Scope

YY 0003-1990 (Manual medical bed) is available as an English-translated PDF.

YY 0003-1990 — This document specifies the requirements, test methods and instructions for manual hospital beds. This document applies to manual beds. This product is intended for use with a height of 146 cm or more, a mass of 40 kg or more, and a weight of Patients with a BMI equal to or greater than 17.

Found what you need?

Document preview — YY 0003-1990

National Standard of the People's Republic of China

ICS
11.040.99
Classification
C 46
Replacing
YY/T 0003-1990

Issued by: State Drug Administration

Contents

  • Preface
  • 1 Scope
  • 2 Normative references
  • 3 Terms and Definitions
  • 4 Requirement
  • 4.1 Angle
  • 4.2 Preventing the patient from being trapped by non-moving parts
  • 4.3 Strength requirements for patient or operator support or suspension systems
  • 4.4 Instability-Imbalance
  • 4.5 Pushing force
  • 4.6 Sidebar Strength and Latch Reliability
  • 4.7 Preventing patients from accidentally falling
  • 4.8 Rough handling
  • 4.9 Headboard/footboard assembly
  • 4.10 Operating device
  • 4.11 Requirements for manual beds
  • 4.12 Annex
  • 4.13 Appearance
  • 4.14 Logo
  • 5 Test methods
  • 5.1 Angle
  • 5.2 Preventing patients from being trapped by non-moving parts
  • 5.3 Strength requirements for patient or operator support or suspension systems
  • 5.4 Instability-Imbalance
  • 5.5 Pushing force
  • 5.6 Sidebar Strength and Latch Reliability
  • 5.7 Preventing patients from accidentally falling
  • 5.8 Rough handling
  • 5.9 Headboard/footboard assembly
  • 5.10 Operating device
  • 5.11 Requirements for manual beds
  • 5.12 Annex
  • 5.13 Appearance
  • 5.14 Logo
  • 6 Instructions
  • Reference

Foreword

This document is in accordance with the provisions of GB/T 1.1-2020 "Guidelines for standardization work Part 1: Structure and drafting rules for standardization documents" Drafting is required.

This document replaces YY/T 0003-1990 "Hospital Beds". Compared with YY/T 0003-1990, except for structural adjustments and editorial changes, in addition, the main technical changes are as follows.

--- Added terminology items and related definitions (see Chapter 3, Chapter 3 of the 1990 edition);

--- Changed the angle requirements (see 4.1, 3.4 of the 1990 edition);

--- Changed the strength requirements (see 4.3, 4.8 and 4.9 of the 1990 edition);

--- Added "preventing patients from falling accidentally" and changed the height of the bed rails (see 4.7, 4.16.4 of the 1990 edition);

--- Added technical requirements "prevent patients from being trapped by non-moving parts" (see 4.2), "instability" (see 4.4), "driving force" (see 4.5), "Sidebar strength and latch reliability" (4.6), "Identification" (4.14);

--- Changed the "operating device" requirements (see 4.10, 4.7 of the 1990 edition);

--- Changed the coordination requirements for manual beds (see 4.11, 4.5 of the 1990 edition);

--- Changed the requirements for the appendix (see 4.12, 4.16 of the 1990 edition);

--- The test method has been changed (see Chapter 5, Chapter 5 of the 1990 edition);

--- Added relevant requirements for instructions (see Chapter 6);

--- Deleted the classification (see Chapter 3 of the 1990 edition);

--- Deleted the inspection rules (see Chapter 6 of the 1990 edition);

--- Deleted marking, packaging, transportation and storage (see Chapter 7 of the 1990 edition);

--- Deleted the warranty period (see Chapter 8 of the 1990 edition).

1 Scope

This document specifies the requirements, test methods and instructions for manual hospital beds.

This document applies to manual beds. This product is intended for use with a height of 146 cm or more, a mass of 40 kg or more, and a weight of Patients with a BMI equal to or greater than 17.

2 Normative references

This document has no normative references.

3 Terms and definitions

The following terms and definitions apply to this document.

Patient

People who are receiving treatment.

[Source. YY 9706.252-2021,.201.3.76, modified] 3.2 in hospitals that provide intensive care, 24-hour medical supervision or continuous monitoring is required, and the use of biological A place where life support systems/equipment are used to maintain or enhance patient function.

[Source. YY 9706.252-2021,.201.3.201] 3.3 in hospitals or other medical institutions providing emergency care, medical supervision and monitoring are required and are often used during medical procedures.

A place where medical equipment is used to maintain or improve a patient's condition.

[Source. YY 9706.252-2021,.201.3.202] 3.4 There are areas with long-term care, where medical supervision and monitoring are required and where necessary, where medical care may be used during the course of treatment.

A place where equipment is used to maintain or improve a patient's condition.

Note. Includes use in nursing homes, rehabilitation facilities and facilities for the elderly.

[Source. YY 9706.252-2021,.201.3.203] 3.5 A place in the home where care is provided where medical equipment is used to relieve or compensate for injury, disability, or illness.

NOTE. When a manual bed is designed solely for use in application environment 4, all other application environments (e.g. nursing homes, rehabilitation facilities and facilities for the elderly) are excluded.

[Source. YY 9706.252-2021,.201.3.204, modified]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.

How to Download YY 0003-1990

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
21 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0003-1990

$494.00