleaning of vomitus), instructions for sterilization (if applicable), and assembly Detailed description of the performance test after operation.
Note: These instructions can also be given in schematic form.
e) recommended methods for cleaning, disinfecting or sterilizing resuscitators, components, and accessories (if applicable);
f) functional tests to be performed before use;
g) operator-replaceable parts list;
h) maintenance requirements;
i) Recommendations for use in hazardous or explosive environments, including if the resuscitator mixes with air or allows the patient to inhale from the atmosphere A warning that gas may be harmful if used in a contaminated environment (unless it is not mixed into the atmosphere). If applicable, the manufacturer shall describe How to prevent contamination or inhalation of contaminated gases, such as the use of filters;
j) Under high oxygen concentration conditions, smoke and open flames can cause danger warnings and warnings against the use of grease in resuscitators;
k) The sterile resuscitator or component shall indicate the sterilization method used.
Appendix A (Normative appendix) experiment method
A.1 General test conditions
Unless otherwise specified, the ambient temperature of the test should be between 20 ° C and 25 ° C, and the relative humidity should be between 45% and 75%.
A.2 Instrument
A.2.1 General
Typical test instruments are shown in Figures A.1 to A.5; test instruments with equal or better accuracy can also be used.
A.2.2 Simulated lung (see Figure A.4 and Figure A.5 for examples), with appropriate compliance and impedance characteristics (see Table A.1 and Table A.2).
A.2.3 Flow resistance if the simulated lungs are not provided together. A.2.4 Instruments for measuring and recording pressure, flow and capacity.
A.2.5 Temperature measuring instrument
A.2.6 Instrument for measuring dead space (see A.3 for a typical example). A.2.7 graduated cylinder, at least 200mL volume.
A.2.8 Oxygen analyzer
A.2.9 A large enough pool to completely submerge the resuscitator.
A.2.10 Environmental test box, which can maintain the temperature at -40 °C ± 1 °C ~ 70 °C ± 1 °C, and maintain the relative humidity at 15% ~ 95%. Hold for up to 7d.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 22 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 4999-2003Anaesthetic and respiratory equipment Vocabulary
- YY/T 0615.1-2007Requirements for medical devices to be designated STERILE - Part 1: Requirements for terminally sterilized medical devices
- YY 0801.1Terminal units for medical gas pipeline systems - Part 1: Terminal units for use with compressed medical gases and vacuum
- YY 1040.1Anaesthetic and respiratory equipment - Conical connectors - Part 1: Cones and sockets
EN 148-1 · EN 868-1 · EN 1041
Similar standards
How to Download YY 0600.4-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 4999-2003 — Anaesthetic and respiratory equipment Vocabulary
YY 0801.1-2010 — Terminal units for medical gas pipeline systems - Part 1: Terminal units for use with compressed medical gases and vacuum