YY 0127.4-2009Biological evaluation of medical devices used in dentistry - Part 2: Test methods - Bone implantation test (English PDF)

口腔医疗器械生物学评价 第2单元:试验方法 骨埋植试验

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for YY 0127.4-2009 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

December 30, 2009

Implementation date

June 1, 2011

Scope

YY 0127.4-2009 is the English-translated version of 口腔医疗器械生物学评价 第2单元:试验方法 骨埋植试验.

YY 0127.4-2009 is the Chinese standard "Biological evaluation of medical devices used in dentistry - Part 2: Test methods - Bone implantation test". Its scope clause reads: This Standard specifies the test method of bone implantation of dental medical devices.

This Standard shall be used to evaluate the biological reactions of bone tissue of which the dental bone issue is planned to contact with dental medical devices in long-term. Its clauses include sample preparation.

Found what you need?

Document preview — YY 0127.4-2009

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope6
  • 2 Normative References6
  • 3 Sample Preparation6
  • 4 Test Cycles7
  • 5 Experimental Animals7
  • 6 Test Position7
  • 7 Surgical Procedure8
  • 8 Post-operative Evaluation8
  • 9 Killing of Animal8
  • 10 Biological Reaction Evaluation8
  • Appendix A13
  • References16
  • Translation References and Original Chinese Documents17

1 Scope

This Standard specifies the test method of bone implantation of dental medical devices.

This Standard shall be used to evaluate the biological reactions of bone tissue of which the dental bone issue is planned to contact with dental medical devices in long-term.

2 Normative References

The articles contained in the following documents have become part of this Standard when they are quoted herein. For the dated documents so quoted, all the modifications (excluding corrections) or revisions made thereafter shall not be applicable to this Standard. For the undated documents so quoted, the latest edition shall be applicable to this Standard.

GB/T 13810 Wrought titanium and titanium alloy for surgical implants

GB/T 16886.6 Biological evaluation of medical devices - Part 6.Tests for local effects after implantation (GB/T 16886.6-1997, ISO 10993-6.1994, IDDT)

GB/T 16886.12 Biological evaluation of medical devices - Part 12.Sample preparation and reference materials (GB/T 16886.12 2005, ISO 10993-12.2005, IDT)

3 Sample Preparation

3.1

CONDUCT preparation, processing, cleaning, and disinfection for the implanted materials according to the same method used for final product. At least 10 samples shall be used for each test cycle.

3.1.1 Solid sample

sample shall be made in cylinder with diameter of 2 mm, height of 6 mm, with smooth surface.

same position; so as to ensure the minimum influence of relative movement between tissue and the implants.

For cylindrical sample, the observation position shall be at the central part of the column; for slotted cylindrical implants, the central part of the slot and the top flat surface are suitable for evaluation.

10.1 Visual evaluation

USE low-magnification loupe to observe the cut tissue from the implanting position; it shall include the evaluation of local draining lymph nodes. RECORD the characteristics and degree of the observed tissue reaction, including hematoma, edema, fiber wrapping and (or) any general findings. RECORD the implant's state, shape and location, including the possible residues of biodegradable materials. Observation results shall be archived with photographs.

In addition, when observing the implanting position, if the animal exhibits pathological or reaction to the implant, gross necropsy shall be carried out if necessary.

10.2 Sampling and preparation of tissue sample

In order to evaluate the local biological reaction, cut tissue samples shall include those sufficient unaffected tissue from the implanting position and surrounding areas.

If the implant is not obvious in the observation position (biodegradable material/ absorbable material), it shall expand the sampling area, including those normal tissue located within few millimeters. For non-degradable materials, it shall include the abnormal draining lymph nodes shown by the result of gross pathology. For biod...

END: Final Published Version (Excerpted from Official PDF)

YY/T 0127.13 | YY/T 0127.7 | YY 0270.1 | YY/T 0127.4 | […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Similar standards

How to Download YY 0127.4-2009

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0127.4-2009

$260.00